Specificity Characteristics of 7 Commercial Creatinine Measurement Procedures by Enzymatic and Jaffe Method Principles

被引:116
作者
Greenberg, Neil [1 ]
Roberts, William L. [2 ]
Bachmann, Lorin M. [3 ]
Wright, Elizabeth C. [4 ]
Dalton, R. Neil [5 ]
Zakowski, Jack J. [6 ]
Miller, W. Greg [3 ]
机构
[1] Ortho Clin Diagnost, Rochester, NY USA
[2] Univ Utah, Salt Lake City, UT USA
[3] Virginia Commonwealth Univ, Richmond, VA USA
[4] NIDDK, NIH, Dept Hlth & Human Serv, Bethesda, MD USA
[5] Evelina Childrens Hosp, Kings Coll London, WellChild Lab, London, England
[6] Beckman Coulter Inc, Brea, CA USA
关键词
GLOMERULAR-FILTRATION-RATE; SERUM CREATININE; STABILITY; EQUATION; DISEASE; ASSAYS;
D O I
10.1373/clinchem.2011.172288
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Standardized calibration does not change a creatinine measurement procedure's susceptibility to potentially interfering substances. METHODS: We obtained individual residual serum or plasma samples (n = 365) from patients with 19 different disease categories associated with potentially interfering substances and from healthy controls. Additional sera at 0.9 mg/dL (80 mu mol/L) and 3.8 mg/dL (336 mu mol/L) creatinine were supplemented with acetoacetate, acetone, ascorbate, and pyruvate. We measured samples by 4 enzymatic and 3 Jaffe commercially available procedures and by a liquid chromatography/isotope dilution/mass spectrometry measurement procedure against which biases were determined. RESULTS: The number of instances when 3 or more results in a disease category had biases greater than the limits of acceptability was 28 of 57 (49%) for Jaffe and 14 of 76 (18%) for enzymatic procedures. For the aggregate group of 59 diabetes samples with increased beta-hydroxybutyrate, glucose, or glycosylated hemoglobin (Hb A(1c)), the enzymatic procedures had 10 biased results of 236 (4.2%) compared with 89 of 177 (50.3%) for the Jaffe procedures, and these interferences were highly procedure dependent. For supplemented sera, interferences were observed in 11 of 24 (46%) of groups for Jaffe and 8 of 32 (25%) of groups for enzymatic procedures and were different at low or high creatinine concentrations. CONCLUSIONS: There were differences in both magnitude and direction of bias among measurement procedures, whether enzymatic or Jaffe. The influence of interfering substances was less frequent with the enzymatic procedures, but no procedure was unaffected. The details of implementation of a method principle influenced its susceptibility to potential interfering substances. (C) 2011 American Association for Clinical Chemistry
引用
收藏
页码:391 / 401
页数:11
相关论文
共 18 条
  • [1] [Anonymous], JOINT COMM TRAC LAB
  • [2] [Anonymous], 2010, COLL AM PATH CREAT S
  • [3] BOWERS LD, 1980, CLIN CHEM, V26, P555
  • [4] BOWERS LD, 1980, CLIN CHEM, V26, P551
  • [5] Reference intervals for serum creatinine concentrations: Assessment of available data for global application
    Ceriotti, Ferruccio
    Boyd, James C.
    Klein, Gerhard
    Henny, Joseph
    Queralto, Josep
    Kairisto, Veli
    Panteghini, Mauro
    [J]. CLINICAL CHEMISTRY, 2008, 54 (03) : 559 - 566
  • [6] CLSI, 2005, EP7A2 CLSI
  • [7] Prime Time for Enzymatic Creatinine Methods in Pediatrics
    Cobbaert, Christa M.
    Baadenhuijsen, Henk
    Weykamp, Cas W.
    [J]. CLINICAL CHEMISTRY, 2009, 55 (03) : 549 - 558
  • [8] Certification of creatinine in a human serum reference material by GC-MS and LC-MS
    Dodder, Nathan G.
    Tai, Susan S. -C.
    Sniegoski, Lorna T.
    Zhang, Nien F.
    Welch, Michael J.
    [J]. CLINICAL CHEMISTRY, 2007, 53 (09) : 1694 - 1699
  • [9] STABILITY OF 25 ANALYTES IN HUMAN SERUM AT 22-DEGREES-C, 4-DEGREES-C, AND -20-DEGREES-C
    DONNELLY, JG
    SOLDIN, SJ
    NEALON, DA
    HICKS, JM
    [J]. PEDIATRIC PATHOLOGY & LABORATORY MEDICINE, 1995, 15 (06): : 869 - 874
  • [10] CHARACTERIZATION OF CREATININE ERROR IN KETOTIC PATIENTS - A PROSPECTIVE COMPARISON OF ALKALINE PICRATE METHODS WITH AN ENZYMATIC METHOD
    GERARD, SK
    KHAYAMBASHI, H
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1985, 84 (05) : 659 - 664