Contribution of usage to endoscope working channel damage and bacterial contamination

被引:15
作者
Santos, L. C. S. [1 ,2 ]
Parvin, F. [1 ]
Huizer-Pajkos, A. [1 ]
Wang, J. [1 ]
Inglis, D. W. [3 ]
Andrade, D. [2 ]
Hu, H. [1 ]
Vickery, K. [1 ]
机构
[1] Macquarie Univ, Fac Med & Hlth Sci, Surg Infect Res Grp, Sydney, NSW, Australia
[2] Univ Sao Paulo, Ribeirao Preto Nursing Sch, Ribeirao Preto, SP, Brazil
[3] Macquarie Univ, Sch Engn, Sydney, NSW, Australia
关键词
Endoscope; Surface damage; Roughness; Bacterial attachment; Biofilm; KLEBSIELLA-PNEUMONIAE; RETROGRADE CHOLANGIOPANCREATOGRAPHY; RESIDUAL CONTAMINATION; FLEXIBLE ENDOSCOPES; TRANSMISSION; DISINFECTANTS; CLEANABILITY; OUTBREAK; BIOFILM; PATIENT;
D O I
10.1016/j.jhin.2020.03.007
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Biofilm formation has been shown to be associated with damaged areas of endoscope channels. It was hypothesized that the passage of instruments and brushes through endoscope channels during procedures and cleaning contributes to channel damage, bacterial attachment and biofilm formation. Aim: To compare surface roughness and bacterial attachment in used and new endoscope channels in vivo and in vitro. Methods: Surface roughness of 10 clinically used (retired) and seven new colonoscope biopsy channels was analysed by a surface profiler. For the in-vitro study, a flexible endoscope biopsy forceps was passed repeatedly through a curved 3.0-mm-diameter Teflon tube 100, 200 and 500 times. Atomic force microscopy was used to determine the degree of inner surface damage. The number of Escherichia coli or Enterococcus faecium attached to the inner surface of the new Teflon tube and the tube with 500 forceps passes in 1 h at 37 degrees C was determined by culture. Results: The average surface roughness of the used biopsy channels was found to be 1.5 times greater than that of the new biopsy channels (P=0.03). Surface roughness of Teflon tubes with 100, 200 and 500 forceps passes was 1.05-, 1.12- and 3.2-fold (P=0.025) greater than the roughness of the new Teflon tubes, respectively. The number of E. coli and E. faecium attached to Teflon tubes with 500 forceps passes was 2.9-fold (P=0.021) and 4.3-fold (P=0.004) higher compared with the number of E. coli and E. faecium attached to the new Teflon tubes, respectively. Conclusion: An association was found between endoscope usage with damage to the biopsy channel and increased bacterial attachment. (C) 2020 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:176 / 182
页数:7
相关论文
共 33 条
  • [11] Designing for cleanability: The effects of material, surface roughness, and the presence of blood test soil and bacteria on devices
    Gonzalez, Elizabeth A.
    Nandy, Poulomi
    Lucas, Anne D.
    Hitchins, Victoria M.
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 2017, 45 (02) : 194 - 196
  • [12] Healthcare Infection Control Practices Advisory Committee, 2017, ESS EL REPR PROGR FL
  • [13] Persistent residual contamination in endoscope channels; a fluorescence epimicroscopy study
    Herve, Rodolphe C.
    Keevil, Charles W.
    [J]. ENDOSCOPY, 2016, 48 (07) : 609 - 616
  • [14] Duodenoscope-Associated Bacterial Infections: A Review and Update
    Jennifer T. Higa
    Michael Gluck
    Andrew S. Ross
    [J]. Current Treatment Options in Gastroenterology, 2016, 14 (2) : 185 - 193
  • [15] Infectious diseases linked to cross-contamination of flexible endoscopes
    Kenters, Nikki
    Huijskens, Elisabeth G. W.
    Meier, Corianne
    Voss, Andreas
    [J]. ENDOSCOPY INTERNATIONAL OPEN, 2015, 3 (04) : E259 - E265
  • [16] An outbreak of carbapenem-resistant OXA-48-producing Klebsiella pneumonia associated to duodenoscopy
    Kola, Axel
    Piening, Brar
    Pape, Ulrich-Frank
    Veltzke-Schlieker, Wilfried
    Kaase, Martin
    Geffers, Christine
    Wiedenmann, Bertram
    Gastmeier, Petra
    [J]. ANTIMICROBIAL RESISTANCE AND INFECTION CONTROL, 2015, 4
  • [17] Is bacteriologic surveillance in endoscope reprocessing stringent enough?
    Kovaleva, J.
    Meessen, N. E. L.
    Peters, F. T. M.
    Been, M. H.
    Arends, J. P.
    Borgers, R. P.
    Degener, J. E.
    [J]. ENDOSCOPY, 2009, 41 (10) : 913 - 916
  • [18] Evaluation of the ability of different detergents and disinfectants to remove and kill organisms in traditional biofilm
    Luciano, Cristiana da Costa
    Olson, Nancy
    Veiga Tipple, Anaclara Ferreira
    Alfa, Michelle
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 2016, 44 (11) : E243 - E249
  • [19] Genomic Epidemiology of an Endoscope-Associated Outbreak of Klebsiella pneumoniae Carbapenemase (KPC)-Producing K. pneumoniae
    Marsh, Jane W.
    Krauland, Mary G.
    Nelson, Jemma S.
    Schlackman, Jessica L.
    Brooks, Anthony M.
    Pasculle, A. William
    Shutt, Kathleen A.
    Doi, Yohei
    Querry, Ashley M.
    Muto, Carlene A.
    Harrison, Lee H.
    [J]. PLOS ONE, 2015, 10 (12):
  • [20] Endoscopy-associated transmission of carbapenem-resistant Klebsiella pneumoniae producing KPC-2 β-lactamase
    Naas, Thierry
    Cuzon, Gaelle
    Babics, Adrien
    Fortineau, Nicolas
    Boytchev, Isabelle
    Gayral, Francois
    Nordmann, Patrice
    [J]. JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2010, 65 (06) : 1305 - 1306