[1] Austin Hlth, Dept Radiol, Melbourne, Vic, Australia
来源:
AUSTRALASIAN RADIOLOGY
|
2007年
/
51卷
/
02期
关键词:
digital subtraction angiography;
interventional radiology;
superior vena cava abnormality;
Swan-Ganz catheterization;
adverse effects;
methods;
D O I:
10.1111/j.1440-1673.2007.01696.x
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Knotting of intravascular catheters is an uncommon but a well-recognized occurrence. The Swan-Ganz catheter (SGC) is the one that knots most commonly. A case of a knotted SGC is described in a patient with a persistent left-sided superior vena cava, and we propose that the presence of a left-sided superior vena cava is a risk factor for knot formation not previously reported. We review the published work on the risk factors for knot formation and on the techniques used to remove knotted SGC. We describe a technique using a gooseneck snare and Omni Flush catheter (Angiodynamics, Queensbury, NY, USA) to loosen and untie a knotted SGC.