Neutron Detector Sensitivity Measurements and Their Variation

被引:0
|
作者
McKinny, Kevin S. [1 ]
Anderson, Thomas R. [1 ]
Johnson, Nathan H. [1 ]
机构
[1] GE Energys Reuter Stokes Prod Line, Twinsburg, OH 44087 USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Historically Helium-3 based neutron detectors have been utilized for many applications including Radiation Portal Monitors and various Nuclear Safeguard applications. With the reduced amount of Helium-3 available alternative technologies are being evaluated and employed for use in these applications. In this study, the performance of Boron-10 lined proportional counters is evaluated experimentally in gamma-ray fields relevant to these applications. The method used to define and measure neutron sensitivity is evaluated, and effects of the ambient gamma field on the measured sensitivity are studied for a variety of detector configurations. Based on the results of this study, a repeatable method of defining and measuring detector sensitivity is reported, and the change in sensitivity that may be observed by systems integrators after incorporating the detectors into completed systems is assessed.
引用
收藏
页码:533 / 537
页数:5
相关论文
共 50 条
  • [1] Detector sensitivity and calibration for yield measurements in fusion neutron sources
    Andola, Sanjay
    Rawool, A. M.
    Kaushik, T. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 944
  • [2] Ultracold neutron detector for neutron lifetime measurements
    Andreev, V.
    Vassiljev, A.
    Ivanov, E.
    Ilyin, D.
    Krivshich, A.
    Serebrov, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 845 : 548 - 551
  • [3] Neutron sensitivity of helium scintillation fission neutron detector
    Zhang, Xiaodong
    Qiu, Mengtong
    Zhang, Jianfu
    Li, Hongyun
    Ouyang, Xiaoping
    Zhang, Xianpeng
    Zhang, Zhongbing
    Chen, Liang
    Liu, Linyue
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2012, 24 (05): : 1151 - 1154
  • [4] High sensitivity neutron detector for Z
    Ruggles, LE
    Porter, JL
    Simpson, WW
    Vargas, MF
    Zagar, DM
    Hartke, R
    Buersgens, F
    Symes, DR
    Ditmire, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3595 - 3597
  • [5] PHOTOGRAPHIC NEUTRON DETECTOR OF HIGH SENSITIVITY
    JOHNSON, TH
    PHYSICAL REVIEW, 1949, 75 (08): : 1301 - 1302
  • [6] DOSE SENSITIVITY OF A RHODIUM NEUTRON DETECTOR
    OBATUROV, GM
    CHUMBARO.YK
    SOVIET ATOMIC ENERGY-USSR, 1971, 30 (04): : 474 - &
  • [7] MEASUREMENTS AND CALCULATIONS OF NEUTRON DETECTOR EFFICIENCIES
    THORNTON, ST
    SMITH, JR
    NUCLEAR INSTRUMENTS & METHODS, 1971, 96 (04): : 551 - &
  • [8] NEUTRON RESONANCE MEASUREMENTS WITH A GERMANIUM DETECTOR
    BIRD, JR
    KENNY, MJ
    ALLEN, BJ
    PHYSICS LETTERS B, 1968, B 27 (10) : 638 - &
  • [9] EFFICIENCY MEASUREMENTS OF A LARGE NEUTRON DETECTOR
    GOULDING, CA
    GREENFIELD, MB
    BAINUM, DE
    RAPAPORT, J
    FOSTER, CC
    WARD, TE
    ZAFIRATOS, CD
    SCHERY, SD
    GOODMAN, CD
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1978, 23 (04): : 558 - 558
  • [10] NEUTRON SENSITIVITY MEASUREMENTS OF LR115 TRACK DETECTOR WITH SOME (N,ALPHA) CONVERTERS
    PALFALVI, J