Repair of Flexor Digitorum Profundus to Distal Phalanx: A Biomechanical Evaluation of Four Techniques

被引:34
作者
Lee, Steve K. [1 ]
Fajardo, Marc [1 ]
Kardashian, George [1 ]
Klein, Jason [1 ]
Tsai, Peter [1 ]
Christoforou, Dimitrios [1 ]
机构
[1] Hosp Special Surg, Hand & Upper Extrem Surg Serv, New York, NY 10021 USA
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2011年 / 36A卷 / 10期
关键词
Anchor; button; jersey finger; repair; tendon; SUTURE ANCHORS; TENDON REPAIR; PULLOUT BUTTON; INJURIES; AVULSION;
D O I
10.1016/j.jhsa.2011.07.017
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Many techniques for repair of the flexor digitorum profundus to the distal phalanx show excessive gapping with variable clinical results. The purpose of this study was to test the biomechanical characteristics of an anchor-button (AB) technique, as compared to 3 other techniques. Methods Twenty-four fresh-frozen human cadaveric fingers were randomized to 4 groups, 6 in each: group 1, 2-strand Bunnell suture button pullout technique; group 2, modified Kessler suture and 2 retrograde anchors; group 3: locking Krakow suture with 2 retrograde anchors; group 4, AB technique incorporating a 2-part repair, consisting of a locking dorsal Krakow suture with 2 retrograde anchors and a locking palmar Krakow suture fixed with a button. Tendon-to-bone gapping was measured after cyclical loading. Ultimate load to failure was measured at the end of 500 cycles. Results The AB technique resulted in significantly less gapping when compared to the other techniques. It also resulted in a significantly stronger repair compared to all the other groups with an average load to failure comparable to the native tendon-to-bone interface. Conclusions The AB repair might allow for early active postoperative motion after repair of flexor digitorum profundus avulsion injuries and tendon reconstruction procedures; however, the soft tissue effects of this multistrand technique are unknown in clinical repairs. (J Hand Surg 2011;36A:1604-1609. Copyright (C) 2011 by the American Society for Surgery of the Hand. All rights reserved.)
引用
收藏
页码:1604 / 1609
页数:6
相关论文
共 17 条
[1]   Bone suture anchors versus the pullout button for repair of distal profundus tendon injuries: A comparison of strength in human cadaveric hands [J].
Brustein, M ;
Pellegrini, J ;
Choueka, J ;
Heminger, H ;
Mass, D .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2001, 26A (03) :489-496
[2]   THE DEADMAN THEORY OF SUTURE ANCHORS - OBSERVATIONS ALONG A SOUTH TEXAS FENCE LINE [J].
BURKHART, SS .
ARTHROSCOPY, 1995, 11 (01) :119-123
[3]  
Evans R B, 1993, J Hand Ther, V6, P266
[4]   COMPLICATIONS EXPERIENCED IN THE REHABILITATION OF ZONE-I FLEXOR TENDON INJURIES WITH DYNAMIC TRACTION SPLINTING [J].
GERBINO, PG ;
SALDANA, MJ ;
WESTERBECK, P ;
SCHACHERER, TG .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1991, 16A (04) :680-686
[5]   PRIMARY REPAIR WITHOUT IMMOBILIZATION OF FLEXOR TENDON DIVISION WITHIN DIGITAL SHEATH - AN EXPERIMENTAL AND CLINICAL STUDY [J].
KESSLER, I ;
NISSIM, F .
ACTA ORTHOPAEDICA SCANDINAVICA, 1969, 40 (05) :587-+
[6]   Cyclic testing of pullout sutures and micro-Mitek suture anchors in flexor digitorum profundus tendon distal fixation [J].
Latendresse, K ;
Dona, E ;
Scougall, PJ ;
Schreuder, FB ;
Puchert, E ;
Walsh, WR .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (03) :471-478
[7]  
Leddy J P, 1977, J Hand Surg Am, V2, P66
[8]   AVULSION OF RING FINGER FLEXOR DIGITORUM PROFUNDUS TENDON - EXPERIMENTAL-STUDY [J].
MANSKE, PR ;
LESKER, PA .
HAND, 1978, 10 (01) :52-55
[9]   Comparison of pullout button versus suture anchor for zone I flexor tendon repair [J].
McCallister, WV ;
Ambrose, HC ;
Katolik, LI ;
Trumble, TE .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2006, 31A (02) :246-251
[10]   Primary flexor tendon repair in zone 1 [J].
Moiemen, NS ;
Elliot, D .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2000, 25B (01) :78-84