Liability of laryngeal mask airway devices to thermal damage from KTP and Nd:YAG lasers

被引:7
作者
Keller, C
Brimacombe, J [1 ]
Coorey, A
Wood, V
Keller, K
机构
[1] Univ Queensland, Cairns Base Hosp, Dept Anaesthesia & Intens Care Med, Cairns 4870, Australia
[2] Leopold Franzens Univ, Dept Anaesthesia & Intens Care Med, A-6020 Innsbruck, Austria
[3] Univ Queensland, Cairns Day Surg, Cairns 4870, Australia
关键词
equipment; lasers; masks anaesthesia; surgery; otolaryngological;
D O I
10.1093/bja/82.2.291
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have compared the liability of four laryngeal mask airway (LMA) devices (standard, flexible, intubating and reusable) and a tracheal tube to thermal damage from KTP and Nd:YAG lasers at two power densities used commonly in airway surgery: 570 W cm(-2) and 1140 W cm(-2) Eighty-five airway devices were tested: 24 standard LMA (silicone-based), 12 flexible LMA (silicone-based, metal wires), 24 disposable LMA (PVC-based), one intubating LMA (silicone and steel-based) and 24 PVC-based tracheal tubes. Comparisons were made during laser strike to eight different targets: the unmarked and marked part of the airway device tube; the unmarked part of the airway device tube after application of blood; the cuff filled with air or methylene blue dye; the unmarked flexible LMA tube on or between the metal wires; and the epiglottic elevator bar of the intubating LMA. The laser strike was continued for 30 s and each target was tested three times. Three different, but identical, impact sites were used for each target. There was no ignition of any airway device with either power density or laser type. The silicone-based LMA were generally more resistant to flaring and penetration than the PVC-based LMA and tracheal tube, but the intubating LMA tube flared more rapidly with the KTP laser, and the disposable LMA cuff was more resistant to penetration. Print markings, blood and the metal wires of the flexible LMA reduced the thermal resistance of the tube. Filling the cuff with methylene blue dye increased the thermal resistance of all airway devices. We conclude that the silicone-based LMA devices were more thermal resistant to KTP and Nd:YAG laser strike than PVC-based devices with the exception of the disposable LMA cuff and the intubating LMA tube.
引用
收藏
页码:291 / 294
页数:4
相关论文
共 15 条
[1]   Damage to the laryngeal mask by residual fluid in the cuff [J].
Asai, T ;
Koga, K ;
Morris, S .
ANAESTHESIA, 1997, 52 (10) :977-981
[2]   The intubating laryngeal mask .1. Development of a new device for intubation of the trachea [J].
Brain, AIJ ;
Verghese, C ;
Addy, EV ;
Kapila, A .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 79 (06) :699-703
[3]   The intubating laryngeal mask .2. a preliminary clinical report of a new means of intubating the trachea [J].
Brain, AIJ ;
Verghese, C ;
Addy, EV ;
Kapila, A ;
Brimacombe, J .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 79 (06) :704-709
[4]   THE INCENDIARY CHARACTERISTICS OF THE LARYNGEAL AND REINFORCED LARYNGEAL MASK AIRWAY TO CO2-LASER STRIKE - A COMPARISON WITH 2 POLYVINYL-CHLORIDE TRACHEAL TUBES [J].
BRIMACOMBE, J .
ANAESTHESIA AND INTENSIVE CARE, 1994, 22 (06) :694-697
[5]   Saline as an alternative to air for filling the laryngeal mask airway cuff [J].
Coorey, A ;
Brimacombe, J ;
Keller, C .
BRITISH JOURNAL OF ANAESTHESIA, 1998, 81 (03) :398-400
[6]   ANESTHETIC MANAGEMENT OF TRACHEAL SURGERY USING THE LARYNGEAL MASK AIRWAY [J].
DIVATIA, JV ;
SAREEN, R ;
UPADHYE, SM ;
SHARMA, KS ;
SHELGAONKAR, JR .
ANAESTHESIA AND INTENSIVE CARE, 1994, 22 (01) :69-73
[7]  
FISHER JC, 1987, ENDOSCOPIC LASER SUR
[8]   Pulsed laser treatment in children and the use of anesthesia [J].
Grevelink, JM ;
White, VR ;
Bonoan, R ;
Denman, WT .
JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 1997, 37 (01) :75-81
[9]   INCENDIARY CHARACTERISTICS OF A NEW LASER-RESISTANT ENDOTRACHEAL-TUBE [J].
HAYES, DM ;
GABA, DM ;
GOODE, RL .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1986, 95 (01) :37-40
[10]   KTP laser-resistant properties of the reinforced laryngeal mask airway [J].
Pandit, JJ ;
Chambers, P ;
OMalley, S .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 78 (05) :594-600