Radiance transfer standard detector (RTSD) is a trap-based facility used as transfer standard detector for radiance measurements. In this paper, a RTSD with band-center wavelength of 5.50nm is calibrated and the result of absolute spectral radiance responsivity (ASRR) calibration is presented. The relative uncertainty in the absolute spectral radiance responsivity calibration of the RTSD is less than 8 x 10(-3).