Influence of handle design on the surgeon's upper limb movements, muscle recruitment, and fatigue during endoscopic suturing

被引:60
作者
Emam, TA
Frank, TG
Hanna, GB
Cuschieri, A [1 ]
机构
[1] Univ Dundee, Ninewells Hosp & Med Sch, Dept Surg, Dundee DD1 9SY, Scotland
[2] Univ Dundee, Ninewells Hosp & Med Sch, Mol Oncol & Surg Skills Unit, Dundee DD1 9SY, Scotland
来源
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES | 2001年 / 15卷 / 07期
关键词
ball handle; rocker handle; needle drivers; motion analysis; telemetric electromyography; ergonomics; endoscopic suturing;
D O I
10.1007/s004640080141
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Thus far, little has been done to investigate the kinematics (motion analysis) and kinetics (muscle work, muscle fatigue, comfort) of surgeons during laparoscopic surgery. Therefore, we set out to study these ergonomic aspects of task performance in the dominant upper limb of surgeons during endoscopic suturing. Methods: Three different handles-conventional finger loop, rocker, and ball handle prototype-were compared in a study involving 10 surgeons suturing porcine enterotomies with each of the three instruments. The endpoints were performance parameters, motion analysis and muscle work, and fatigue of the surgeon's dominant upper limb; subjective scores for comfort level and maneuverability were also elicited from the subjects. Results: Task quality and efficiency during endoscopic suturing, were significantly better with the ball and rocker handle needle drivers than with the finger loop instrument, with lower angular velocity at the elbow and shoulder joints, more pronation, and less supination. The integrated muscle work was much lower for both the rocker and the ball handles. Significant muscle fatigue, especially of the arm flexors and deltoid, was observed only with finger loop instruments. Comfort and maneuverability rating scores were higher with both handles than with the conventional finger loop. The ball handle was easier to maneuver, but it was somewhat less comfortable than the rocker system. Conclusion: A different pattern of joint movements, a reduction in muscle power exerted during endoscopic suturing, and hence an absence of muscle fatigue were documented with ergonomic needle drivers (rocker and ball) when compared to the conventional finger loop instruments. These differences translate to better and more efficient task performance with enhanced comfort.
引用
收藏
页码:667 / 672
页数:6
相关论文
共 20 条
  • [1] [Anonymous], CURR ORTHOPAED
  • [2] DE LUCA CJ, 1984, CRIT REV BIOMED ENG, V11, P251
  • [3] Rocker handle for endoscopic needle drivers - Technical and ergonomic evaluation by infrared motion analysis system
    Emam, TA
    Frank, TG
    Hanna, GB
    Stockham, G
    Cuschieri, A
    [J]. SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1999, 13 (07): : 658 - 661
  • [4] Effect of intracorporeal-extracorporeal instrument length ratio on endoscopic task performance and surgeon movements
    Emam, TA
    Hanna, GB
    Kimber, C
    Dunkley, P
    Cuschieri, A
    [J]. ARCHIVES OF SURGERY, 2000, 135 (01) : 62 - 65
  • [5] HALL SJ, 1995, BASIC BIOMECHANICS, P294
  • [6] HAMILL J, 1995, BIOMECHANICAL BASIS, P325
  • [7] Optimal port locations for endoscopic intracorporeal knotting
    Hanna, GB
    Shimi, S
    Cuschieri, A
    [J]. SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1997, 11 (04): : 397 - 401
  • [8] Neuropraxia following laparoscopic procedures: An occupational injury
    Horgan, LF
    ORiordan, DC
    Doctor, N
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 1997, 6 (01): : 33 - 35
  • [9] JONSSON B, 1978, SCAND J REHABIL MED, P69
  • [10] KADEFORS R, 1978, SCAND J REHABIL MED, P75