The validity of adenosine triphosphate measurement in detecting endoscope contamination

被引:9
|
作者
McCafferty, C. E. [1 ,2 ]
Abi-Hanna, D. [3 ,4 ]
Aghajani, M. J. [1 ,2 ]
Micali, G. T. [1 ,2 ]
Lockart, I [3 ]
Vickery, K. [5 ]
Gosbell, I. B. [1 ,2 ]
Jensen, S. O. [1 ,2 ]
机构
[1] Ingham Inst Appl Med Res, Sydney, NSW, Australia
[2] Western Sydney Univ, Sch Med, Sydney, NSW, Australia
[3] Liverpool Hosp, Dept Gastroenterol & Hepatol, Sydney, NSW, Australia
[4] Univ New South Wales, Sch Med, Sydney, NSW, Australia
[5] Macquarie Univ, Australian Sch Adv Med, Sydney, NSW, Australia
关键词
Gastrointestinal endoscopy; Adenosine triphosphate; Endoscope reprocessing; GASTROINTESTINAL ENDOSCOPES; KLEBSIELLA-PNEUMONIAE; DUODENOSCOPES; BIOLUMINESCENCE; DECONTAMINATION; DISINFECTION; TRANSMISSION; UPDATE; RISK;
D O I
10.1016/j.jhin.2018.08.004
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Endoscopic procedures are vital to gastrointestinal disease diagnosis and management, but risk infection transmission. In Australia, endoscopes undergo monthlyto-quarterly microbiological testing, to prevent patient infection. Endoscopes are used more frequently, meaning contamination may not be detected by this surveillance before infection transmission occurs. Aim: To evaluate the use of adenosine triphosphate (ATP) measurement, alongside standard microbiological cultures, in detecting endoscope contamination before high-level disinfection. Using these results, we also aimed to confirm the efficacy of manual cleaning in reducing levels of ATP and cfu/mL. Methods: Seventeen in-clinical-use gastroscopes and 24 in-clinical-use colonoscopes from the Liverpool Hospital Endoscopy unit were sampled across three separate cleaning stages before high-level disinfection. Colony counts and ATP measurements were then performed on these samples. Findings: The correlation between the cfu/mL and RLU of samples collected from colonoscopes was 0.497 (95% confidence interval: 0.28-0.66; P < 0.0001). The correlation between cfu/mL and RLU for samples collected from gastroscopes was 0.377 (0.08-0.61; P = 0.0138). RLU and cfu/mL values were shown to fall significantly (P < 0.005) following precleaning and manual cleaning. Conclusion: There was a significant correlation between ATP and cfu/mL measured from samples collected before high-level disinfection. Precleaning and manual cleaning were shown to reduce ATP and microbiological load significantly. ATP measurement can be performed within minutes with little training and produces results that are easy to interpret. These findings warrant further research on the utility of ATP measurement as a screening tool for detecting endoscope contamination after high-level disinfection. (C) 2018 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E142 / E145
页数:4
相关论文
共 50 条
  • [1] Detecting adenosine triphosphate in the pericellular space
    Falzoni, Simonetta
    Donvito, Giovanna
    Di Virgilio, Francesco
    INTERFACE FOCUS, 2013, 3 (03)
  • [2] Use of adenosine triphosphate to audit reprocessing of flexible endoscopes with an elevator mechanism
    Quan, Erik
    Mahmood, Rizwan
    Naik, Amar
    Sargon, Peter
    Shastri, Nikhil
    Venu, Mukund
    Parada, Jorge P.
    Gupta, Neil
    AMERICAN JOURNAL OF INFECTION CONTROL, 2018, 46 (11) : 1272 - 1277
  • [3] Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels
    Alfa, Michelle J.
    Fatima, Iram
    Olson, Nancy
    AMERICAN JOURNAL OF INFECTION CONTROL, 2013, 41 (03) : 245 - 248
  • [4] Adenosine Triphosphate Quantification Correlates Poorly with Microbial Contamination of Duodenoscopes
    Olafsdottir, Lovisa B.
    Wright, Sharon B.
    Smithey, Anne
    Heroux, Riley
    Hirsch, Elizabeth B.
    Chen, Alice
    Lane, Benjamin
    Sawhney, Mandeep S.
    Snyder, Graham M.
    INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2017, 38 (06) : 678 - 684
  • [5] Highly-Sensitive Detecting of Adenosine Triphosphate Based on Bioluminescence
    He Bao-Shan
    Yue Wei-Wei
    Luo Jin-Ping
    Zhou Ai-Yu
    Wang Mi-Xia
    Cai Xin-Xia
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2008, 36 (10) : 1427 - 1430
  • [6] High-quality endoscope reprocessing decreases endoscope contamination
    Decristoforo, P.
    Kaltseis, J.
    Fritz, A.
    Edlinger, M.
    Posch, W.
    Wilflingseder, D.
    Lass-Floerl, C.
    Orth-Hoeller, D.
    CLINICAL MICROBIOLOGY AND INFECTION, 2018, 24 (10) : 1101.e1 - 1101.e6
  • [7] Persistent residual contamination in endoscope channels; a fluorescence epimicroscopy study
    Herve, Rodolphe C.
    Keevil, Charles W.
    ENDOSCOPY, 2016, 48 (07) : 609 - 616
  • [8] MEASUREMENT OF ADENOSINE-TRIPHOSPHATE IN THE BRAIN OF MICE EXPOSED TO NORMOBARIC HYPEROXYGENATION
    ITO, T
    KATAYAMA, Y
    EDAMATSU, R
    YUFU, K
    MORI, A
    MEDICAL SCIENCE RESEARCH, 1995, 23 (06): : 373 - 374
  • [9] Can adenosine triphosphate be a proxy measure in evaluation of hand disinfection effect?
    Breidablik, H. J.
    Lysebo, D. E.
    Johannesen, L.
    Skare, A.
    Andersen, J. R.
    Kleiven, O. T.
    JOURNAL OF HOSPITAL INFECTION, 2020, 105 (03) : 558 - 560
  • [10] A systematic review of adenosine triphosphate as a surrogate for bacterial contamination of duodenoscopes used for endoscopic retrograde cholangiopancreatography
    Olafsdottir, Lovisa B.
    Whelan, Julia
    Snyder, Graham M.
    AMERICAN JOURNAL OF INFECTION CONTROL, 2018, 46 (06) : 697 - 705