Traceable and accurate quantification of iron in seawater using isotope dilution calibration strategies by triple quadrupole ICP-MS/MS: Characterization measurements of iron in a candidate seawater CRM

被引:15
作者
An, Betul [1 ,2 ]
Can, Suleyman Z. [1 ]
Bakirdere, Sezgin [2 ]
机构
[1] TUBITAK Natl Metrol Inst, TR-41470 Kocaeli, Turkey
[2] Yildiz Tech Univ, Fac Art & Sci, Chem Dept, TR-34210 Istanbul, Turkey
关键词
Isotope dilution mass spectrometry; Single dilution; Double dilution; Triple dilution; Seawater; Iron; Triethylamine; Mg(OH)(2) co-precipitation; PLASMA-MASS SPECTROMETRY; REDUCED GRAPHENE OXIDE; DETECTOR DEAD-TIME; VOLTAMMETRIC DETERMINATION; CHELATING RESIN; DISSOLVED IRON; METALS; FE; PRECONCENTRATION; ELEMENTS;
D O I
10.1016/j.talanta.2019.120503
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Accurate quantification of iron is crucial for the characterization of a candidate seawater certified reference material. Isotope dilution mass spectrometry as a primary method is one of the option for this purpose. In this study, triethylamine (TEA) assisted Mg(OH)(2) co-precipitation is developed and combined with isotope dilution mass spectrometry for the characterization of candidate certified reference material, UME CRM 1206. The analytical protocol of co-precipitation was optimized to achieve reproducible yields of similar to 100% Fe with high salt rejection (residual salinity similar to 0.003%). Three calibration strategies, single, double and triple isotope dilution mass spectrometry (IDMS) coupled with co-precipitation were investigated in terms of metrological aspects. The determination of Fe was carried out by ICP-MS with tandem configuration (ICP-MS/MS) using NH3/He cell gas which was optimized for the minimum background equivalence concentration to be able to measure highly accurate Fe-56/Fe-57 ratio with the highest possible sensitivity. The background corrected typical procedural blank levels and limit of detection calculated based on three times the standard deviation of blanks were found as 9 ng/kg (0.16 nM) and 11 ng/kg (0.20 nM) for single IDMS, respectively. Trueness of the developed analytical methods has been demonstrated using CASS-6 and NMIA MX014 with complete and very precise recoveries. SI-traceable values with combined uncertainty (k = 2) were obtained for single, double and triple IDMS as (12.70 +/- 0.16) ng/g, (12.728 +/- 0.084) ng/g and (12.732 +/- 0.062) ng/g, respectively. The relative combined uncertainty for each approach was calculated as 0.64%, 0.33% and 0.24%, respectively.
引用
收藏
页数:10
相关论文
共 59 条
[1]   DETERMINATION OF IRON AND 10 OTHER TRACE-ELEMENTS IN THE OPEN OCEAN SEAWATER REFERENCE MATERIAL NASS-3 BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
AKATSUKA, K ;
MCLAREN, JW ;
LAM, JW ;
BERMAN, SS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :889-894
[2]  
[Anonymous], 2010, APPL NOTE 1
[3]  
[Anonymous], [No title captured]
[4]  
[Anonymous], 2008, JCGM, V100, DOI DOI 10.1373/CLINCHCM.2003.030528
[5]  
[Anonymous], 2012, EURACHEM CITAC QUANT
[6]   Determination of ultratrace levels of dissolved metals in seawater by reaction cell inductively coupled plasma mass spectrometry after ammonia induced magnesium hydroxide coprecipitation [J].
Ardini, Francisco ;
Magi, Emanuele ;
Grotti, Marco .
ANALYTICA CHIMICA ACTA, 2011, 706 (01) :84-88
[7]   Triethylamine-assisted Mg(OH)2 coprecipitation/preconcentration for determination of trace metals and rare earth elements in seawater by inductively coupled plasma mass spectrometry (ICP-MS) [J].
Arslan, Zikri ;
Oymak, Tulay ;
White, Jeremy .
ANALYTICA CHIMICA ACTA, 2018, 1008 :18-28
[8]   Widespread iron limitation of phytoplankton in the South Pacific Ocean [J].
Behrenfeld, MJ ;
Kolber, ZS .
SCIENCE, 1999, 283 (5403) :840-843
[9]   Analysis of Mn, Fe, Co, Ni, Cu, Zn, Cd, and Pb in seawater using the Nobias-chelate PA1 resin and magnetic sector inductively coupled plasma mass spectrometry (ICP-MS) [J].
Biller, Dondra V. ;
Bruland, Kenneth W. .
MARINE CHEMISTRY, 2012, 130 :12-20
[10]   IRON(II) AND IRON(III) DETERMINATION IN SEA-WATER AT THE NANOMOLAR LEVEL WITH SELECTIVE ONLINE PRECONCENTRATION AND SPECTROPHOTOMETRIC DETERMINATION [J].
BLAIN, S ;
TREGUER, P .
ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) :425-432