Study of precursors of ventricular tachycardia from data stored in the memory of a dual chamber implantable cardioverter defibrillator

被引:9
|
作者
Leenhardt, A
Sadoul, N
Mabo, P
Kacet, S
Lavergne, T
Saoudi, N
Iscolo, N
机构
[1] Hop Lariboisiere, Serv Cardiol, F-75475 Paris 10, France
[2] Brabois Hosp, Vandoeuvre Les Nancy, France
[3] Cardiol Hosp, Rennes, France
[4] Cardiol Hosp, Lille, France
[5] European Georges Pompidou Hosp, Paris, France
[6] Princess Grace Hosp, Monaco, Monaco
[7] ELA Med, Le Plessis Robinson, France
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 07期
关键词
ventricular tachycardia; implantable cardioverter defibrillator; arrhythmias onset;
D O I
10.1046/j.1460-9592.2003.t01-1-00210.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study was performed to examine precursors of ventricular tachyarrhythmias in patients who experienced a sustained ventricular tachyarrhythmia and received appropriate therapy by ICD. From an overall consecutive population of 77 patients, 18 patients (1 woman, mean age 61.7 +/- 10.8 years) were selected for having experienced a sustained ventricular tochyarrhythmia and received at least one appropriate ICD therapy preceded by 20 minutes of internal information. The number of premature ventricular complexes (PVCs)/min for each of the 20 minutes preceding the onset of ventricular tachyarrhythmia, the shortest coupling intervals between PVC and normal sinus beat, and the presence of short-long-short (SLS) interval sequences were examined. Data were stratified according to underlying disease, left ventricular ejection fraction, rate of ventricular tachyarrhythmia, and antiorrhythmic therapy. One hundred twenty-eight episodes of spontaneous ventricular tachyarrhythmia were retrieved. Rapid ventricular tachyarrhythmia (> 160 beats/min) were preceded by a significantly greater mean number (3.71 +/- 6.36) of PVCs than slower ventricular tachyarrhythmia (less than or equal to160 beats/min) (0.63 +/- 0.88, P = 0.0004). The mean shortest PVC coupling interval was significantly shorter in patients with (588 +/- 99ms) versus without (643 +/- 111 ms, P = 0.03)ischemic heart disease, before episodes of rapid (527 55 ms) versus slower (636 105 ins, P = 0.0001) ventricular tachyarrhythmia, and in the absence (538 +/- 80 ms) versus the presence (620 +/- 105 ms, P = 0.006) of amiodarone. SLS sequences preceded 29% of rapid ventricular tachyorrhythmic episodes, versus 8% of the slower ventricular tachyarrhythmia (P < 0.01). Significant differences were found in the characteristics of PVCs preceding ventricular tachyarrhythmic episodes in accordance to their rate and the underlying cardiomyopathy. Though insufficient in isolation, these findings may be helpful when combined with other observations to develop preventive algorithms, or to refine the programming of implantable devices.
引用
收藏
页码:1454 / 1460
页数:7
相关论文
共 50 条
  • [1] Safety and efficacy of a dual chamber implantable cardioverter defibrillator capable of slow ventricular tachycardia discrimination:: A randomized study
    Mletzko, R
    Aliot, E
    Seidl, K
    Lavergne, T
    Mabo, P
    Hohnloser, S
    Böcker, D
    Schoels, W
    Sadoul, N
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (12): : 2275 - 2282
  • [2] A new classification algorithm for discrimination of ventricular from supraventricular tachycardia in a dual chamber implantable cardioverter defibrillator
    Korte, T
    Jung, W
    Wolpert, C
    Spehl, S
    Schumacher, B
    Esmailzadeh, B
    Lüderitz, B
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (01) : 70 - 73
  • [3] Initiation of Ventricular Tachycardia by Interruption of Pacemaker-Mediated Tachycardia in a Patient with a Dual-Chamber Implantable Cardioverter Defibrillator
    Van Gelder, Berry M.
    Meijer, Albert
    Dekker, Lukas R.
    Bracke, Frank A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (09): : 1227 - 1230
  • [4] Delayed detection of ventricular tachycardia in a dual chamber rate adaptive pacing implantable cardioverter defibrillator: A case of intradevice interaction
    Shalaby, AA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (08): : 1164 - 1166
  • [5] Failure of an Implantable Cardioverter Defibrillator to Terminate Ventricular Tachycardia: Why?
    Conti, Letizia
    Tran, Nam
    Burri, Haran
    Zimmermann, Marc
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (02): : 228 - 230
  • [6] Use of atrial and ventricular electrograms from a dual chamber implantable cardioverter defibrillator to elucidate a complex dysrhythmia
    Greenberg, RM
    Degeratu, FT
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10): : 2002 - 2004
  • [7] Evaluation of a dual chamber implantable cardioverter defibrillator for the treatment of atrial and ventricular arrhythmias
    Boriani, G
    Wollmann, C
    Biffi, M
    Kühl, M
    Schuchert, A
    Sperzel, J
    Stiller, S
    Gasparini, G
    Böcker, D
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 461 - 465
  • [8] Sexual Activity as a Trigger for Ventricular Tachycardia in a Patient with Implantable Cardioverter Defibrillator
    Arshad M. Safi
    Maurice Rachko
    Dima Yeshou
    Richard A. Stein
    Archives of Sexual Behavior, 2002, 31 : 295 - 299
  • [9] Sexual activity as a trigger for ventricular tachycardia in a patient with implantable cardioverter defibrillator
    Safi, AM
    Rachko, M
    Yeshou, D
    Stein, RA
    ARCHIVES OF SEXUAL BEHAVIOR, 2002, 31 (03) : 295 - 299
  • [10] Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection
    Shivkumar, K
    Feliciano, Z
    Boyle, NG
    Wiener, I
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (11) : 1285 - 1288