Analysis of plasticizers in poly(vinyl chloride) medical devices for infusion and artificial nutrition: comparison and optimization of the extraction procedures, a pre-migration test step

被引:28
作者
Bernard, Lise [1 ,2 ]
Cueff, Regis [2 ]
Bourdeaux, Daniel [1 ,2 ]
Breysse, Colette [3 ]
Sautou, Valerie [1 ,2 ]
机构
[1] CHU Clermont Ferrand, Pole Pharm, F-63003 Clermont Ferrand, France
[2] Clermont Univ, Univ Auvergne, C BIOSENSS, EA 4676, F-63000 Clermont Ferrand, France
[3] Ctr Ressources Technol CASIMIR, F-63173 Aubiere, France
关键词
Biomaterials; GC; Extraction (SFE vertical bar SPE vertical bar SPME); IR spectroscopy; Polymers; CHROMATOGRAPHY; SEPARATION; PHTHALATE;
D O I
10.1007/s00216-014-8426-z
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Medical devices (MDs) for infusion and enteral and parenteral nutrition are essentially made of plasticized polyvinyl chloride (PVC). The first step in assessing patient exposure to these plasticizers, as well as ensuring that the MDs are free from di(2-ethylhexyl) phthalate (DEHP), consists of identifying and quantifying the plasticizers present and, consequently, determining which ones are likely to migrate into the patient's body. We compared three different extraction methods using 0.1 g of plasticized PVC: Soxhlet extraction in diethyl ether and ethyl acetate, polymer dissolution, and room temperature extraction in different solvents. It was found that simple room temperature chloroform extraction under optimized conditions (30 min, 50 mL) gave the best separation of plasticizers from the PVC matrix, with extraction yields ranging from 92 to 100 % for all plasticizers. This result was confirmed by supplemented Fourier transform infrared spectroscopy-attenuated total reflection (FTIR-ATR) and gravimetric analyses. The technique was used on eight marketed medical devices and showed that they contained different amounts of plasticizers, ranging from 25 to 36 % of the PVC weight. These yields, associated with the individual physicochemical properties of each plasticizer, highlight the need for further migration studies.
引用
收藏
页码:1651 / 1659
页数:9
相关论文
共 20 条
[1]   Test Method for 6 Phthalates in Polyvinyl Chloride [J].
Abe, Yutaka ;
Mutsuga, Motoh ;
Hirahara, Yoshichika ;
Kawamura, Yoko .
FOOD HYGIENE AND SAFETY SCIENCE, 2011, 52 (05) :309-313
[2]  
[Anonymous], 2006, PLASTICS TECHNOLOGY
[3]  
[Anonymous], RESULTS PROFICIENCY
[4]  
[Anonymous], 2010, REV EXPOSURE TOXICIT
[5]  
[Anonymous], PVC HEALTHC APPL
[6]   Analytical methods for the determination of DEHP plasticizer alternatives present in medical devices: A review [J].
Bernard, L. ;
Decaudin, B. ;
Lecoeur, M. ;
Richard, D. ;
Bourdeaux, D. ;
Cueff, R. ;
Sautou, V. .
TALANTA, 2014, 129 :39-54
[7]   Determination of adipate plasticizers in poly(vinyl chloride) by microwave-assisted extraction [J].
Cano, JM ;
Marín, ML ;
Sánchez, A ;
Hernandis, V .
JOURNAL OF CHROMATOGRAPHY A, 2002, 963 (1-2) :401-409
[8]  
Danish Environmental Protection Agency, 2014, ALT CLASS PHTHAL MED
[9]   Liquid phase microextraction-ion exchange-high performance thin layer chromatography for the preconcentration, separation, and determination of plasticizers in aqueous samples [J].
Faraji, Hakim ;
Mirzaie, Afshin ;
Waqif-Husain, Syed .
JOURNAL OF SEPARATION SCIENCE, 2013, 36 (08) :1486-1492
[10]   Experimental study on infusion devices containing polyvinyl chloride: To what extent are they di(2-ethylhexyl)phthalate-free? [J].
Genay, Stephanie ;
Luciani, Chiara ;
Decaudin, Bertrand ;
Kambia, Nicolas ;
Dine, Thierry ;
Azaroual, Nathalie ;
Di Martino, Piera ;
Barthelemy, Christine ;
Odou, Pascal .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2011, 412 (1-2) :47-51