Assessing Image Quality of Low-Cost Laparoscopic Box Trainers: Options for Residents Training at Home

被引:9
作者
Kiely, Daniel J. [1 ]
Stephanson, Kirk [1 ]
Ross, Sue [1 ]
机构
[1] Univ Calgary, Res Program, Dept Obstet & Gynaecol, Foothills Med Ctr, Calgary, AB T2N 4J8, Canada
来源
SIMULATION IN HEALTHCARE-JOURNAL OF THE SOCIETY FOR SIMULATION IN HEALTHCARE | 2011年 / 6卷 / 05期
关键词
Laparoscopy; Computer simulation; Internship and residency; Graduate medical education; Minimally invasive surgical procedures; SKILLS; PERFORMANCE; SIMULATION; BUILD;
D O I
10.1097/SIH.0b013e31821cdb68
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Introduction: Low-cost laparoscopic box trainers built using home computers and webcams may provide residents with a useful tool for practice at home. This study set out to evaluate the image quality of low-cost laparoscopic box trainers compared with a commercially available model. Methods: Five low-cost laparoscopic box trainers including the components listed were compared in random order to one commercially available box trainer: A (high-definition USB 2.0 webcam, PC laptop), B (Firewire webcam, Mac laptop), C (high-definition USB 2.0 webcam, Mac laptop), D (standard USB webcam, PC desktop), E (Firewire webcam, PC desktop), and F (the TRLCD03 3-DMEd Standard Minimally Invasive Training System). Participants observed still image quality and performed a peg transfer task using each box trainer. Participants rated still image quality, image quality with motion, and whether the box trainer had sufficient image quality to be useful for training. Results: Sixteen residents in obstetrics and gynecology took part in the study. The box trainers showing no statistically significant difference from the commercially available model were A, B, C, D, and E for still image quality; A for image quality with motion; and A and B for usefulness of the simulator based on image quality. The cost of the box trainers A-E is approximately $ 100 to $ 160 each, not including a computer or laparoscopic instruments. Conclusions: Laparoscopic box trainers built from a high-definition USB 2.0 webcam with a PC (box trainer A) or from a Firewire webcam with a Mac (box trainer B) provide image quality comparable with a commercial standard. (Sim Healthcare 6: 292-298, 2011)
引用
收藏
页码:292 / 298
页数:7
相关论文
共 28 条
  • [1] A Novel Home Laparoscopic Simulator
    Al-Abed, Yahya
    Cooper, David G.
    [J]. JOURNAL OF SURGICAL EDUCATION, 2009, 66 (01) : 1 - 2
  • [2] Anderson D, 2001, UNIVERSAL SERIAL BUS, P19
  • [3] Anderson D, 1999, FIREWIRE SYSTEM ARCH, P15
  • [4] [Anonymous], FUNDAMENTALS BIOSTAT
  • [5] How to build an inexpensive laparoscopic webcam-based trainer
    Beatty, JD
    [J]. BJU INTERNATIONAL, 2005, 96 (04) : 679 - 682
  • [6] How to build your own laparoscopic trainer
    Blacker, AJR
    [J]. JOURNAL OF ENDOUROLOGY, 2005, 19 (06) : 748 - 752
  • [7] Integrating simulation into a surgical residency program - Is voluntary participation effective?
    Chang, L.
    Petros, J.
    Hess, D. T.
    Rotondi, C.
    Babineau, T. J.
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2007, 21 (03): : 418 - 421
  • [8] CONOVER WJ, 1980, PRACTICAL NONPARAMET, P280
  • [9] Development of a model for training and evaluation of laparoscopic skills
    Derossis, AM
    Fried, GM
    Abrahamowicz, M
    Sigman, HH
    Barkun, JS
    Meakins, JL
    [J]. AMERICAN JOURNAL OF SURGERY, 1998, 175 (06) : 482 - 487
  • [10] Reverse Alignment "Mirror Image" Visualization as a Laparoscopic Training Tool Improves Task Performance
    Dunnican, Ward J.
    Singh, T. Paul
    Ata, Ashar
    Bendana, Emma E.
    Conlee, Thomas D.
    Dolce, Charles J.
    Ramakrishnan, Rakesh
    [J]. SURGICAL INNOVATION, 2010, 17 (02) : 108 - 113