Incorporation of the electrode-electrolyte interface into finite-element models of metal microelectrodes

被引:89
作者
Cantrell, Donald R. [1 ,2 ]
Inayat, Samsoon [1 ,5 ]
Taflove, Allen [3 ]
Ruoff, Rodney S. [4 ]
Troy, John B. [1 ,2 ]
机构
[1] Northwestern Univ, Dept Biomed Engn, Evanston, IL 60208 USA
[2] Northwestern Univ, Inst Neurosci, Evanston, IL 60208 USA
[3] Northwestern Univ, Dept Elect Engn & Comp Sci, Evanston, IL 60208 USA
[4] Northwestern Univ, Dept Engn Mech, Evanston, IL 60208 USA
[5] Univ Engn & Technol, Dept Mech & Control Engn, Lahore 54890, Pakistan
关键词
D O I
10.1088/1741-2560/5/1/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An accurate description of the electrode-electrolyte interfacial impedance is critical to the development of computational models of neural recording and stimulation that aim to improve understanding of neuro-electric interfaces and to expedite electrode design. This work examines the effect that the electrode-electrolyte interfacial impedance has upon the solutions generated from time-harmonic finite-element models of cone- and disk-shaped platinum microelectrodes submerged in physiological saline. A thin-layer approximation is utilized to incorporate a platinum-saline interfacial impedance into the finite-element models. This approximation is easy to implement and is not computationally costly. Using an iterative nonlinear solver, solutions were obtained for systems in which the electrode was driven at ac potentials with amplitudes from 10 mV to 500 mV and frequencies from 100 Hz to 100 kHz. The results of these simulations indicate that, under certain conditions, incorporation of the interface may strongly affect the solutions obtained. This effect, however, is dependent upon the amplitude of the driving potential and, to a lesser extent, its frequency. The solutions are most strongly affected at low amplitudes where the impedance of the interface is large. Here, the current density distribution that is calculated from models incorporating the interface is much more uniform than the current density distribution generated by models that neglect the interface. At higher potential amplitudes, however, the impedance of the interface decreases, and its effect on the solutions obtained is attenuated.
引用
收藏
页码:54 / 67
页数:14
相关论文
共 57 条
[31]   How does the double layer at a disk electrode charge? [J].
Myland, JC ;
Oldham, KB .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2005, 575 (01) :81-93
[34]   RESISTANCE FOR FLOW OF CURRENT TO A DISK [J].
NEWMAN, J .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1966, 113 (05) :501-&
[35]   TRANSIENT-RESPONSE OF A DISK ELECTRODE WITH CONTROLLED POTENTIAL [J].
NISANCIOGLU, K ;
NEWMAN, J .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1973, 120 (10) :1356-1358
[36]  
NISANCIOGLU K, 1973, J ELECTROCHEM SOC, V120, P1339, DOI 10.1149/1.2403258
[37]  
Patrick James F, 2006, Trends Amplif, V10, P175, DOI 10.1177/1084713806296386
[38]   TIME-RESOLVED IMAGING OF CURRENT-DENSITY AT INLAID DISK ELECTRODES [J].
PHARR, CM ;
ENGSTROM, RC ;
TOPLE, RA ;
BEE, TK ;
UNZELMAN, PL .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1990, 278 (1-2) :119-128
[39]   Numerical determination of transmembrane voltage induced on irregularly shaped cells [J].
Pucihar, G ;
Kotnik, T ;
Valic, B ;
Miklavcic, D .
ANNALS OF BIOMEDICAL ENGINEERING, 2006, 34 (04) :642-652
[40]   Harmonic analysis of low-frequency bioelectrode behavior [J].
Richardot, A ;
McAdams, ET .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :604-612