Clinical value of intra-operative transit-time flow measurement for coronary artery bypass grafting: a prospective angiography-controlled study

被引:61
作者
Jokinen, Janne J. [1 ]
Werkkala, Kalervo [1 ]
Vainikka, Tiina [1 ]
Perakyla, Tarja [1 ]
Simpanen, Jarmo [1 ]
Ihlberg, Leo [1 ]
机构
[1] Helsinki Univ Hosp, Dept Cardiothorac Surg, FI-00029 Helsinki, Finland
关键词
Coronary artery bypass grafting; Transit-time flow measurement; Graft patency; Receiver operating characteristic analysis; SURGERY; PATENCY; SURVIVAL; REVISION; FAILURE; TRIAL;
D O I
10.1016/j.ejcts.2010.10.006
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Transit-time flow measurement (TTFM) is the most widely used method for intra-operative graft quality control in coronary artery bypass surgery. Although it may provide the opportunity for the surgeon to promptly revise the graft before the patient is discharged from the operating room, controlled clinical data on the ultimate usefulness of the TTFM are scarce. Clear cut-off values for when to revise grafts have not been set. Methods: A total of 204 consecutive grafts (left internal mammary artery (n = 46), vein graft (n = 155), and radial artery (n = 3)) underwent TTFM in 75 elective coronary artery bypass grafting (CABG) patients. The following parameters were recorded: mean graft flow (MGF), pulsatility index (PI), and insufficiency ratio (IR). After a mean follow-up of 199 +/- 42 days, coronary angiography was performed for assessment of graft patency. Results: A total of 166 grafts were found to be patent (85%), and 29 (15%) were completely occluded. The median and interquartile range (IQR) of MGF for the occluded grafts at the time of surgery was 38 ml min(-1) (IQR, 25 49 ml min(-1)) and for the patent grafts 45 ml min(-1) (IQR, 31-71 ml min(-1); p = ns]. The corresponding PI values were 3.3 (IQR, 2.8-5.0) and 2.2 (IQR, 1.7-3.2; p = 0.003), and the IR values were 1.6 (IQR, 0.6-6.1) and 0.2 (IQR, 0-2.2; p = 0.03). By receiver operating characteristic (ROC) analysis, the highest sensitivity (72%) and specificity (70%) were associated with a PI value > 3.0. However, 49 out of 70 such grafts (70%) were found to be patent. Furthermore, 10 out of 16 (63%) grafts, that had a combination of low flow (MGF < 15 ml min(-1)) and high PI (> 3.0), were patent at control angiography. Conclusions: TTFM predicts graft failure within the 6 months after CABG. However, specific cut-off recommendations for when to revise a graft cannot be set on the basis of TTFM. The cut-off values suggested in the literature lead to unnecessary graft revisions in the majority of cases, and, on the other hand, many technical defects probably remain unnoticed. Better methods to assess the quality of coronary artery bypass grafts are needed. (C) 2010 European Association for Cardio-Thoracic Surgery. Published by Elsevier B. V. All rights reserved.
引用
收藏
页码:918 / 923
页数:6
相关论文
共 25 条
[1]   Analyses of coronary graft patency after aprotinin use: Results from the international multicenter aprotinin graft patency experience (IMAGE) trial [J].
Alderman, EL ;
Levy, JH ;
Rich, JB ;
Nili, M ;
Vidne, B ;
Schaff, H ;
Uretzky, G ;
Pettersson, G ;
Thiis, JJ ;
Hantler, CB ;
Chaitman, B ;
Nadel, A .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1998, 116 (05) :716-729
[2]  
Alexander JH, 2005, JAMA-J AM MED ASSOC, V294, P2446
[3]   A comparison of transit-time flowmetry and intraoperative fluorescence imaging for assessing coronary artery bypass graft patency [J].
Balacumaraswami, L ;
Abu-Omar, Y ;
Choudhary, B ;
Pigott, D ;
Taggart, DP .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2005, 130 (02) :315-320
[4]   Intraoperative imaging techniques to assess coronary artery bypass graft patency [J].
Balacumaraswami, Lognathen ;
Taggart, David P. .
ANNALS OF THORACIC SURGERY, 2007, 83 (06) :2251-2257
[5]   High-frequency epicardial ultrasound: review of a multipurpose intraoperative tool for coronary surgery [J].
Budde, Ricardo P. J. ;
Bakker, Patricia F. A. ;
Gruendeman, Paul F. ;
Borst, Cornelius .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2009, 23 (03) :467-476
[6]   Graft revision after transit time flow measurement in off-pump coronary artery bypass grafting [J].
D'Ancona, G ;
Karamanoukian, HL ;
Ricci, M ;
Schmid, S ;
Bergsland, J ;
Salerno, TA .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 17 (03) :287-293
[7]   A randomized comparison of intraoperative indocyanine green angiography and transit-time flow measurement to detect technical errors in coronary bypass grafts [J].
Desai, Nimesh D. ;
Miwa, Senri ;
Kodama, David ;
Koyama, Taadaki ;
Cohen, Gideon ;
Pelletier, Marc P. ;
Cohen, Eric A. ;
Christakis, George T. ;
Goldman, Bernard S. ;
Fremes, Stephen E. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2006, 132 (03) :585-594
[8]   Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery [J].
Di Giammarco, Gabriele ;
Pano, Marco ;
Cirmeni, Sergio ;
Pelini, Piero ;
Vitolla, Giuseppe ;
Di Mauro, Michele .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2006, 132 (03) :468-474
[9]   Coronary bypass graft fate and patient outcome: Angiographic follow-up of 5,065 grafts related to survival and reoperation in 1,388 patients during 25 years [J].
FitzGibbon, GM ;
Kafka, HP ;
Leach, AJ ;
Keon, WJ ;
Hooper, GD ;
Burton, JR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (03) :616-626
[10]   STARTING ASPIRIN THERAPY AFTER OPERATION - EFFECTS ON EARLY GRAFT PATENCY [J].
GOLDMAN, S ;
COPELAND, J ;
MORITZ, T ;
HENDERSON, W ;
ZADINA, K ;
OVITT, T ;
KERN, KB ;
SETHI, G ;
SHARMA, GVRK ;
KHURI, S ;
RICHARDS, K ;
GROVER, F ;
MORRISON, D ;
WHITMAN, G ;
CHESLER, E ;
SAKO, Y ;
PACOLD, I ;
MONTOYA, A ;
DEMOTS, H ;
FLOTEN, S ;
DOHERTY, J ;
READ, R ;
SCOTT, S ;
SPOONER, T ;
MASUD, Z ;
HAAKENSON, C ;
HARKER, LA .
CIRCULATION, 1991, 84 (02) :520-526