Outbreaks of contaminated broncho-alveolar lavage related to intrinsically defective bronchoscopes

被引:29
作者
Cêtre, JC
Nicolle, MC
Salord, H
Pérol, M
Tigaud, S
David, G
Bourjault, M
Vanhems, P
机构
[1] Hop Croix Rousse, Hyg & Epidemiol Unit, F-69317 Lyon, France
[2] Hop Croix Rousse, Dept Microbiol, F-69317 Lyon, France
[3] Hop Croix Rousse, Dept Resp Med, F-69317 Lyon, France
[4] Hop Edouard Herriot, Dept Hosp Epidemiol & Infect Control, Lyon, France
[5] INSERM, U271, F-69008 Lyon, France
关键词
bronchoscopy; outbreak; equipment contamination; nosocomial infection; infection control;
D O I
10.1016/j.jhin.2004.12.020
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
From 5 March 2001 to 19 October 2001, outbreaks of broncho-alveolar lavage (BAL) contamination with Enterobacteraceae were detected in our 700-bed institution. We report the investigation of these outbreaks. A case was defined as the occurrence of pairs of specific Enterobacteraceae in BAL specimens among any patients who underwent bronchoscopy in the respiratory unit during the period of the outbreak. Contamination was identified in 117 BAL samples during three outbreaks among 418 patients, and was associated with bronchoscopes 11 and 12 (P < 0.001). The other five devices in use were not linked with the outbreaks. During the first outbreak, particular pairs of micro-organisms were associated with a specific bronchoscope (Klebsiella pneumoniae/Proteus vulgaris with bronchoscope 11, and Morganella morganii/Proteus mirabilis with bronchoscope 12). Cultures of sputa from two patients also yielded M. morganii some days after bronchoscopic examination. Isolates from contaminated BAL samples and bronchoscope 11 had similar patterns by pulsed-field get electrophoresis. No further cases occurred after removal of the implicated bronchoscopes. No deficiencies in disinfection procedures were detected and the source of contamination was found to be a loose port of the biopsy channel of the bronchoscope. Our findings underscore the urgent need to test bronchoscopic samples regularly and to improve the design and structure of bronchoscopes. (c) 2005 The Hospital Infection Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:39 / 45
页数:7
相关论文
共 28 条
  • [1] Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis - Community outbreak and nosocomial transmission via a contaminated bronchoscope
    Agerton, T
    Valway, S
    Gore, B
    Pozsik, C
    Plikaytis, B
    Woodley, C
    Onorato, I
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1997, 278 (13): : 1073 - 1077
  • [2] Future of flexible bronchoscopy
    Ahmad, M
    Dweik, RA
    [J]. CLINICS IN CHEST MEDICINE, 1999, 20 (01) : 1 - +
  • [3] Importance of lumen flow in liquid chemical sterilization
    Alfa, MJ
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 1999, 27 (04) : 373 - 374
  • [4] Alvarado CJ, 2000, AM J INFECT CONTROL, V28, P138
  • [5] AYLIFFE GAJ, 1999, HOSP EPIDEMIOLOGY IN, P680
  • [6] Blanc DS, 1997, INFECT CONT HOSP EP, V18, P134
  • [7] *CDCP, 2002, MMWR-MORBID MORTAL W, V51, P190
  • [8] Centers for Disease Control and Prevention (CDC), 1999, MMWR Morb Mortal Wkly Rep, V48, P557
  • [9] Cetre JC, 2003, NEW ENGL J MED, V348, P2039, DOI 10.1056/NEJM200305153482021
  • [10] Infection control in the bronchoscopy suite - A review of outbreaks and guidelines for prevention
    Culver, DA
    Gordon, SM
    Mehta, AC
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2003, 167 (08) : 1050 - 1056