Analysis of primary aromatic amines using precolumn derivatization by HPLC fluorescence detection and online MS identification

被引:22
作者
Zhao, Xianen [1 ,2 ]
Suo, Yourui [1 ]
机构
[1] Chinese Acad Sci, NW Inst Plateau Biol, Xining 810001, Peoples R China
[2] Chinese Acad Sci, Grad Sch, Beijing, Peoples R China
关键词
aromatic amines; derivatization; 2-(2-phenyl-1H-phenanthro-[9,10-d]imidazole-1-yl)acetic; acid; 2-9-acridoneyacetic acid; HPLC; MS;
D O I
10.1002/jssc.200700400
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
2-(2-Phenyl-1H-phenanthro-[9,10-d]imidazole-1-yl)-acetic acid (PPIA) and 2-(9-acridone)-acetic acid (AAA), two novel precolumn fluorescent derivatization reagents, have been developed and compared for analysis of primary aromatic amines by high performance liquid chromatographic fluorescence detection coupled with online mass spectrometric identification. PPIA and AAA react rapidly and smoothly with the aromatic amines on the basis of a condensation reaction using 1-ethyl-3-(3dimethylaminopropyl)-carbodiimide (EDC) as dehydrating catalyst to form stable derivatives with emission wavelengths at 380 and 440 nm, respectively. Taking six primary aromatic amines (aniline, 2-methylaniline, 2-methoxyaniline, 4-methylaniline, 4-chloroaniline, and 4-bromoaniline) as testing compounds, derivatization conditions such as coupling reagent, basic catalyst, reaction temperature and time, reaction solvent, and fluorescent labeling reagent concentration have also been investigated. With the better PPIA method, chromatographic separation of derivatized aromatic amines exhibited a good baseline resolution on an RP column. At the same time, by online mass spectrometric identification with atmospheric pressure chemical ionization (APCI) source in positive ion mode, the PPIA-labeled derivatives were characterized by easy-to-interpret mass spectra due to the prominent protonated molecular ion m/z [M + H](+) and specific fragment ions (MS/MS) m/z 335 and 295. The linear range is 24.41 fmol-200.0 pmol with correlation coefficients in the range of 0.9996-0.9999, and detection limits of PPIA-labeled aromatic amines are 0.12-0.21 nmol/L (S/N = 3). Method repeatability, precision, and recovery were evaluated and the results were excellent for the efficient HPLC analysis. The most important argument, however, was the high sensitivity and ease-of-handling of the PPIA method. Preliminary experiments with wastewater samples collected from the waterspout of a paper mill and its nearby soil where pollution with aromatic amines may be expected show that the method is highly validated with little interference in the chromatogram.
引用
收藏
页码:646 / 658
页数:13
相关论文
共 33 条
[1]   DERIVATIZATION WITH 4-CHLORO-7-NITROBENZOFURAZAN FOR LIQUID-CHROMATOGRAPHIC DETERMINATION OF HYDROXYPROLINE IN COLLAGEN HYDROLYSATE [J].
AHNOFF, M ;
GRUNDEVIK, I ;
ARFWIDSSON, A ;
FONSELIUS, J ;
PERSSON, BA .
ANALYTICAL CHEMISTRY, 1981, 53 (03) :485-489
[2]  
AIGHABSHA TS, 1976, ANAL CHIM ACTA, V85, P189
[3]  
AKIRA T, 2000, ANAL CHEM, V72, P732
[4]   Determination of aromatic amines in water samples by capillary electrophoresis with electrochemical and fluorescence detection [J].
Asthana, A ;
Bose, D ;
Durgbanshi, A ;
Sanghi, SK ;
Kok, WT .
JOURNAL OF CHROMATOGRAPHY A, 2000, 895 (1-2) :197-203
[5]  
BHATIA IS, 1976, ANN CHIM FRANCE, V1, P7
[6]   Liquid chromatography of aromatic amines with photochemical derivatization and tris(bipyridine)ruthenium(III) chemiluminescence detection [J].
Bolden, ME ;
Danielson, ND .
JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) :421-430
[7]   IMPROVEMENT IN THE METHOD OF SAMPLE STACKING FOR GRAVITY INJECTION IN CAPILLARY ZONE ELECTROPHORESIS [J].
BURGI, DS ;
CHIEN, RL .
ANALYTICAL BIOCHEMISTRY, 1992, 202 (02) :306-309
[8]   A new selective method for dimethylamine in water analysis by liquid chromatography using solid-phase microextraction and two-stage derivatization with o-phthalaldialdehyde and 9-fluorenylmethyl chloroformate [J].
Cháfer-Pericás, C ;
Herráez-Hernández, R ;
Campís-Falcó, P .
TALANTA, 2005, 66 (05) :1139-1145
[9]  
DJOZAN D, 1995, CHROMATOGRAPHIA, V41, P568, DOI 10.1007/BF02269721
[10]   SEPARATION OF AMINO-ACID ENANTIOMERS AND CHIRAL AMINES USING PRECOLUMN DERIVATIZATION WITH (+)-1-(9-FLUORENYL)ETHYL CHLOROFORMATE AND REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
EINARSSON, S ;
JOSEFSSON, B ;
MOLLER, P ;
SANCHEZ, D .
ANALYTICAL CHEMISTRY, 1987, 59 (08) :1191-1195