Accurate detection of heart rate using in-ear photoplethysmography in a clinical setting

被引:4
作者
Adams, Tim [1 ]
Wagner, Sophie [1 ]
Baldinger, Melanie [2 ]
Zellhuber, Incinur [1 ]
Weber, Michael [1 ]
Nass, Daniel [3 ]
Surges, Rainer [3 ]
机构
[1] Cosinuss GmbH, Munich, Germany
[2] Tech Univ Munich, Sport Equipment & Sport Mat, Munich, Germany
[3] Univ Hosp Bonn, Dept Epileptol, Bonn, Germany
来源
FRONTIERS IN DIGITAL HEALTH | 2022年 / 4卷
关键词
validation; heart rate; photoplethysmography; in-ear; wearable sensor; continuous monitoring; epilepsy; EPILEPSY; RELIABILITY; AGREEMENT; VALIDITY; MONITORS; ADULTS;
D O I
10.3389/fdgth.2022.909519
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Recent research has shown that photoplethysmography (PPG) based wearable sensors offer a promising potential for chronic disease monitoring. The aim of the present study was to assess the performance of an in-ear wearable PPG sensor in acquiring valid and reliable heart rate measurements in a clinical setting, with epileptic patients. Methods: Patients undergoing video-electroencephalography (EEG) monitoring with concomitant one-lead electrocardiographic (ECG) recordings were equipped with an in-ear sensor developed by cosinuss degrees. Results: In total, 2,048 h of recording from 97 patients with simultaneous ECG and in-ear heart rate data were included in the analysis. The comparison of the quality-filtered in-ear heart rate data with the reference ECG resulted in a bias of 0.78 bpm with a standard deviation of +2.54 bpm; Pearson's Correlation Coefficient PCC = 0.83; Intraclass Correlation Coefficient ICC = 0.81 and mean absolute percentage error MAPE = 2.57%. Conclusion: These data confirm that the in-ear wearable PPG sensor provides accurate heart rate measurements in comparison with ECG under realistic clinical conditions, especially with a signal quality indicator. Further research is required to investigate whether this technology is helpful in identifying seizure-related cardiovascular changes.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Evaluation of the accuracy and reliability for photoplethysmography based heart rate and beat-to-beat detection during daily activities
    Pietila, J.
    Mehrang, S.
    Tolonen, J.
    Helander, E.
    Jimison, H.
    Pavel, M.
    Korhonen, I.
    EMBEC & NBC 2017, 2018, 65 : 145 - 148
  • [42] Accurate Heart Rate Monitoring During Physical Exercises Using PPG
    Temko, Andriy
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (09) : 2016 - 2024
  • [43] Optimizing Remote Photoplethysmography Using Adaptive Skin Segmentation for Real-Time Heart Rate Monitoring
    Fouad, R. M.
    Omer, Osama A.
    Aly, Moustafa H.
    IEEE ACCESS, 2019, 7 : 76513 - 76528
  • [44] Video-based Noncontact Heart Rate Measurement Using Ear Features
    Cai, Xi
    Han, Guang
    Wang, Jinkuan
    PROCEEDINGS OF 2015 IEEE INTERNATIONAL CONFERENCE ON PROGRESS IN INFORMATCS AND COMPUTING (IEEE PIC), 2015, : 262 - 265
  • [45] Heart Rate Detection Using Microsoft Kinect: Validation and Comparison to Wearable Devices
    Gambi, Ennio
    Agostinelli, Angela
    Belli, Alberto
    Burattini, Laura
    Cippitelli, Enea
    Fioretti, Sandro
    Pierleoni, Paola
    Ricciuti, Manola
    Sbrollini, Agnese
    Spinsante, Susanna
    SENSORS, 2017, 17 (08)
  • [46] Seizure detection based on heart rate variability using a wearable electrocardiography device
    Jeppesen, Jesper
    Fuglsang-Frederiksen, Anders
    Johansen, Peter
    Christensen, Jakob
    Wustenhagen, Stephan
    Tankisi, Hatice
    Qerama, Erisela
    Hess, Alexander
    Beniczky, Sandor
    EPILEPSIA, 2019, 60 (10) : 2105 - 2113
  • [47] Heart Rate Measurement Using Face Detection in Video
    Nadrag, Carmen
    Poenaru, Vlad
    Suciu, George
    2018 12TH INTERNATIONAL CONFERENCE ON COMMUNICATIONS (COMM), 2018, : 131 - 134
  • [48] Assessing heart rate and blood pressure estimation from image photoplethysmography using a digital blood pressure meter
    Trirongjitmoah, Suchin
    Promking, Arphorn
    Kaewdang, Khanittha
    Phansiri, Nisarut
    Treeprapin, Kriengsak
    HELIYON, 2024, 10 (05)
  • [49] A review of significant research on epileptic seizure detection and prediction using heart rate variability
    Behbahani, Soroor
    TURK KARDIYOLOJI DERNEGI ARSIVI-ARCHIVES OF THE TURKISH SOCIETY OF CARDIOLOGY, 2018, 46 (05): : 414 - 421
  • [50] Comparative study on the effect of color spaces and color formats on heart rate measurement using the imaging photoplethysmography (IPPG) method
    Zhang, Chi
    Tian, Jing
    Li, Deyu
    Hou, Xiaoxu
    Wang, Li
    TECHNOLOGY AND HEALTH CARE, 2022, 30 : S391 - S402