Biomimetic hydroxyapatite (HAp) coatings on pure Mg and their physiological corrosion behavior

被引:32
作者
Hernandez, Leonardo [1 ]
Gonzalez, Jesus E. [2 ]
Barranco, Violeta [3 ]
Veranes-Pantoja, Yaymarilis [2 ]
Galvan, J. C. [3 ]
Rodriguez Gattorno, Geonel [1 ]
机构
[1] Natl Polytech Inst CINVESTAV IPN, Appl Phys Dept, Ctr Res & Adv Studies, Carr Ant Progreso Km 6, Merida 97310, Yucatan, Mexico
[2] Univ Havana, Biomat Ctr, Dept Ceram & Metall Biomat, Av Univ S-N Entre G & Ronda, Havana 6323, Cuba
[3] CSIC, Ctr Nacl Invest Metalurg, CENIM, Av Gregorio Amo 8, Madrid 28040, Spain
关键词
Magnesium; Hydroxyapatite; Biomimetic deposition; Supersaturated calcification solutions; Corrosion; EIS; CALCIUM-PHOSPHATE COATINGS; CA-P COATINGS; MAGNESIUM ALLOY; TRICALCIUM PHOSPHATE; DEGRADATION; SULFATE; BIODEGRADATION; IMPEDANCE; XPS; MICROSTRUCTURE;
D O I
10.1016/j.ceramint.2021.09.206
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The deposition of bioactive hydroxyapatite (HAp) coatings on pure magnesium (Mg) surfaces was studied. A modification of the biomimetic deposition method was used, which consisted in the incorporation of a supersaturated calcification solution (SCS). Different preparation parameters were modulated to obtain HAp and/or doped apatite coatings with different characteristics. The influence of the surface pre-treatments and soaking time in SCS on the physicochemical properties and the corrosion performance of the biomimetic HAp coatings were evaluated. Two different pre-treatments were explored to condition the sample's surface for the biomimetic HAp deposition. After the surface pre-treatments, the Mg samples were covered with HAp coatings by soaking in SCS during two immersion periods specifically chosen for this study: 2 h and 6 h, respectively. Scanning electron microscopy, coupled with energy-dispersive X-ray spectroscopy (SEM/EDS), X-ray photoelectron spectroscopy (XPS), X-ray diffraction (XRD), Raman Spectroscopy, and Fourier transform infrared (FTIR), were applied for the chemical, morphological, and structural characterization of the resulting biomimetic HAp-coatings. The in vitro corrosion behavior was studied by open circuit potential (OCP) measurements and electrochemical impedance spectroscopy (EIS) during immersion tests in Ringer's physiological solution for up to 14 days. The results showed that HAp coatings were successfully deposited on the pure Mg substrates. Furthermore, the corrosion studies in Ringer's solution showed that HAp coating behavior depends not only on the pre-treatment type but also on the immersion time in SCS. These results open the door to propose different strategies to control-on demand-the corrosion rate of Mg coated with biomimetic HAp. These systems could be a promising alternative for designing and developing new temporary implants for the human body.
引用
收藏
页码:1208 / 1222
页数:15
相关论文
共 81 条
[21]  
2-1
[22]   THE EFFECT OF CALCIUM-PHOSPHATE CERAMIC COMPOSITION AND STRUCTURE ON INVITRO BEHAVIOR .1. DISSOLUTION [J].
DUCHEYNE, P ;
RADIN, S ;
KING, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (01) :25-34
[23]   Preparation and bioactivity evaluation of bone-like hydroxyapatite nanopowder [J].
Fathi, M. H. ;
Hanifi, A. ;
Mortazavi, V. .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2008, 202 (1-3) :536-542
[24]   Electrochemical Impedance Spectroscopy for the Measurement of the Corrosion Rate of Magnesium Alloys: Brief Review and Challenges [J].
Feliu Jr, Sebastian .
METALS, 2020, 10 (06) :1-23
[25]   In vitro corrosion behaviour of surgical 316LVM stainless steel modified by Si+ ion implantation - An electrochemical impedance spectroscopy study [J].
Galvan, J. C. ;
Larrea, M. T. ;
Braceras, I. ;
Multigner, M. ;
Gonzalez-Carrasco, J. L. .
JOURNAL OF ALLOYS AND COMPOUNDS, 2016, 676 :414-427
[26]  
Gayathri B., 2019, Materials Today: Proceedings, V18, P1678, DOI 10.1016/j.matpr.2019.05.264
[27]   Magnesium degradation under physiological conditions - Best practice [J].
Gonzalez, Jorge ;
Hou, Rui Qing ;
Nidadavolu, Eshwara P. S. ;
Willumeit-Roemer, Regine ;
Feyerabend, Frank .
BIOACTIVE MATERIALS, 2018, 3 (02) :174-185
[28]   The mechanism of deposition of calcium phosphate coatings from solution onto magnesium alloy AZ31 [J].
Gray-Munro, J. E. ;
Strong, M. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2009, 90A (02) :339-350
[29]   Biological performance of uncoated and octacalcium phosphate-coated Ti6A14V [J].
Habibovic, P ;
Li, JP ;
van der Valk, CM ;
Meijer, G ;
Layrolle, P ;
van Blitterswijk, CA ;
de Groot, K .
BIOMATERIALS, 2005, 26 (01) :23-36
[30]   Hydroxyapatite coating of AZ31 magnesium alloy by a solution treatment and its corrosion behavior in NaCl solution [J].
Hiromoto, Sachiko ;
Tomozawa, Masanari .