Spore Film Dosimeters Are Feasible for UV Dose Monitoring During Heliotherapy

被引:3
作者
Vahavihu, Katja [1 ,2 ]
Ylianttila, Lasse [3 ]
Kautiainen, Hannu [4 ,5 ]
Tuohimaa, Pentti [6 ]
Reunala, Timo [2 ,7 ]
Snellman, Erna [1 ,2 ]
机构
[1] Paijat Hame Cent Hosp, Dept Dermatol, Lahti, Finland
[2] Tampere Univ Hosp, Dept Dermatol, Tampere, Finland
[3] Non Ionizing Radiat Lab, Radiat & Nucl Safety Author, Helsinki, Finland
[4] Cent Finland Cent Hosp, Unit Family Practice, Jyvaskyla, Finland
[5] ORTON & Rehabil Unit, Helsinki, Finland
[6] Univ Tampere, Dept Anat, Sch Med, FIN-33101 Tampere, Finland
[7] Univ Tampere, Dept Dermatol, Sch Med, FIN-33101 Tampere, Finland
关键词
ULTRAVIOLET-RADIATION EXPOSURE; VITAMIN-D BALANCE; ERYTHEMA; SUNLIGHT;
D O I
10.1111/j.1751-1097.2010.00769.x
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The objective of the study was to compare Bacillus subtilis spore film dosimeters with a Robertson Berger UV meter (RB meter) and diary records for assessing personal UV-B doses during a 13-day heliotherapy (HT) for atopic dermatitis (AD). In addition, the relationship between the personal UV-B dose and change in serum 25-hydroxyvitamin D (25(OH)D) was studied. Altogether 21 adult patients with AD completed the study arranged in the Canary Islands, either in January or March 2005. The spore film dosimeters were used throughout the day during the HT. Serum 25(OH)D was analyzed using radioimmunoassay. The mean personal UV-B dose measured with the dosimeters was 75 SED in January and 131 SED in March. The respective results gained from the RB meter combined with diary records were 63 SED and 119 SED showing a close correlation with the dosimeter results. Serum 25(OH)D concentration increased by 9.7 nmol L-1 in January and by 26.0 7 nmol L-1 in March. The increase in serum 25(OH)D correlated with the UV-B dose received. The patients complied well to use the dosimeters. We conclude spore films to be a feasible and reliable personal UV dosimeter in vivo in field conditions.
引用
收藏
页码:1174 / 1178
页数:5
相关论文
共 31 条
[1]   Biological UV dosimeters in the assessment of the biological hazard from environmental radiation [J].
Bérces, A ;
Fekete, A ;
Gáspár, S ;
Gróf, P ;
Rettberg, P ;
Horneck, G ;
Rontó, G .
JOURNAL OF PHOTOCHEMISTRY AND PHOTOBIOLOGY B-BIOLOGY, 1999, 53 (1-3) :36-43
[2]  
Commission Internationale de l'Eclairage (CIE), 2006, 174 CIE, V174
[3]  
Commission Internationale de lEclairage (CIE), 1999, 171661999E CIE ISO
[4]   POSSIBLE DOSIMETER FOR ULTRAVIOLET-RADIATION [J].
DAVIS, A ;
DEANE, GHW ;
DIFFEY, BL .
NATURE, 1976, 261 (5556) :169-170
[5]   The standard erythema dose: a new photobiological concept [J].
Diffey, BL ;
Jansen, CT ;
Urbach, F ;
Wulf, HC .
PHOTODERMATOLOGY PHOTOIMMUNOLOGY & PHOTOMEDICINE, 1997, 13 (1-2) :64-66
[6]   Fast simulation tool for ultraviolet radiation at the earth's surface [J].
Engelsen, O ;
Kylling, A .
OPTICAL ENGINEERING, 2005, 44 (04) :1-7
[7]  
Even-Paz Z, 2003, ISRAEL MED ASSOC J, V5, P87
[8]   Determination of erythema-effective solar radiation in Japan and Germany with a spore monolayer film optimized for the detection of UVB and UVA - results of a field campaign [J].
Furusawa, Y ;
Quintern, LE ;
Holtschmidt, H ;
Koepke, P ;
Saito, M .
APPLIED MICROBIOLOGY AND BIOTECHNOLOGY, 1998, 50 (05) :597-603
[9]   'Vitamin D' biodosimeter: basic characteristics and potential applications [J].
Galkin, ON ;
Terenetskaya, IP .
JOURNAL OF PHOTOCHEMISTRY AND PHOTOBIOLOGY B-BIOLOGY, 1999, 53 (1-3) :12-19
[10]   Ultraviolet protection factor of fabrics: comparison of laboratory and field-based measurements [J].
Gambichler, T ;
Hatch, KL ;
Avermaete, A ;
Bader, A ;
Herde, M ;
Altmeyer, P ;
Hoffmann, K .
PHOTODERMATOLOGY PHOTOIMMUNOLOGY & PHOTOMEDICINE, 2002, 18 (03) :135-140