An Efficient Method for Electrical Conductivity Measurement in the RF Range

被引:6
作者
Giovannetti, Giulio [1 ]
Frijia, Francesca [2 ]
Menichetti, Luca [1 ]
Hartwig, Valentina [1 ,3 ]
Viti, Vittorio [4 ]
Landini, Luigi [5 ]
机构
[1] CNR, Inst Clin Physiol, Pisa, Italy
[2] G Monasterio Fdn, Pisa, Italy
[3] Univ Pisa, Interdept Res Ctr E Piaggio, Pisa, Italy
[4] Esaote Biomed Spa, Genoa, Italy
[5] Univ Pisa, Dept Informat Engn, Pisa, Italy
关键词
electrical conductivity; radiofrequency measurements; magnetic resonance phantom; saline solution; dielectric properties; TO-NOISE RATIO; DIELECTRIC-PROPERTIES; BIOLOGICAL TISSUES; EQUIVALENT; SURFACE; COILS; DESIGN;
D O I
10.1002/cmr.b.20165
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Many of the hardware procedure like radiofrequency (RE) coils testing and simulation experiments such as the estimation of temperature changes in magnetic resonance imaging experiments require saline phantoms. These kinds of experimental models are generally obtained using a well-defined container filled with conductive salt solution (i.e., diluted NaCl solution) or using a combination of conductive gel with a semi-solid consistency. These phantoms are commonly prepared according to the standards values of the principal parameter concerning theirs dielectric properties (i.e., permittivity and electrical conductivity). Although several permittivity measurement methods have already been reported in literature, most of the data describing the electrical conductivity of these saline solutions provided by different authors are often conflicting and the frequencies of each measure are not always specified. Besides, low-cost commercial conductimeters could provide measurements overlooking the conductivity dependence on the frequency, while the measurements in the RF range require costly coaxial probes which work with network analyzers with multi-electrodes working in solution. In this article, we describe a system based on self-made coil designed to measure the electrical conductivity of conductive solutions in the RE range, without any contact between the sample and the measurement probe. The estimated value of conductivity has been performed combining theoretical calculation of induced resistance in the phantom volume and experimental measurements using workbench instrumentations. The method can be applied to the whole RE range measurements by tuning the capacitor value of the resonant circuit. (C) 2010 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 37B: 160-166, 2010
引用
收藏
页码:160 / 166
页数:7
相关论文
共 38 条
[1]  
Baker J, 2005, 1536 NIST
[2]   POLARIZATION IN ELECTROLYTIC SOLUTIONS .1. THEORY [J].
CHANG, HC ;
JAFFE, G .
JOURNAL OF CHEMICAL PHYSICS, 1952, 20 (07) :1071-1077
[3]  
CHOU CK, 1984, BIOELECTROMAGNETICS, V5, P435, DOI 10.1002/bem.2250050408
[4]   QUICK MEASUREMENT OF NMR-COIL SENSITIVITY WITH A DUAL-LOOP PROBE [J].
DARRASSE, L ;
KASSAB, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) :1841-1844
[5]   Influence of salts and phenolic compounds on olive mill wastewater detoxification using superabsorbent polymers [J].
Davies, LC ;
Novais, JM ;
Martins-Dias, S .
BIORESOURCE TECHNOLOGY, 2004, 95 (03) :259-268
[6]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[7]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[8]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[9]   Dielectric measurements using a coaxial resonator opened to a waveguide below cut-off [J].
Galwas, BA ;
Piotrowski, JK ;
Skulski, J .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) :511-514
[10]   Conductor geometry and capacitor quality for performance optimization of low-frequency birdcage coils [J].
Giovannetti, G ;
Francesconi, R ;
Landini, L ;
Santarelli, MF ;
Positano, V ;
Viti, V ;
Benassi, A .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2004, 20B (01) :9-16