Intraocular lens bicylindric power calculation method: Using both flat and steep K readings to improve intraocular lens power prediction

被引:6
作者
Calvo-Sanz, Jorge A. [1 ]
Ruiz-Alcocer, Javier [2 ]
Sanchez-Tena, Miguel A. [2 ]
机构
[1] Inst Ciencias Visuales INCIVI, Dept Optometry, Madrid, Spain
[2] Univ Europea Madrid, Madrid, Spain
关键词
Lenses; intraocular; meridian keratometry; astigmatism; refraction; SURGICALLY INDUCED ASTIGMATISM; INDUCED REFRACTIVE CHANGE; CATARACT-SURGERY; CORNEAL ASTIGMATISM; HOFFER-Q; HAIGIS FORMULAS; VECTOR ANALYSIS; SHORT EYES; SRK/T; ACCURACY;
D O I
10.1177/1120672117754170
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare and analyze the accuracy of the refractive outcomes obtained in intraocular lens power calculation using the classical calculation method with mean keratometry (K) and the calculation method with both K meridians presented in this article. Methods: A total of 62 eyes of 62 subjects who were undergoing cataract surgery were included in this study. Optical biometry was performed using mean K and Haigis formula for classical intraocular lens calculation methods to achieve intraocular lens power; 4 weeks after surgery, prior to medical discharge, subjective refraction was made. Alternatively, intraocular lens power was calculated with bicylindric method using both keratometry readings, obtaining spherocylindrical refractive expected outcomes. Finally, results obtained with intraocular lens calculation methods, bicylindric method, and Haigis formula were compared. Results: Spherical equivalent calculated by classical intraocular lens calculation methods using Haigis formula (H-SE) was -0.027 +/- 0.115 D and using bicylindric method (B-SE) was -0.080 +/- 0.222 D. Achieved spherical equivalent obtained 4 weeks after surgery (A-SE) was -0.144 +/- 0.268 D. Difference between H-SE and A-SE was -0.117 D (p = 0.002). Difference between B-SE and A-SE was not significant (-0.054 D, p = 0.109). Analysis in refraction groups showed a positive correlation between A-SE confronted to B-SE and H-SE (r = 0.313; p = 0.013 and r = 0.562; p < 0.001, respectively). This indicated a reliability in ametropic group prediction of 0.767 in H-SE and 0.843 in B-SE. Conclusion: Intraocular lens calculation with bicylindric method could be more accurate and had more reliability than classical intraocular lens calculation method. Bicylindric method adds astigmatism control and provides a reliable expected spherocylindrical refraction.
引用
收藏
页码:559 / 565
页数:7
相关论文
共 33 条
[1]   SURGICALLY INDUCED ASTIGMATISM [J].
ALPINS, N .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF OPHTHALMOLOGY, 1994, 22 (03) :217-217
[2]   Intraocular lens power calculation after corneal refractive surgery: Double-K method [J].
Aramberri, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (11) :2063-2068
[3]   Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry [J].
Aristodemou, Petros ;
Cartwright, Nathaniel E. Knox ;
Sparrow, John M. ;
Johnston, Robert L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (01) :63-71
[4]   Effect on refractive outcomes after cataract surgery of intraocular lens constant personalization using the Haigis formula [J].
Charalampidou, Sofia ;
Cassidy, Lorraine ;
Ng, Eugene ;
Loughman, James ;
Nolan, John ;
Stack, Jim ;
Beatty, Stephen .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (07) :1081-1089
[5]  
CRAVY TV, 1979, OPHTHALMIC SURG LAS, V10, P38
[6]   Refractive Stabilization and Corneal Swelling After Cataract Surgery [J].
de Juan, Victoria ;
Maria Herreras, Jose ;
Perez, Inmaculada ;
Morejon, Angela ;
Rio-San Cristobal, Ana ;
Martin, Raul ;
Fernandez, Itziar ;
Rodriguez, Guadalupe .
OPTOMETRY AND VISION SCIENCE, 2013, 90 (01) :31-36
[7]  
Doshi D, 2017, J CLIN DIAGN RES, V11, pNC1, DOI 10.7860/JCDR/2017/22095.9136
[8]   Comparison of SRK/T and Haigis formulas for predicting corneal astigmatism correction with tonic intraocular lenses [J].
Eom, Youngsub ;
Song, Jong Suk ;
Kim, Yong Yeon ;
Kim, Hyo Myung .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (08) :1650-1657
[9]   Comparison of Hoffer Q and Haigis Formulae for Intraocular Lens Power Calculation According to the Anterior Chamber Depth in Short Eyes [J].
Eom, Youngsub ;
Kang, Su-Yeon ;
Song, Jong Suk ;
Kim, Yong Yeon ;
Kim, Hyo Myung .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2014, 157 (04) :818-824
[10]   Meridional analysis for calculating the expected spherocylindrical refraction in eyes with toric intraocular lenses [J].
Fam, Han Bor ;
Lim, Kooi Ling .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (12) :2072-2076