Development of a flow standard to enable highly reproducible measurements of deformability of stored red blood cells in a microfluidic device

被引:15
|
作者
Robidoux, Jonathan [1 ]
Laforce-Lavoie, Audrey [1 ]
Charette, Steve J. [2 ]
Shevkoplyas, Sergey S. [3 ]
Yoshida, Tatsuro [4 ]
Brouard, Danny [1 ]
Lewin, Antoine [1 ]
机构
[1] Hema Quebec, Med Affairs & Innovat, Montreal, PQ, Canada
[2] Univ Laval, Biochem Microbiol & Bioinformat Dept, Montreal, PQ, Canada
[3] Univ Houston, Dept Biomed Engn, Houston, TX USA
[4] Hemanext Inc, Lexington, MA USA
关键词
ERYTHROCYTE DEFORMABILITY; MICROVASCULAR NETWORK; STORAGE LESION; TRANSFUSION; RHEOLOGY; PERFUSION; TOOL;
D O I
10.1111/trf.15770
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Great deformability allows red blood cells (RBCs) to flow through narrow capillaries in tissues. A number of microfluidic devices with capillary-like microchannels have been developed to monitor storage-related impairment of RBC deformability during blood banking operations. This proof-of-concept study describes a new method to standardize and improve reproducibility of the RBC deformability measurements using one of these devices. STUDY DESIGN AND METHODS The rate of RBC flow through the microfluidic capillary network of the microvascular analyzer (MVA) device made of polydimethylsiloxane was measured to assess RBC deformability. A suspension of microbeads in a solution of glycerol in phosphate-buffered saline was developed to be used as an internal flow rate reference alongside RBC samples in the same device. RBC deformability and other in vitro quality markers were assessed weekly in six leukoreduced RBC concentrates (RCCs) dispersed in saline-adenine-glucose-mannitol additive solution and stored over 42 days at 4 degrees C. RESULTS The use of flow reference reduced device-to-device measurement variability from 10% to 2%. Repeated-measure analysis using the generalized estimating equation (GEE) method showed a significant monotonic decrease in relative RBC flow rate with storage from Week 0. By the end of storage, relative RBC flow rate decreased by 22 +/- 6% on average. CONCLUSIONS The suspension of microbeads was successfully used as a flow reference to increase reproducibility of RBC deformability measurements using the MVA. Deformability results suggest an early and late aging phase for stored RCCs, with significant decreases between successive weeks suggesting a highly sensitive measurement method.
引用
收藏
页码:1032 / 1041
页数:10
相关论文
共 16 条
  • [1] Deformability of Stored Red Blood Cells
    Barshtein, Gregory
    Pajic-Lijakovic, Ivana
    Gural, Alexander
    FRONTIERS IN PHYSIOLOGY, 2021, 12
  • [2] Reduced deformability of stored red blood cells is associated with generation of extracellular vesicles
    McVey, M. J.
    Kuebler, W. M.
    Orbach, A.
    Arbell, D.
    Zelig, O.
    Barshtein, G.
    Yedgar, S.
    TRANSFUSION AND APHERESIS SCIENCE, 2020, 59 (05)
  • [3] A Microfluidic Deformability Assessment of Pathological Red Blood Cells Flowing in a Hyperbolic Converging Microchannel
    Faustino, Vera
    Rodrigues, Raquel O.
    Pinho, Diana
    Costa, Elisio
    Santos-Silva, Alice
    Miranda, Vasco
    Amaral, Joana S.
    Lima, Rui
    MICROMACHINES, 2019, 10 (10)
  • [4] Microfluidic analysis of cellular deformability of normal and oxidatively damaged red blood cells
    Kwan, Joanne M.
    Guo, Quan
    Kyluik-Price, Dana L.
    Ma, Hongshen
    Scott, Mark D.
    AMERICAN JOURNAL OF HEMATOLOGY, 2013, 88 (08) : 682 - 689
  • [5] Deformability based sorting of stored red blood cells reveals donor-dependent aging curves
    Islamzada, Emel
    Matthews, Kerryn
    Guo, Quan
    Santoso, Aline T.
    Duffy, Simon P.
    Scott, Mark D.
    Ma, Hongshen
    LAB ON A CHIP, 2020, 20 (02) : 226 - 235
  • [6] Modeling the Effect of Red Blood Cells Deformability on Blood Flow Conditions in Human Carotid Artery Bifurcation
    Urevc, Janez
    Zun, Iztok
    Brumen, Milan
    Stok, Boris
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2017, 139 (01):
  • [7] Washing of stored red blood cells by an autotransfusion device before transfusion
    de Vroege, R.
    Wildevuur, W. R.
    Muradin, J. A. G.
    Graves, D.
    van Oeveren, W.
    VOX SANGUINIS, 2007, 92 (02) : 130 - 135
  • [8] Microfluidic evaluation of red cells collected and stored in modified processing solutions used in blood banking
    Wang, Yimeng
    Giebink, Adam
    Spence, Dana M.
    INTEGRATIVE BIOLOGY, 2014, 6 (01) : 65 - 75
  • [9] The mechanical properties of stored red blood cells measured by a convenient microfluidic approach combining with mathematic model
    Wang, Ying
    You, Guoxing
    Chen, Peipei
    Li, Jianjun
    Chen, Gan
    Wang, Bo
    Li, Penglong
    Han, Dong
    Zhou, Hong
    Zhao, Lian
    BIOMICROFLUIDICS, 2016, 10 (02):
  • [10] Dynamics of Individual Red Blood Cells Under Shear Flow: A Way to Discriminate Deformability Alterations
    Atwell, Scott
    Badens, Catherine
    Charrier, Anne
    Helfer, Emmanuele
    Viallat, Annie
    FRONTIERS IN PHYSIOLOGY, 2022, 12