Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs

被引:76
作者
Cristofolini, L
Teutonico, AS
Monti, L
Cappello, A
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Fac Engn, Bologna, Italy
关键词
hip prostheses; in vitro testing; fatigue; migration; cement damage;
D O I
10.1016/S0021-9290(03)00191-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The long-term clinical success of cemented hip stems is influenced both by the implant design, and by the surgical procedure. A methodology is proposed for discriminating between implant designs with different clinical outcomes. The protocol was designed with industrial pre-clinical validation in mind. Two cemented stem types were tested, one (Lubinus SPII) having good and the other (Muller Curved) having poor clinical outcomes. Three implants for each type were subjected to a mechanical in vitro test of one million loading cycles. Each cycle reproduced the load components of stair climbing. Interface shear micromotion was measured during the test in the direction of rotation and along the stem axis. The stem roughness before and after the test was compared. After the test, the cement mantles were retrieved and inspected through dye penetrants to detect evidences of micro-damage. For each specimen, the events of the loosening process were examined, based on the in vitro data available, so as to analyze the whole failure mechanism. The protocol developed was sensitive to the implant design, with significantly different results being found for the two stem types, both in terms of stem-cement micromotions, surface roughness alteration, and cement mantle damage. The information yielded by the three different investigation techniques was consistent for each of the two groups of specimens tested, allowing a better understanding of the failure process. In vitro inducible micromotion and permanent migration measurements, together with cement-stem interface fretting damage and cement fatigue damage, can help predicting the clinical performance of cemented stems. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1603 / 1615
页数:13
相关论文
共 48 条
  • [1] ASEPTIC LOOSENING OF CEMENT-ANCHORED TOTAL HIP REPLACEMENTS
    ALDINGER, G
    GEKELER, J
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1982, 100 (01) : 19 - 25
  • [2] [Anonymous], 67 ANN M AM AC ORTH
  • [3] Baleani M, 2000, J BIOMED MATER RES, V50, P605, DOI 10.1002/(SICI)1097-4636(20000615)50:4<605::AID-JBM17>3.0.CO
  • [4] 2-P
  • [5] IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1995, 28 (05) : 535 - 553
  • [6] STEM CURVATURE AND LOAD ANGLE INFLUENCE THE INITIAL RELATIVE BONE-IMPLANT MOTION OF CEMENTLESS FEMORAL STEMS
    BERZINS, A
    SUMNER, DR
    ANDRIACCHI, TP
    GALANTE, JO
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1993, 11 (05) : 758 - 769
  • [7] Three-dimensional primary stability of cementless femoral stems
    Buhler, DW
    Berlemann, U
    Lippuner, K
    Jaeger, P
    Nolte, LP
    [J]. CLINICAL BIOMECHANICS, 1997, 12 (02) : 75 - 86
  • [8] SOME FATIGUE CHARACTERISTICS OF THERMOPLASTICS
    CRAWFORD, RJ
    BENHAM, PP
    [J]. POLYMER, 1975, 16 (12) : 908 - 914
  • [9] A critical analysis of stress shielding evaluation of hip prostheses
    Cristofolini, L
    [J]. CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) : 409 - 483
  • [10] Mechanical validation of whole bone composite femur models
    Cristofolini, L
    Viceconti, M
    Cappello, A
    Toni, A
    [J]. JOURNAL OF BIOMECHANICS, 1996, 29 (04) : 525 - 535