Prospective long-term follow-up of the cementless bicontact hip stem with plasmapore coating

被引:21
作者
Ochs, U.
Eingartner, C.
Volkmann, R.
Ochs, B. G.
Huber, C.
Weller, S.
Weise, K.
机构
[1] Univ Tubingen, BG Trauma Ctr, Tubingen, Germany
[2] Caritas Hosp, Unit Trauma & Reconstruct Surg, Bad Mergentheim, Germany
[3] Ctr Clin, Dept Trauma & Reconstruct Surg, Bad Hersfeld, Germany
来源
ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE | 2007年 / 145卷
关键词
total hip arthroplasty; cementless; proximal fixation; long-term results;
D O I
10.1055/s-2007-965652
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aim: A prospective long-term study of the first 250 cementless Bicontact stems implanted in the BG Trauma Centre, Tuebingen, Germany. Method: All of the patients in this series (treatment period June 1987 to April 1990) who were still contactable were invited for clinical and radiological follow-up. The radiographs were analysed for signs of loosening in the form of Gruen lysis zones, stress shielding, subsidence behaviour, heterotopic ossification and spot welds. Results: The average follow-up period was 17.8 (16.7-19.5) years. The overall rate of follow-up was 65% (162 of 250) and 91% of patients who were still alive (162 of 779). The average patient age was 56.2 years at the time of operation and 74.0 years at follow-up. The average HHS was 81.6 points. In the course of the first 10 years (up to 03/1998), a total of 8 stem revisions had to be performed. In the period from 03/1998 to 0l/ 2007, 2 cases of loosening requiring revision occurred in the patients still alive at the time of follow-up. The survival rate calculated was thus 95.6%. Conclusion: These outstanding results provide enduring support for the philosophy of the cementless and bone-preserving fixation principles underlying the Bicontact hip stem with proximal intertrochanteric transmission of forces and high primary rotational stability.
引用
收藏
页码:S3 / S8
页数:6
相关论文
共 45 条
[21]  
ENGH CA, 2001, CLIN ORTHOP RELAT R, V393, P137
[22]  
EVANS BG, 1993, ORTHOP CLIN N AM, V24, P599
[24]  
HELLMAN EJ, 1999, CLIN ORTHOP RELAT R, V364, P164
[25]  
Huggler A H, 1993, Acta Orthop Belg, V59 Suppl 1, P215
[26]   CLINICAL AND RADIOGRAPHIC EVALUATION OF TOTAL HIP-REPLACEMENT - A STANDARD SYSTEM OF TERMINOLOGY FOR REPORTING RESULTS [J].
JOHNSTON, RC ;
FITZGERALD, RH ;
HARRIS, WH ;
POSS, R ;
MULLER, ME ;
SLEDGE, CB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (02) :161-168
[27]  
JONES LC, 1987, CLIN ORTHOP RELAT R, V225, P192
[28]   The porous coated anatomic total hip replacement - A ten to fourteen-year follow-up study of a cementless total hip arthroplasty [J].
Kawamura, H ;
Dunbar, MJ ;
Murray, P ;
Bourne, RB ;
Rorabeck, CH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (09) :1333-1338
[29]  
KEISU KS, 2001, CLIN ORTHOP RELAT R, V382, P133
[30]   Primary total hip arthroplasty with a cementless porous-coated anatomic total hip prosthesis - 10-to 12-year results of prospective and consecutive series [J].
Kim, YH ;
Kim, JS ;
Cho, SH .
JOURNAL OF ARTHROPLASTY, 1999, 14 (05) :538-548