Validation of aberrometry-based relative peripheral refraction measurements

被引:34
作者
Berntsen, David A. [1 ]
Mutti, Donald O. [1 ]
Zadnik, Karla [1 ]
机构
[1] Ohio State Univ, Coll Optometry, Columbus, OH 43210 USA
关键词
aberrometry-based refraction; autorefraction; ocular shape; relative peripheral refraction;
D O I
10.1111/j.1475-1313.2007.00535.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To validate aberrometry-based relative peripheral refraction (RPR) measurements and to examine RPR differences in the nasal and temporal horizontal visual field. Methods: Thirty subjects underwent cycloplegia in the right eye. Ten measurements each were made centrally, 30 degrees nasally on the retina and 30 degrees temporally on the retina using the Complete Ophthalmic Analysis System (COAS) aberrometer and the Grand Seiko WR-5100K autorefractor in random order. A 2-mm COAS analysis diameter was used to approximate the Grand Seiko autorefractor measurement beam diameter. Nasal and temporal RPR were calculated as the difference between the average nasal or temporal spherical equivalent and the average central spherical equivalent. Repeated measures ANOVAs were used to test for significant differences between the RPR readings of each instrument in each retinal location (nasal or temporal) and to examine refractive error in each direction of gaze for both instruments. Results: Central spherical equivalent refractive error as measured by the Grand Seiko autorefractor ranged from +0.63 D to -8.41 D (mean +/- S.D. = -2.63 D +/- 2.05 D). There were no significant differences between the instruments for RPR measurements (p = 0.34). The nasal RPR was significantly more hyperopic than the temporal RPR (mean +/- S.D. = +0.45 D +/- 1.04 D, p = 0.02). Spherical equivalent refractive error in all directions of gaze was more myopic with the COAS than with the Grand Seiko autorefractor (mean +/- S.D. = -0.41 D +/- 0.61 D, p < 0.0001). Conclusions: RPR measurements with the COAS are feasible and equivalent to those made with the Grand Seiko autorefractor. The COAS can be used to simultaneously collect RPR and peripheral aberration data. Nasal and temporal RPR measurements were significantly different in our sample.
引用
收藏
页码:83 / 90
页数:8
相关论文
共 42 条
  • [1] Peripheral refraction along the horizontal and vertical visual fields in myopia
    Atchison, DA
    Pritchard, N
    Schmid, KL
    [J]. VISION RESEARCH, 2006, 46 (8-9) : 1450 - 1458
  • [2] Shape of the retinal surface in emmetropia and myopia
    Atchison, DA
    Pritchard, N
    Schmid, KL
    Scott, DH
    Jones, CE
    Pope, JM
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (08) : 2698 - 2707
  • [3] Monochromatic aberrations of human eyes in the horizontal visual field
    Atchison, DA
    Scott, DH
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2002, 19 (11): : 2180 - 2184
  • [4] Repeatability of autorefraction and axial length measurements after laser in situ keratomileusis
    Bailey, MD
    Twa, MD
    Mitchell, GL
    Dhaliwal, DK
    Jones, LA
    McMahon, TT
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (05) : 1025 - 1034
  • [5] BENNETT AG, 1998, BENNETT RABBETTS CLI, P220
  • [6] The effect of overnight contact lens corneal reshaping on higher-order aberrations and best-corrected visual acuity
    Berntsen, DA
    Barr, JT
    Mitchell, GL
    [J]. OPTOMETRY AND VISION SCIENCE, 2005, 82 (06) : 490 - 497
  • [7] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [8] Validation of a clinical Shack-Hartmann aberrometer
    Cheng, X
    Himebaugh, NL
    Kollbaum, PS
    Thibos, LN
    Bradley, A
    [J]. OPTOMETRY AND VISION SCIENCE, 2003, 80 (08) : 587 - 595
  • [9] Local changes in eye growth induced by imposed local refractive error despite active accommodation
    Diether, S
    Schaeffel, F
    [J]. VISION RESEARCH, 1997, 37 (06) : 659 - 668
  • [10] COMPARISON OF CYCLOPENTOLATE VERSUS TROPICAMIDE CYCLOPLEGIA IN CHILDREN
    EGASHIRA, SM
    KISH, LL
    TWELKER, JD
    MUTTI, DO
    ZADNIK, K
    ADAMS, AJ
    [J]. OPTOMETRY AND VISION SCIENCE, 1993, 70 (12) : 1019 - 1026