Data acquisition with a fast digitizer for large volume HPGe detectors

被引:35
作者
Mihailescu, L. C.
Borcea, C.
Plompen, A. J. M. [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
[2] Horia Hulubei Natl Inst Phys & Nucl Engn, Bucharest 76900, Romania
关键词
fast digitizer; large volume HPGe detectors; time-of-flight; neutron inelastic scattering;
D O I
10.1016/j.nima.2007.05.231
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 12-bits and 420 MSample/s waveform digitizer was used with large volume High Purity Germanium (HPGe) detectors for measurements of gamma-ray production cross-section from inelastic neutron scattering. For these time-of-flight measurements at the GELINA white neutron source, the use of a fast digitizer significantly increases the efficiency compared with a conventional data acquisition system. First, the pulse processing time required to obtain the amplitude with high resolution is significantly reduced so that pulse pile-up with the prompt gamma-burst is eliminated for neutron-induced events. Second, an improved time response is obtained for which the amplitude and rise time dependence is strongly reduced compared to that of a conventional constant fraction discriminator. Excellent energy and time resolution is obtained with algorithms suitable for on-line signal processing, so that data storage is under control. Bench tests are presented that compare methods of signal processing. For the best method, the data acquisition system based on the fast digitizer was tested during measurements of gamma production cross-sections for Pb-206 and Pb-208. A direct comparison was made with results obtained with conventional electronics operated in parallel. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:298 / 305
页数:8
相关论文
共 10 条
[1]  
DEJONGE S, 1987, GEDER2487 JRC CBNM
[2]   DIGITAL PULSE PROCESSING IN HIGH-RESOLUTION, HIGH-THROUGHPUT GAMMA-RAY SPECTROSCOPY [J].
GEORGIEV, A ;
GAST, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :770-779
[3]  
GONZALEZ J, 1997, MODULAR MULTIPARAMET
[4]   DIGITAL-TECHNIQUES FOR REAL-TIME PULSE SHAPING IN RADIATION MEASUREMENTS [J].
JORDANOV, VT ;
KNOLL, GF ;
HUBER, AC ;
PANTAZIS, JA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :261-264
[5]  
Knoll GF., 2000, Radiation Detection and Measurement, Vthird
[6]   High resolution measurement of neutron inelastic scattering and (n, 2n) cross-sections for 52Cr [J].
Mihailescu, L. C. ;
Borcea, C. ;
Koning, A. J. ;
Plompen, A. J. M. .
NUCLEAR PHYSICS A, 2007, 786 :1-23
[7]   A new HPGe setup at Gelina for measurement of gamma-ray production cross-sections from inelastic neutron scattering [J].
Mihailescu, LC ;
Oláh, L ;
Borcea, C ;
Plompen, AM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 531 (03) :375-391
[8]  
MIHAILESCU LC, 2006, 22343 EUR
[9]  
MIHAILESCU LC, 2005, P ENL WORKSH NEUTR M, P109