Patient doses in abdominal aortogram and aorta femoral runoff examinations

被引:4
作者
Chu, RYL
Parry, C
Thompson, W
Loeffler, C
机构
[1] Vet Affairs Med Ctr, Serv Radiol, Oklahoma City, OK 73104 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Dept Radiol Sci, Oklahoma City, OK 73190 USA
来源
HEALTH PHYSICS | 1998年 / 75卷 / 05期
关键词
fluoroscopy; effective dose; medical radiation; x rays;
D O I
10.1097/00004032-199811000-00004
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Radiation doses to adult male patients from abdominal aortogram and aorta femoral runoff examinations in a medical center were determined with the help of a dose-area product meter. The abdominal aortogram and aorta femoral runoff examination consisted of scout radiographs, fluoroscopy (to position a catheter near the area of interest), and serial films (to record the flow of contrast media). Measurements were converted to effective doses with the help of published results from Monte Carlo simulation calculations. Data from 19 male adult patients weighing 53 to 86 kg were analyzed. The resulting total effective dose had a value of 14.0 +/- 4.4 mSv (mean and standard deviation). The percent contribution by fluoroscopy was 18.5 +/- 9.9%. The fluoroscopy effective dose had a stronger correlation with the dose-area product (correlation coefficient of 0.97) than with duration of exposure (correlation coefficient of 0.84). Most of the radiation exposure in the observed abdominal aortogram and aorta femoral runoff examination was attributed to radiography.
引用
收藏
页码:487 / 491
页数:5
相关论文
共 17 条
[1]   ESTIMATION OF DOSES TO PATIENTS FROM COMPLEX CONVENTIONAL X-RAY EXAMINATIONS [J].
CALZADO, A ;
VANO, E ;
MORAN, P ;
CASTELLOTE, C ;
RUIZ, S ;
GONZALEZ, L .
BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (762) :539-546
[2]   ASSESSMENT OF ORGAN RADIATION-DOSES AND ASSOCIATED RISK FOR DIGITAL BIFEMORAL ARTERIOGRAPHY [J].
CASTELLANO, IA ;
MCNEILL, JG ;
THORP, NC ;
DANCE, DR ;
RAPHAEL, MJ .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (809) :502-507
[3]   Evaluation of an on-line patient exposure meter in neuroradiology [J].
Gkanatsios, NA ;
Huda, W ;
Peters, KR ;
Freeman, JA .
RADIOLOGY, 1997, 203 (03) :837-842
[4]  
Gray N., 1994, Australasian Physical and Engineering Sciences in Medicine, V17, P79
[5]  
HART D, 1995, NRPBSR262
[6]  
KERIAKES JG, 1980, HDB RAD DOSES NUCL M
[7]   REDUCTION OF PATIENT DOSES FROM BARIUM MEAL AND BARIUM ENEMA EXAMINATIONS THROUGH CHANGES IN EQUIPMENT FACTORS [J].
MARTIN, CJ ;
HUNTER, S .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (804) :1196-1205
[8]  
*NUCL ASS, 1987, RADC XRAY ENTR SKIN
[9]   PATIENT EFFECTIVE DOSE RESULTING FROM RADIOGRAPHIC EXAMINATIONS [J].
OKKALIDES, D ;
FOTAKIS, M .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (798) :564-572
[10]   Measurement of skin entrance exposure with a dose-area-product meter at chest radiography [J].
Parry, CK ;
Chu, RYL ;
Eaton, BG ;
Chen, CY .
RADIOLOGY, 1996, 201 (02) :574-575