Primary calibration of coiled 103Pd brachytherapy sources

被引:8
作者
Paxton, Adam B. [1 ]
Culberson, Wesley S. [1 ]
DeWerd, Larry A. [1 ]
Micka, John A. [1 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
关键词
calibration; brachytherapy; Pd-103;
D O I
10.1118/1.2815628
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Coiled Pd-103 brachytherapy sources have been developed by RadioMed Corporation for use as low-dose-rate (LDR) interstitial implants. The coiled sources are provided in integer lengths from 1 to 6 cm and address many common issues seen with traditional LDR brachytherapy sources. The current standard for determining the air-kerma strength (S-K) of low-energy LDR brachytherapy sources is the National Institute of Standards and Technology's Wide-Angle Free-Air Chamber (NIST WAFAC). Due to geometric limitations, however, the NIST WAFAC is unable to determine the SK of sources longer than 1 cm. This project utilized the University of Wisconsin's Variable-Aperture Free-Air Chamber (UW VAFAC) to determine the S-K of the longer coiled sources. The UW VAFAC has shown agreement in S-K values of 1 cm length coils to within 1% of those determined with the NIST WAFAC, but the UW VAFAC does not share the same geometric limitations as the NISTWAFAC. A new source holder was constructed to hold the coiled sources in place during measurements with the UW VAFAC. Correction factors for the increased length of the sources have been determined and applied to the measurements. Using the new source holder and corrections, the S-K of 3 and 6 cm coiled sources has been determined. Corrected UW VAFAC data and ionization current measurements from well chambers have been used to determine calibration coefficients for use in the measurement of 3 and 6 cm coiled sources in well chambers. Thus, the UW VAFAC has provided the first transferable, primary measurement of low-energy LDR brachytherapy sources with lengths greater than 1 cm. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:32 / 38
页数:7
相关论文
共 15 条
  • [1] [Anonymous], LAUR031987 LOS AL NA
  • [2] Large-volume ionization chamber with variable apertures for air-kerma measurements of low-energy radiation sources
    Culberson, WS
    DeWerd, LA
    Anderson, DR
    Micka, JA
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (01) : 1 - 9
  • [3] Calibration of multiple LDR brachytherapy sources
    DeWerd, Larry A.
    Micka, John A.
    Holmes, Shannon M.
    Bohm, Tim D.
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3804 - 3813
  • [4] Fuller Donald B, 2004, Brachytherapy, V3, P10, DOI 10.1016/j.brachy.2004.02.003
  • [5] *LOS AL NAT LAB, 2003, LAUR031019 LOS AL NA
  • [6] Feasibility of calibrating elongated brachytherapy sources using a well-type ionization chamber
    Meigooni, Ali S.
    Awan, Shahid B.
    Dou, Kai
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4184 - 4189
  • [7] Dosimetric characteristics of the new RadioCoil™ 103Pd wire line source for use in permanent brachytherapy implants
    Meigooni, AS
    Zhang, H
    Clark, JR
    Rachabatthula, V
    Koona, RA
    [J]. MEDICAL PHYSICS, 2004, 31 (11) : 3095 - 3105
  • [8] Seed fixity in the prostate/periprostatic region following brachytherapy
    Merrick, GS
    Butler, WM
    Dorsey, AT
    Lief, JH
    Benson, ML
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (01): : 215 - 220
  • [9] TRANSPERINEAL PALLADIUM-103 PROSTATE BRACHYTHERAPY - ANALYSIS OF MORBIDITY AND SEED MIGRATION
    NAG, S
    SCAPEROTH, DD
    BADALAMENT, R
    HALL, SA
    BURGERS, J
    [J]. UROLOGY, 1995, 45 (01) : 87 - 92
  • [10] Radioactive implant migration in patients treated for localized prostate cancer with interstitial brachytherapy
    Older, RA
    Synder, B
    Krupski, TL
    Glembocki, DJ
    Gillenwater, JY
    [J]. JOURNAL OF UROLOGY, 2001, 165 (05) : 1590 - 1592