Comparative study on AISI 440 and AISI 420B stainless steel for dental drill performance

被引:13
作者
Albertini, M. [2 ]
Herrero-Climent, M. [3 ]
Lazaro, P. [3 ]
Rios, J. V. [3 ]
Gil, F. J. [1 ]
机构
[1] Univ Politecn Cataluna, Dept Mat Sci & Met Engn, ETSEIB, E-08028 Barcelona, Spain
[2] Catedra UPC Klockner, Inst Padros, Barcelona, Spain
[3] Fac Odontol Sevilla, Seville, Spain
关键词
Stainless steel; Drill; Biomaterials; Wear; TITANIUM IMPLANTS; BONE;
D O I
10.1016/j.matlet.2012.04.006
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
One of the most usual causes of the failure in dental implant osseointegration is the increase of the temperature above 47 degrees C during the drilling of the bone because it produces irreversible osteonecrosis. The result is weakened contact of implants with bone and possible loss of rigid fixation. Many studies about the design of the drill bit, contact forces, drill speed, drill point angle and feed-rate have been made. However, in all studies a martensitic stainless steel (AISI 440) has been used. The aim of this study was to determine the behavior of AISI420B stainless steel with higher hardness, higher heat conductivity and lower wear than the conventional stainless steel used in oral implantology. This steel showed adequate temperature increases, with or without irrigation, and a decrease of the weight lost of the drills with use. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:163 / 165
页数:3
相关论文
共 13 条
[1]  
Abouzgia MB, 1984, J MAXILLOFAC SURG, V53, P1308
[2]   Interspecies differences in bone composition, density, and quality:: Potential implications for in vivo bone research [J].
Aerssens, J ;
Boonen, S ;
Lowet, G ;
Dequeker, J .
ENDOCRINOLOGY, 1998, 139 (02) :663-670
[3]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[4]   Osseointegration of grit-blasted and bioactive titanium implants:: Histomorphometry in minipigs [J].
Aparicio, C ;
Gil, FJ ;
Thams, U ;
Muñoz, F ;
Padrós, A ;
Planell, JA .
BIOCERAMICS 16, 2004, 254-2 :737-740
[5]   Human-osteoblast proliferation and differentiation on grit-blasted and bioactive titanium for dental applications [J].
Aparicio, C ;
Gil, FJ ;
Planell, JA ;
Engel, E .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2002, 13 (12) :1105-1111
[6]   Thermal osteonecrosis and bone drilling parameters revisited [J].
Augustin, Goran ;
Davila, Slavko ;
Mihoci, Kristijan ;
Udiljak, Toma ;
Vedrina, Denis Stjepan ;
Antabak, Anko .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (01) :71-77
[7]  
Bonfield W, 1972, J BIOMECH, V1, P323
[8]   INFLUENCE OF SURFACE CHARACTERISTICS ON BONE INTEGRATION OF TITANIUM IMPLANTS - A HISTOMORPHOMETRIC STUDY IN MINIATURE PIGS [J].
BUSER, D ;
SCHENK, RK ;
STEINEMANN, S ;
FIORELLINI, JP ;
FOX, CH ;
STICH, H .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :889-902
[9]  
Cordioli G, 1997, INT J ORAL MAX IMPL, V12, P186
[10]  
Harris BH, 2001, INT J ORAL MAX IMPL, V16, P819