New information on the characteristics of 1 in. x 1 in. cerium bromide scintillation detectors

被引:47
作者
Billnert, R. [1 ,2 ]
Oberstedt, S. [1 ]
Andreotti, E. [1 ]
Hult, M. [1 ]
Marissens, G. [1 ]
Oberstedt, A. [2 ,3 ]
机构
[1] European Commiss, DG JRC, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
[2] Chalmers Tekniska Hgsk, S-41296 Gothenburg, Sweden
[3] Univ Orebro, Akad Nat Vetenskap Tekn, S-70182 Orebro, Sweden
关键词
Prompt fission gamma-rays; Cerium bromide; Lanthanum chloride; Scintillation detector; Energy resolution; Intrinsic activity; Timing resolution; NAI(TL) SCINTILLATORS; X-RAY; LACL3-CE; LABR3-CE;
D O I
10.1016/j.nima.2011.05.034
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In view of highly demanded new and accurate data on prompt gamma-ray emission in nuclear fission a major part of investigations is directed towards the selection of suitable detector systems. Here we have studied a new type of crystal scintillation detectors made from cerium bromide (CeBr(3)). For the first time a full characterization of such a detector is presented in terms of energy resolution, pulse-height linearity, intrinsic activity and intrinsic timing resolution. In particular the latter one is very important for prompt fission gamma-ray studies, because the presence of fast neutrons, emitted in fission too, requires the time-of-flight method for their discrimination. The energy resolution has been found to be comparable to that of cerium-doped LaCl(3):Ce detectors at an efficiency comparable to the one of a LaBr(3):Ce detector of the same size. The intrinsic activity of the CeBr(3) crystal was observed to be much lower compared to lanthanum halide crystals. The intrinsic timing resolution of a coaxial 1 in. x 1 in. sized detector was measured relative to that of a previously characterized LaCl(3):Ce detector and found to be (326 +/- 7) Ps at (60)Co energies, which is in between those of a LaBr(3):Ce and a LaCl(3):Ce detector of same size. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:94 / 99
页数:6
相关论文
共 21 条
[1]  
ANDREOTTI E, P 3 INT C CURR PROBL
[2]  
[Anonymous], NUCL DATA HIGH PRIOR
[3]  
[Anonymous], GENP0120080113 JRC
[4]   Determination of absolute detection efficiencies for detectors of interest in homeland security [J].
Ayaz-Malerhafer, Birsen ;
DeVol, Timothy A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01) :410-413
[5]   Comparison of LaCl3:Ce and NaI(Tl) scintillators in γ-ray spectrometry [J].
Balcerzyk, M ;
Moszynski, M ;
Kapusta, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 537 (1-2) :50-56
[6]   Measurements of high-energy γ-rays with LaBr3 : Ce detectors [J].
Ciemala, M. ;
Balabanski, D. ;
Csatlos, M. ;
Daugas, J. M. ;
Georgiev, G. ;
Gulyas, J. ;
Kmiecik, M. ;
Krasznahorkay, A. ;
Lalkovski, S. ;
Lefebvre-Schuhl, A. ;
Lozeva, R. ;
Maj, A. ;
Vitez, A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (01) :76-79
[7]   Wide energy range efficiency calibration for a lanthanum bromide scintillation detector [J].
Favalli, A. ;
Mehner, H. -C. ;
Simonelli, F. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :506-509
[8]  
GILMORE G, 1978, PRACTICAL GAMMARAY S
[9]   CeBr3 as a room-temperature, high-resolution gamma-ray detector [J].
Guss, Paul ;
Reed, Michael ;
Yuan, Ding ;
Reed, Alexis ;
Mukhopadhyay, Sanjoy .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (02) :297-304
[10]   Observations on the background spectra of four LaCl3(Ce) scintillation detectors [J].
Hartwell, JK ;
Gehrke, RJ .
APPLIED RADIATION AND ISOTOPES, 2005, 63 (02) :223-228