Test-retest reliability of the GAITRite system in people with stroke undergoing rehabilitation

被引:60
作者
Kuys, Suzanne S. [1 ,2 ,3 ]
Brauer, Sandra G. [2 ]
Ada, Louise [4 ]
机构
[1] Prince Charles Hosp, Allied Hlth Res Collaborat, Metro N No Cluster Hlth Serv, Chermside, Qld 4032, Australia
[2] Univ Queensland, Div Physiotherapy, Brisbane, Qld, Australia
[3] Griffith Univ, Sch Physiotherapy & Exercise Sci, Nathan, Qld 4111, Australia
[4] Univ Sydney, Discipline Physiotherapy, Sydney, NSW 2006, Australia
基金
澳大利亚国家健康与医学研究理事会;
关键词
Gait; stroke; GAITRite; measurement; reliability; OUTCOME MEASURES; WALKWAY SYSTEM; PARAMETERS; WALKING; IMPAIRMENTS; VALIDITY;
D O I
10.3109/09638288.2010.549895
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Purpose. To determine the test-retest reliability of the GAITRite system in measuring spatio-temporal gait parameters in people with stroke undergoing inpatient rehabilitation and to investigate whether walking ability influences test-retest reliability. Method. A within-subject repeated measures study was conducted. Twenty-one individuals with first stroke undergoing inpatient rehabilitation participated. The Motor Assessment Scale Item 5 score was used to categorise participants into poorer (score 3 or 4) or better walkers (score 5 or 6). Participants made two passes at a self-selected gait speed over the GAITRite mat on two separate occasions. Intra-class correlation coefficients (ICC(2,1)) and 95% limits of agreement for spatio-temporal gait parameters between the two occasions and groups were determined. Results. No differences in gait parameters between the two sessions (p>0.15) were found. Test-retest reliability for the whole group was good to excellent, ICC(2,1) ranging from 0.72 to 0.94. Walking ability did influence these findings with fair to good reliability demonstrated for both poorer (ICC(2,1) > 0.57) and better walkers (ICC(2,1) > 0.41). Conclusions. It appears that the GAITRite has good test-retest reliability in this group of people with first stroke undergoing rehabilitation. Test reliability seems to be slightly higher for poorer walkers than better walkers.
引用
收藏
页码:1848 / 1853
页数:6
相关论文
共 26 条
[1]   Variability in spatiotemporal step characteristics and its relationship to walking performance post-stroke [J].
Balasubramanian, Chitralakshmi K. ;
Neptune, Richard R. ;
Kautz, Steven A. .
GAIT & POSTURE, 2009, 29 (03) :408-414
[2]   Concurrent related validity of the GAITRite® walkway system for quantification of the spatial and temporal parameters of gait [J].
Bilney, B ;
Morris, M ;
Webster, K .
GAIT & POSTURE, 2003, 17 (01) :68-74
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Comfortable and maximum walking speed of adults aged 20-79 years: Reference values and determinants [J].
Bohannon, RW .
AGE AND AGEING, 1997, 26 (01) :15-19
[5]   INVESTIGATION OF A NEW MOTOR-ASSESSMENT SCALE FOR STROKE PATIENTS [J].
CARR, JH ;
SHEPHERD, RB ;
NORDHOLM, L ;
LYNNE, D .
PHYSICAL THERAPY, 1985, 65 (02) :175-180
[6]  
Dean C, 1992, Aust J Physiother, V38, P31, DOI 10.1016/S0004-9514(14)60548-1
[7]   MINI-MENTAL STATE - PRACTICAL METHOD FOR GRADING COGNITIVE STATE OF PATIENTS FOR CLINICIAN [J].
FOLSTEIN, MF ;
FOLSTEIN, SE ;
MCHUGH, PR .
JOURNAL OF PSYCHIATRIC RESEARCH, 1975, 12 (03) :189-198
[8]   Mobility on discharge from an aged care unit [J].
Gorgon, Edward ;
Said, Catherine ;
Galea, Mary .
PHYSIOTHERAPY RESEARCH INTERNATIONAL, 2007, 12 (02) :72-81
[9]   Hemiparetic gait parameters in overground versus treadmill walking [J].
Harris-Love, ML ;
Forrester, LW ;
Macko, RF ;
Silver, KHC ;
Smith, GV .
NEUROREHABILITATION AND NEURAL REPAIR, 2001, 15 (02) :105-112
[10]  
Hill Keith, 1997, Aust J Physiother, V43, P173