Magnetic Resonance Imaging Predicts Adverse Local Tissue Reaction Histologic Severity in Modular Neck Total Hip Arthroplasty

被引:10
作者
Barlow, Brian T. [1 ]
Ortiz, Philippe A. [1 ]
Fields, Kara G. [2 ]
Burge, Alissa J. [3 ]
Potter, Hollis G. [3 ]
Westrich, Geoffrey H. [1 ]
机构
[1] Hosp Special Surg, Dept Orthopaed Surg, Div Adult Reconstruct & Joint Replacement Surg, 535 E 70th St, New York, NY 10021 USA
[2] Hosp Special Surg, Epidemiol & Biostat Core, 535 E 70th St, New York, NY 10021 USA
[3] Hosp Special Surg, Dept Radiol & Imaging, 535 E 70th St, New York, NY 10021 USA
关键词
MRI; modular; revision; ALTR; synovial; histology; CORROSION; TAPER; JUNCTION; FAILURE; STEM;
D O I
10.1016/j.arth.2016.03.022
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The association between advanced imaging, serum metal ion levels, and histologic adverse local tissue reaction (ALTR) severity has not been previously reported for Rejuvenate modular neck femoral stems. Methods: A cohort of 90 patients with 98 Rejuvenate modular neck femoral stems was revised by a single surgeon from July 2011 to December 2014. Before revision, patients underwent multiacquisition variable resonance image combination sequence magnetic resonance imaging (MRI), and serum cobalt and chromium ion levels were measured. Histologic samples from the revision surgery were scored for synovial lining, inflammatory infiltrate, and tissue organization as proposed by Campbell. Regression based on the generalized estimating equations approach was used to assess the univariate association between each MRI, demographic, and metal ion measure and ALTR severity while accounting for the correlation between bilateral hips. Random forest analysis was then used to determine the relative importance of MRI characteristics, demographics, and metal ion levels in predicting ALTR severity. Results: Synovial thickness as measured on MRI was found to be the strongest predictor of ALTR histologic severity in a recalled modular neck femoral stem. Conclusion: MRI can accurately describe ALTR in modular femoral neck total hip arthroplasty. MRI characteristics, particularly maximal synovial thickness and synovitis volume, predicted histologic severity. Serum metal ion levels do not correlate with histologic severity in Rejuvenate modular neck total hip arthroplasty. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:2325 / 2331
页数:7
相关论文
共 20 条
  • [1] Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system
    Anderson, Helen
    Toms, Andoni Paul
    Cahir, John G.
    Goodwin, Richard W.
    Wimhurst, James
    Nolan, John F.
    [J]. SKELETAL RADIOLOGY, 2011, 40 (03) : 303 - 307
  • [2] Barlow BT, 2015, J ARTHROPLASTY
  • [3] Random forests
    Breiman, L
    [J]. MACHINE LEARNING, 2001, 45 (01) : 5 - 32
  • [4] Burge AJ, 2015, RADIOLOGY
  • [5] Histological Features of Pseudotumor-like Tissues From Metal-on-Metal Hips
    Campbell, Pat
    Ebramzadeh, Edward
    Nelson, Scott
    Takamura, Karren
    De Smet, Koen
    Amstutz, Harlan C.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (09) : 2321 - 2327
  • [6] Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck
    Cooper, H. John
    Urban, Robert M.
    Wixson, Richard L.
    Meneghini, R. Michael
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) : 865 - 872
  • [7] Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty
    Cooper, H. John
    Della Valle, Craig J.
    Berger, Richard A.
    Tetreault, Matthew
    Paprosky, Wayne G.
    Sporer, Scott M.
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) : 1655 - 1661
  • [8] Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation
    Gill, I. P. S.
    Webb, J.
    Sloan, K.
    Beaver, R. J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (07): : 895 - 900
  • [9] Guilfoyle J, 2012, STRYKER INITIATES VO
  • [10] MRI Findings in Painful Metal-on-Metal Hip Arthroplasty
    Hayter, Catherine L.
    Gold, Stephanie L.
    Koff, Matthew F.
    Perino, Giorgio
    Nawabi, Danyal H.
    Miller, Theodore T.
    Potter, Hollis G.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (04) : 884 - 893