Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses

被引:17
作者
Kleinmann, G
Marcovich, AL
Apple, DJ
Mamalis, N
机构
[1] Univ Utah, Dept Ophthalmol & Visual Sci, A Moran Eye Ctr, Labs Ophthalm Devices Res, Salt Lake City, UT 84132 USA
[2] Kaplan Med Ctr, Dept Ophthalmol, Rehovot, Israel
关键词
D O I
10.1136/bjo.2005.071738
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To describe unique linear deposits on the surface of posterior chamber intraocular lenses (IOLs) occurring after implantation through a hexagonal cartridge. Methods: Five ACR6D SE IOLs ( Corneal Laboratories, Pringy, France) were injected/ implanted through hexagonal cartridges. Two of these were injected into a petri dish and the remaining three were inserted into the evacuated capsular bag of cadaver eyes. In addition, three other ACR6D SE IOLs were injected into a petri dish through round cartridges. The latter served as controls. Results: All lenses that were injected/ implanted through the hexagonal cartridges demonstrated linear deposits on the posterior surface of the IOL optic component. The IOLs that were injected through the round cartridges had no deposits. All the hexagonal cartridges showed signs of internal cracking. None of the round cartridges were cracked. Conclusion: Implantation of at least one IOL model, the ACR6D SE IOL, through a hexagonal cartridge can result in linear deposits on the posterior optical surface of the IOL. The shape of the cartridge appears to be a significant factor in causing the depositions.
引用
收藏
页码:1474 / 1477
页数:4
相关论文
共 12 条
[1]  
Carlson KH, 1995, OPHTHALMIC SURG LAS, V26, P572
[2]  
Dick HB, 1999, J CATARACT REFR SURG, V25, P463
[4]   Surface abnormalities on hydrophilic acrylic intraocular lenses implanted by an injector [J].
Faschinger, CW .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (06) :845-849
[5]   Cracked cartridges during foldable intraocular lens implantation [J].
Habib, NE ;
Singh, J ;
Adams, AD ;
Bartholomew, RS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (05) :630-632
[6]   Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses [J].
Kohnen, T ;
Magdowski, G ;
Koch, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1342-1350
[7]   Alterations to the AcrySof intraocular lens during folding [J].
Milazzo, S ;
Turut, P ;
Blin, H .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1351-1354
[8]   Cartridge cracks with different viscoelastic agents [J].
Olson, RJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (04) :465-465
[9]   Stress fractures after folding an acrylic intraocular lens [J].
Pfister, DR .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 121 (05) :572-574
[10]   Cartridge cracks during foldable intraocular lens insertion [J].
Rath, R ;
Singh, AD ;
Singh, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1999, 25 (03) :306-306