Utilization of steerable sheath improves the efficiency of atrial fibrillation ablation guided by robotic magnetic navigation compared with fixed-curve sheath

被引:6
作者
Luo, Qingzhi [1 ]
Xie, Yun [1 ]
Bao, Yangyang [1 ]
Wei, Yue [1 ]
Lin, Changjian [1 ]
Zhang, Ning [1 ]
Ling, Tianyou [1 ]
Chen, Kang [1 ]
Pan, Wenqi [1 ]
Wu, Liqun [1 ]
Jin, Qi [1 ]
机构
[1] Shanghai Jiao Tong Univ, Ruijin Hosp, Dept Cardiovasc Med, Sch Med, Shanghai, Peoples R China
关键词
atrial fibrillation; catheter ablation; magnetic navigation; steerable sheath; CATHETER ABLATION; RADIOFREQUENCY ABLATION; EXPERIENCE; EFFICACY; SAFETY; SYSTEM;
D O I
10.1002/clc.23801
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The objective of this study was to assess the impact of steerable sheaths compared with fixed-curve sheaths on the procedural outcomes of atrial fibrillation (AF) ablation guided by robotic magnetic navigation (RMN). Methods and Results In this retrospective case-control study, 110 patients scheduled for AF catheter ablation were enrolled and divided into two groups. Fifty-five patients (paroxysmal, 70%) were treated with RMN-guided ablation utilizing a steerable sheath and another 55 patients (paroxysmal, 70%) were ablated with RMN using a fixed-curve sheath. Clinical characteristics were similar between the two groups. Compared with the fixed-curve sheath group, the steerable sheath group procedure time (111.9 +/- 25.2 vs. 90.4 +/- 20.7 min, p < .001) and radiofrequency (RF) time (35.9 +/- 9.0 vs. 30.5 +/- 7.4 min, p < .001) were significantly shortened. Additionally, the navigation index was significantly improved (0.41 +/- 0.06 vs. 0.48 +/- 0.08, p < .001) in the steerable sheath group. By employing a large catheter loop for targeting the right pulmonary veins (PVs), the steerable sheath group significantly reduced the RF delivery time (15.0 +/- 3.0 vs. 12.0 +/- 2.1 min, p < .001) during right-side PV isolation (PVI). However, total fluoroscopy time was similar between the two groups (5.6 +/- 2.6 vs. 5.0 +/- 2.0 min, p > .05). Acute PVI success rates were similar between the two groups. No major or minor complications occurred in either group. Conclusion Appropriate utilization of steerable sheath technology can improve the efficiency of AF ablation guided by RMN, primarily by reducing the total procedure and RF delivery times of right-side PVI without compromising safety.
引用
收藏
页码:482 / 487
页数:6
相关论文
共 19 条
[1]   Robotic navigation for catheter ablation: benefits and challenges [J].
Aagaard, Philip ;
Natale, Andrea ;
Di Biase, Luigi .
EXPERT REVIEW OF MEDICAL DEVICES, 2015, 12 (04) :457-469
[2]   Contact force with magnetic-guided catheter ablation [J].
Bessiere, Francis ;
Zikry, Christopher ;
Rivard, Lena ;
Dyrda, Katia ;
Khairy, Paul .
EUROPACE, 2018, 20 :1-4
[3]   Magnetic guidance versus manual control: comparison of radiofrequency lesion dimensions and evaluation of the effect of heart wall motion in a myocardial phantom [J].
Bhaskaran, Abhishek ;
Barry, M. A. ;
Al Raisi, Sara I. ;
Chik, William ;
Doan Trang Nguyen ;
Pouliopoulos, Jim ;
Nalliah, Chrishan ;
Hendricks, Roger ;
Thomas, Stuart ;
McEwan, Alistair L. ;
Kovoor, Pramesh ;
Thiagalingam, Aravinda .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2015, 44 (01) :1-8
[4]   Catheter Ablation Utilizing Remote Magnetic Navigation: A Review of Applications and Outcomes [J].
Bradfield, Jason ;
Tung, Roderick ;
Mandapati, Ravi ;
Boyle, Noel G. ;
Shivkumar, Kalyanam .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (08) :1021-1034
[5]   Radiofrequency catheter ablation of atrial fibrillation: Electrical modification suggesting transmurality is faster achieved with remote magnetic catheter in comparison with contact force use [J].
Bun, Sok-Sithikun ;
Ayari, Anis ;
Latcu, Decebal Gabriel ;
Errahmouni, Abdelkarim ;
Saoudi, Nadir .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (07) :745-753
[6]   Substantial superiority of Niobe ES over Niobe II system in remote-controlled magnetic pulmonary vein isolation [J].
Da Costa, Antoine ;
Guichard, Jean Baptiste ;
Maillard, Nicolas ;
Romeyer-Bouchard, Cecile ;
Gerbay, Antoine ;
Isaaz, Karl .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 230 :319-323
[7]   The impact of steerable sheaths on unblinded contact force during catheter ablation for atrial fibrillation [J].
Deyell, Marc W. ;
Wen, Glen ;
Laksman, Zachary ;
Bennett, Matthew T. ;
Chakrabarti, Santabhanu ;
Yeung-Lai-Wah, John A. ;
Krahn, Andrew D. ;
Andrade, Jason G. .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 57 (03) :417-424
[8]   Remote magnetic navigation versus manual control navigation for atrial fibrillation ablation: A systematic review and meta-analysis [J].
Jia, Kangni ;
Jin, Qi ;
Liu, Ao ;
Wu, Liqun .
JOURNAL OF ELECTROCARDIOLOGY, 2019, 55 :78-86
[9]   Efficacy and Safety of Atrial Fibrillation Ablation Using Remote Magnetic Navigation: Experience from 1,006 Procedures [J].
Jin, Qi ;
Pehrson, Steen ;
Jacobsen, Peter Karl ;
Chen, Xu .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2016, 27 :S23-S28
[10]  
Kirchhof P, 2016, EUROPACE, V18, DOI [10.1093/europace/euw295, 10.1016/j.rec.2016.11.033, 10.5603/KP.2016.0172]