The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations

被引:78
作者
van Etmpt, Wouter [1 ]
Nijsten, Sebastiaan [1 ]
Mijnheer, Ben [1 ]
Dekker, Andre [1 ]
Lambin, Philippe [1 ]
机构
[1] Univ Hosp Maastricht, Dept Radiat Oncol MAASTRO, GROW, NL-6229 ET Maastricht, Netherlands
关键词
patient-specific QA; EPID dosimetry; Monte Carlo dose calculation; dose verification;
D O I
10.1016/j.radonc.2007.11.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: A method was evaluated to reconstruct the 3D dose distribution in patients using their planning CT-scan in combination with a Monte Carlo calculation, and the energy fluence of the actual treatment beams measured pre-treatment with an EPID without the patient or a phantom in the beam. Materials and methods: Nine plans of lung cancer patients treated with a 3D conformal technique, calculated using a simple convolution algorithm (CA), as well as five IMRT treatments of head-and-neck cancer patients, calculated with a more advanced superposition algorithm (SA), were verified. Differences between planned and reconstructed dose distributions were quantified in terms of DVH parameters. Results: For the lung cancer group, differences between the reconstructed mean PTV dose and the values calculated with the TPS were 5.0 +/- 4.2% (1SD) and -1.4 +/- 1.5% for the CA and SA algorithm, respectively. No large differences in the lung and spinal cord DVH parameters were found. For the IMRT treatments, the average dose differences in the PTV were generally below 3%. The reconstructed mean parotid gland dose was 3.2 +/- 1.2% lower, while the maximum spinal cord dose was on average 3.1 +/- 1.9% higher. Conclusions: EPID dosimetry combined with 3D dose reconstruction is a useful procedure for patient-specific QA of complex treatments. DVH parameters can be used to interpret the dose distribution delivered to the patient in the same way as during standard treatment plan evaluation. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:86 / 92
页数:7
相关论文
共 23 条
[11]   Routine individualised patient dosimetry using electronic portal imaging devices [J].
Nijsten, Sebastiaan M. J. J. G. ;
Mijnheer, Ben J. ;
Dekker, Andre L. A. J. ;
Lambin, Philippe ;
Minken, Andre W. H. .
RADIOTHERAPY AND ONCOLOGY, 2007, 83 (01) :65-75
[12]   A dose delivery verification method for conventional and intensity-modulated radiation therapy using measured field fluence distributions [J].
Renner, WD ;
Sarfaraz, M ;
Earl, MA ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (11) :2996-3005
[13]   Converting absorbed dose to medium to absorbed dose to water for Monte Carlo based photon beam dose calculations [J].
Siebers, JV ;
Keall, PJ ;
Nahum, AE ;
Mohan, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (04) :983-995
[14]   Three-dimensional IMRT verification with a flat-panel EPID [J].
Steciw, S ;
Warkentin, B ;
Rathee, S ;
Fallone, BG .
MEDICAL PHYSICS, 2005, 32 (02) :600-612
[15]   Experimental verification of a portal dose prediction model [J].
van Elmpt, WJC ;
Nijsten, SMJJG ;
Mijnheer, BJ ;
Minken, AWH .
MEDICAL PHYSICS, 2005, 32 (09) :2805-2818
[16]   Treatment verification in the presence of inhomogeneities using EPID-based three-dimensional dose reconstruction [J].
van Elmpt, Wouter J. C. ;
Nijsten, Sebastiaan M. J. J. G. ;
Dekker, Andre L. A. J. ;
Mijnheer, Ben J. ;
Lambin, Philippe .
MEDICAL PHYSICS, 2007, 34 (07) :2816-2826
[17]   A Monte Carlo based three-dimensional dose reconstruction method derived from portal dose images [J].
van Elmpt, Wouter J. C. ;
Nijsten, Sebastiaan M. J. J. G. ;
Schiffeleers, Robert F. H. ;
Dekker, Andre L. A. J. ;
Mijnheer, Ben J. ;
Lambin, Philippe ;
Minken, Andre W. H. .
MEDICAL PHYSICS, 2006, 33 (07) :2426-2434
[18]   The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields [J].
Van Esch, A ;
Depuydt, T ;
Huyskens, DP .
RADIOTHERAPY AND ONCOLOGY, 2004, 71 (02) :223-234
[19]   3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID [J].
van Zijtveld, Mathilda ;
Dirkx, Maarten L. P. ;
de Boer, Hans C. J. ;
Heijmen, Ben J. M. .
RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) :201-207
[20]   Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience [J].
van Zijtveld, Mathilda ;
Dirkx, Maarten L. P. ;
de Boer, Hans C. J. ;
Heijmen, Ben J. M. .
RADIOTHERAPY AND ONCOLOGY, 2006, 81 (02) :168-175