The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations

被引:78
作者
van Etmpt, Wouter [1 ]
Nijsten, Sebastiaan [1 ]
Mijnheer, Ben [1 ]
Dekker, Andre [1 ]
Lambin, Philippe [1 ]
机构
[1] Univ Hosp Maastricht, Dept Radiat Oncol MAASTRO, GROW, NL-6229 ET Maastricht, Netherlands
关键词
patient-specific QA; EPID dosimetry; Monte Carlo dose calculation; dose verification;
D O I
10.1016/j.radonc.2007.11.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: A method was evaluated to reconstruct the 3D dose distribution in patients using their planning CT-scan in combination with a Monte Carlo calculation, and the energy fluence of the actual treatment beams measured pre-treatment with an EPID without the patient or a phantom in the beam. Materials and methods: Nine plans of lung cancer patients treated with a 3D conformal technique, calculated using a simple convolution algorithm (CA), as well as five IMRT treatments of head-and-neck cancer patients, calculated with a more advanced superposition algorithm (SA), were verified. Differences between planned and reconstructed dose distributions were quantified in terms of DVH parameters. Results: For the lung cancer group, differences between the reconstructed mean PTV dose and the values calculated with the TPS were 5.0 +/- 4.2% (1SD) and -1.4 +/- 1.5% for the CA and SA algorithm, respectively. No large differences in the lung and spinal cord DVH parameters were found. For the IMRT treatments, the average dose differences in the PTV were generally below 3%. The reconstructed mean parotid gland dose was 3.2 +/- 1.2% lower, while the maximum spinal cord dose was on average 3.1 +/- 1.9% higher. Conclusions: EPID dosimetry combined with 3D dose reconstruction is a useful procedure for patient-specific QA of complex treatments. DVH parameters can be used to interpret the dose distribution delivered to the patient in the same way as during standard treatment plan evaluation. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:86 / 92
页数:7
相关论文
共 23 条
[1]   Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm [J].
Fippel, M .
MEDICAL PHYSICS, 1999, 26 (08) :1466-1475
[2]   On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities:: comparison with Monte Carlo calculations [J].
Fogliata, Antonella ;
Vanetti, Eugenio ;
Albers, Dirk ;
Brink, Carsten ;
Clivio, Alessandro ;
Knoos, Tommy ;
Nicolini, Giorgia ;
Cozzi, Luca .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (05) :1363-1385
[3]   An inter-centre quality assurance network for IMRT verification: Results of the ESTRO QUASIMODO project [J].
Gillis, S ;
De Wagter, C ;
Bohsung, J ;
Perrin, B ;
Williams, P ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (03) :340-353
[4]   Patient-specific dosimetry of conventional and intensity modulated radiation therapy using a novel full Monte Carlo phase space reconstruction method from electronic portal images [J].
Jarry, Genevieve ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (08) :2277-2299
[5]   On the accuracy and effectiveness of dose reconstruction for tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Balog, JP ;
Keller, H ;
Reckwerdt, PJ ;
Mackie, TR .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) :943-966
[6]   Using fluence separation to account for energy spectra dependence in computing dosimetric a-Si EPID images for IMRT fields [J].
Li, Weidong ;
Siebers, Jeffrey V. ;
Moore, Joseph A. .
MEDICAL PHYSICS, 2006, 33 (12) :4468-4480
[7]   Three-dimensional dose reconstruction of breast cancer treatment using portal imaging [J].
Louwe, RJW ;
Damen, EMF ;
van Herk, M ;
Minken, AWH ;
Törzsök, O ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2003, 30 (09) :2376-2389
[8]   Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification [J].
McDermott, L. N. ;
Wendling, M. ;
van Asselen, B. ;
Stroom, J. ;
Sonke, J. -J. ;
van Herk, M. ;
Mijnheer, B. J. .
MEDICAL PHYSICS, 2006, 33 (10) :3921-3930
[9]   Implementation of FFT convolution and multigrid superposition models in the FOCUS RTP system [J].
Miften, M ;
Wiesmeyer, M ;
Monthofer, S ;
Krippner, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (04) :817-833
[10]   A global calibration model for a-Si EPIDs used for transit dosimetry [J].
Nijsten, S. M. J. J. G. ;
van Elmpt, W. J. C. ;
Jacobs, M. ;
Mijnheer, B. J. ;
Dekker, A. L. A. J. ;
Lambin, P. ;
Minken, A. W. H. .
MEDICAL PHYSICS, 2007, 34 (10) :3872-3884