Radiation Dose and Image Quality of a High-Pitch Prospective Spiral First Approach in Coronary Computed Tomography Angiography (CCTA)

被引:5
作者
Finck, Tom [1 ]
Klambauer, Konstantin [1 ]
Hendrich, Eva [1 ]
Will, Albrecht [1 ]
Martinoff, Stefan [1 ]
Hadamitzky, Martin [1 ]
机构
[1] German Heart Ctr Munich, Inst Radiol & Nucl Med, Lazarettstr 36, D-80636 Munich, Germany
关键词
coronary computed tomography angiography; high-pitch-spiral scan; radiation exposure; coronary artery disease; DUAL-SOURCE CT; ACQUISITION; PROTECTION; ACCURACY;
D O I
10.3390/jcdd8100119
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To investigate a high-pitch spiral first (HPSF) approach for coronary computed tomography angiography (CCTA) in an unselected patient cohort and compare diagnostic yield and radiation exposure to CCTAs acquired via conventional, non-high-pitch spiral first (NHPSF) scan regimes. Materials and Methods: All consecutive patients from 1 January 2015 to 31 December 2017 were included. Two investigation protocols (HPSF/NHPSF) were used with the aim to achieve diagnostic image quality of all coronary segments. Low-pitch secondary scans followed the initial examination if image quality was unsatisfactory. Dosage and image quality were compared between both regimes. Results: 1410 patients were subject to a HPSF and 236 patients to a NHPSF approach. While the HPSF approach led to a higher fraction of re-scans (35% vs. 11%, p < 0.001), the fraction of aggregate scans that remained non-diagnostic after considering the initial and secondary scan was comparably low for the HPSF and NHPSF approach (0.78 vs. 0%, p = 0.18). Aggregate radiation exposure in the HPSF protocol was significantly lower (1.12 mSv (IQR: 0.73, 2.10) vs. 3.96 mSv (IQR: 2.23, 8.33) p < 0.001). Conclusions: In spite of a higher number of re-scans, a HPSF approach leads to a reduction in overall radiation exposure with diagnostic yields similar to a NHPSF approach.
引用
收藏
页数:13
相关论文
共 25 条
[11]   Radiation dose and diagnostic accuracy of high-pitch dual-source coronary angiography in the evaluation of coronary artery stenoses [J].
Koplay, M. ;
Erdogan, H. ;
Avci, A. ;
Sivri, M. ;
Demir, K. ;
Guler, I. ;
Demir, L. S. ;
Paksoy, Y. .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2016, 97 (04) :461-469
[12]   Prospectively ECG-triggered High-pitch Spiral Acquisition for Cardiac CT Angiography in Routine Clinical Practice Initial Results [J].
Kroepil, Patric ;
Rojas, Carlos A. ;
Ghoshhajra, Brian ;
Lanzman, Rotem S. ;
Miese, Falk R. ;
Scherer, Axel ;
Kalra, Mannudeep ;
Abbara, Suhny .
JOURNAL OF THORACIC IMAGING, 2012, 27 (03) :194-201
[13]   Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques - A comparative study with intracoronary ultrasound [J].
Leber, AW ;
Knez, A ;
Becker, A ;
Becker, C ;
von Ziegler, F ;
Nikolaou, K ;
Rist, C ;
Reiser, M ;
White, C ;
Steinbeck, G ;
Boekstegers, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (07) :1241-1247
[14]   Comparison of radiation doses using weight-based protocol and dose modulation techniques for patients undergoing biphasic abdominal computed tomography examinations [J].
Livingstone, Roshan S. ;
Dinakaran, Paul M. ;
Cherian, Rekha S. ;
Eapen, Anu .
JOURNAL OF MEDICAL PHYSICS, 2009, 34 (04) :217-222
[15]   128-slice dual-source CT coronary angiography with prospectively electrocardiography-triggered high-pitch spiral mode: radiation dose, image quality, and diagnostic acceptability [J].
Matsubara, Kosuke ;
Sakuda, Keita ;
Nunome, Haruka ;
Takata, Tadanori ;
Koshida, Kichiro ;
Gabata, Toshifumi .
ACTA RADIOLOGICA, 2016, 57 (01) :25-32
[16]   The diagnostic and prognostic value of coronary CT angiography in asymptomatic high-risk patients: a cohort study [J].
Plank, Fabian ;
Friedrich, Guy ;
Dichtl, Wolfgang ;
Klauser, Andrea ;
Jaschke, Werner ;
Franz, Wolfgang-Michael ;
Feuchtner, Gudrun .
OPEN HEART, 2014, 1 (01)
[17]   Comparison of sensitivity, specificity, positive predictive value, and negative predictive value of stress testing versus 64-multislice coronary computed tomography angiography in predicting obstructive coronary artery disease diagnosed by coronary angiography [J].
Ravipati, Gautham ;
Aronow, Wilbert S. ;
Lai, Hoang ;
Shao, John ;
DeLuca, Albert J. ;
Weiss, Melvin B. ;
Pucillo, Anthony L. ;
Kalapatapu, Kumar ;
Monsen, Craig E. ;
Belkin, Robert N. .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 101 (06) :774-775
[18]   Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance [J].
Scheffel, H. ;
Alkadhi, H. ;
Leschka, S. ;
Plass, A. ;
Desbiolles, L. ;
Guber, I. ;
Krauss, T. ;
Gruenenfelder, J. ;
Genoni, M. ;
Luescher, T. F. ;
Marincek, B. ;
Stolzmann, P. .
HEART, 2008, 94 (09) :1132-1137
[19]   Ultra-fast, low dose high-pitch (FLASH) versus prospectively-gated coronary computed tomography angiography: Comparison of image quality and patient radiation exposure [J].
Smettei, Osama A. ;
Sayed, Sawsan ;
Al Habib, Abdullah M. ;
Alharbi, Fahad ;
Abazid, Rami M. .
JOURNAL OF THE SAUDI HEART ASSOCIATION, 2018, 30 (03) :165-171
[20]   Low-dose, Prospective Triggered High-pitch Spiral Coronary Computed Tomography Angiography: Comparison with Retrospective Spiral Technique [J].
Srichai, Monvadi B. ;
Lim, Ruth P. ;
Donnino, Robert ;
Mannelli, Lorenzo ;
Hiralal, Rajesh ;
Avery, Ryan ;
Ho, Corey ;
Babb, James S. ;
Jacobs, Jill E. .