Dose-current discharge correlation analysis in a Mather type Plasma Focus device for medical applications

被引:7
作者
Sumini, M. [1 ,4 ]
Mostacci, D. [1 ]
Tartari, A. [4 ]
Mazza, A. [1 ]
Cucchi, G. [1 ]
Isolan, L. [1 ]
Buontempo, F. [3 ]
Zironi, I. [2 ,4 ]
Castellani, G. [2 ,4 ]
机构
[1] Univ Bologna, Ind Engn Dept, Via Colli 16, I-40136 Bologna, Italy
[2] Univ Bologna, Phys & Astron Dept, Via B Pichat 6-2, I-40126 Bologna, Italy
[3] Univ Bologna, Biomed Sci Dept, Via Irnerio 48, I-40126 Bologna, Italy
[4] INFN, Rome, Italy
关键词
Plasma focus; Signal analysis; Wigner-Ville transform; Dose rate; Medical applications; ELECTRON-BEAM;
D O I
10.1016/j.radphyschem.2017.03.022
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
In a Plasma Focus device the plasma collapses into the pinch where it reaches thermonuclear conditions for a few tens of nanoseconds, becoming a multi-radiation source. The nature of the radiation generated depends on the gas filling the chamber and the device working parameters. The self-collimated electron beam generated in the backward direction with respect to the plasma motion is one of the main radiation sources of interest also for medical applications. The electron beam may be guided against a high Z material target to produce an X-ray beam. This technique offers an ultra-high dose rate source of X-rays, able to deliver during the pinch a massive dose (up to 1 Gy per discharge for the PFMA-3 test device), as measured with EBT3 Gafchromicefilm tissue equivalent dosimeters. Given the stochastic behavior of the discharge process, a reliable on-line estimate of the dose-delivered is a very challenging task, in some way preventing a systematic application as a potentially interesting therapy device. This work presents an approach to linking the dose registered by the EBT3 Gafchromic films with the information contained in the signal recorded during the current discharge process. Processing the signal with the Wigner-Ville distribution, a spectrogram was obtained, displaying the information on intensity at various frequency scales, identifying the band of frequencies representative of the pinch events and define some patterns correlated with the dose.
引用
收藏
页码:452 / 457
页数:6
相关论文
共 14 条
[1]  
[Anonymous], THESIS
[2]  
[Anonymous], 2007, THESIS
[3]  
[Anonymous], P 5 INT C FRONT PLAS
[4]  
Bracewell R. N., 1986, The Fourier Transform and its Applications
[5]   Monte Carlo simulation of the dose deposited into GAFCHROMIC® film by X-rays produced by the electron beam of PFMA-3 [J].
Ceccolini, E. ;
Rocchi, F. ;
Mostacci, D. ;
Sumini, M. ;
Tartani, A. .
RADIATION MEASUREMENTS, 2012, 47 (10) :1000-1004
[6]   EBT2 dosimetry of x-rays produced by the electron beam from a Plasma Focus for medical applications [J].
Ceccolini, E. ;
Rocchi, F. ;
Mostacci, D. ;
Sumini, M. ;
Tartari, A. ;
Mariotti, F. .
JOURNAL OF APPLIED PHYSICS, 2012, 112 (05)
[7]   TIME FREQUENCY-DISTRIBUTIONS - A REVIEW [J].
COHEN, L .
PROCEEDINGS OF THE IEEE, 1989, 77 (07) :941-981
[8]   Dimensions and lifetime of the plasma focus pinch [J].
Lee, S ;
Serban, A .
IEEE TRANSACTIONS ON PLASMA SCIENCE, 1996, 24 (03) :1101-1105
[9]   Wigner spectrograms for femtosecond pulse-shaped heterodyne and autocorrelation measurements [J].
Mukamel, S ;
CiordasCiurdariu, C ;
Khidekel, V .
IEEE JOURNAL OF QUANTUM ELECTRONICS, 1996, 32 (08) :1278-1288
[10]   Spectral study of the electron beam emitted from a 3 kJ plasma focus [J].
Patran, A ;
Tan, LC ;
Stoenescu, D ;
Rafique, MS ;
Rawat, RS ;
Springham, SV ;
Tan, TL ;
Lee, P ;
Zakaullah, M ;
Lee, S .
PLASMA SOURCES SCIENCE & TECHNOLOGY, 2005, 14 (03) :549-560