A systematic review of variability and reliability of manual and automated blood pressure readings

被引:54
作者
Skirton, Heather [1 ]
Chamberlain, Wendy [2 ]
Lawson, Caroline [3 ]
Ryan, Helen
Young, Emma [4 ]
机构
[1] Univ Plymouth, Fac Hlth, Taunton TA1 5YD, Somerset, England
[2] Musgrove Pk Hosp, Taunton & Somerset Fdn Trust, Taunton, Somerset, England
[3] Yeovil Dist Hosp NHS Fdn Trust, Queensway Treatment Ctr, Higher Kingston, England
[4] Yeovil Dist Hosp NHS Fdn Trust, Yeovil, Somerset, England
关键词
blood pressure; nursing; observations; systematic review; vital signs; HYPERTENSION; SPHYGMOMANOMETERS; SOCIETY; DEVICE; POPULATION; VALIDATION; ACCURACY; ABC;
D O I
10.1111/j.1365-2702.2010.03528.x
中图分类号
R47 [护理学];
学科分类号
1011 ;
摘要
Aims and objectives. To compare the accuracy and appropriateness of auscultatory (manual) and oscillometric (automated) devices for measuring blood pressure in clinical settings. Background. Accurate measurement of blood pressure is integral to early recognition of deterioration in the condition of a patient. Despite recommendations regarding the use of auscultatory devices in situations where treatment decisions are made dependent on blood readings, the use of automated machines is becoming common practice. Design. Systematic review. Methods. A search of the Medline, CINAHLPlus and The Cochrane Library databases was undertaken for papers published in English between January 1997-May 2009. Sixteen studies were identified that fulfilled the inclusion criteria. After quality assessment, all were included in the review. Results are presented in tabular and narrative form. Results. In 10 of the studies reviewed, the authors came to the conclusion that oscillometric devices were less accurate than auscultatory devices. However, in most cases the oscillometric device appears sufficiently accurate for clinical use, the exceptions being use with hypertensive patients, patients with arrhythmia and after trauma. Only two studies assessed the comparative accuracy of aneroid devices, and these indicated that they were more accurate than oscillometric devices, but the differences were not clinically important. Conclusions. There are situations where the substitution of oscillometric for auscultatory devices could have particularly serious repercussions for the patient, such as when the patient is either hypertensive or hypotensive. However, further research is required on the use of aneroid sphygmomanometers as a replacement for mercury devices. Relevance to clinical practice. Practitioners should be made aware of the need to use auscultatory devices in specific circumstances, such as in management of hypertension, after the patient has experienced trauma or where there is significant potential for deterioration in the patient's condition.
引用
收藏
页码:602 / 614
页数:13
相关论文
共 40 条
  • [1] ABC of hypertension - Blood pressure measurement - Part I - Sphygmomanometry: factors common to all techniques
    Beevers, G
    Lip, GYH
    O'Brien, E
    [J]. BRITISH MEDICAL JOURNAL, 2001, 322 (7292): : 981 - 985
  • [2] Britain) N. P. S. A., 2007, SAFER CARE ACUTELY I
  • [3] Automated assessment of blood pressure using BpTRU compared with assessments by a trained technician and a clinic nurse
    Campbell, NRC
    Conradson, HE
    Kang, J
    Brant, R
    Anderson, T
    [J]. BLOOD PRESSURE MONITORING, 2005, 10 (05) : 257 - 262
  • [4] Carney S L, 1999, J Qual Clin Pract, V19, P95, DOI 10.1046/j.1440-1762.1999.00308.x
  • [5] The validity of emergency department triage blood pressure measurements
    Cienki, JJ
    DeLuca, LA
    Daniel, N
    [J]. ACADEMIC EMERGENCY MEDICINE, 2004, 11 (03) : 237 - 243
  • [6] CONWAY J, 1989, J HYPERTENS, V7, pS29
  • [7] Clinical evaluation of an automated oscillometric blood pressure wrist device
    Cuckson, AC
    Moran, P
    Seed, P
    Reinders, A
    Shennan, AH
    [J]. BLOOD PRESSURE MONITORING, 2004, 9 (01) : 31 - 37
  • [8] Are automated blood pressure measurements accurate in trauma patients?
    Davis, JW
    Davis, IC
    Bennink, LD
    Bilello, JF
    Kaups, KL
    Parks, SN
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2003, 55 (05): : 860 - 863
  • [9] Clinical assessment of the DINAMAP ProCare monitor in an adult population according to the British Hypertension Society Protocol
    de Greeff, Annemarie
    Reggiori, Fayrouge
    Shennan, Andrew H.
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (01) : 51 - 55
  • [10] Department of Health, 2000, COMPR CRIT CAR REV A