A flow-through amperometric sensor for micro-analytical systems

被引:20
作者
Pijanowska, DG
Sprenkels, AJ
Olthuis, W
Bergveld, P
机构
[1] Univ Twente, MESA Res Inst, NL-7500 AE Enschede, Netherlands
[2] Polish Acad Sci, Inst Biocybernet & Biomed Engn, PL-02109 Warsaw, Poland
关键词
biosensors; amperometric detection; flow-through systems; glucose sensors; lab-on-a-chip; mu TAS;
D O I
10.1016/S0925-4005(03)00072-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, the performance of the flow-through amperometric sensor based on semi-permeable dialysis tubing implemented in silicon is presented. The sensor is designed in the form to be an integral component of a lab-on-a-chip systems and it has been successfully incorporated into an integrated microdialysis system. Results concerning glucose sensors based on a flow-through sensor implemented in silicon operating as part of the integrated microdialysis system are presented. Moreover, comparison of the performances of the glucose sensor implemented in Perspex in the earlier work and the sensor implemented in silicon are discussed. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:98 / 102
页数:5
相关论文
共 22 条
[1]   ELECTROCHEMICAL IMMOBILIZATION OF ENZYMES .1. THEORY [J].
BARTLETT, PN ;
WHITAKER, RG .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1987, 224 (1-2) :27-35
[2]   ELECTROCHEMICAL IMMOBILIZATION OF ENZYMES .3. IMMOBILIZATION OF GLUCOSE-OXIDASE IN THIN-FILMS OF ELECTROCHEMICALLY POLYMERIZED PHENOLS [J].
BARTLETT, PN ;
TEBBUTT, P ;
TYRRELL, CH .
ANALYTICAL CHEMISTRY, 1992, 64 (02) :138-142
[3]   A flow-through amperometric sensor based on dialysis tubing and free enzyme reactors [J].
Böhm, S ;
Pijanowska, D ;
Olthuis, W ;
Bergveld, P .
BIOSENSORS & BIOELECTRONICS, 2001, 16 (06) :391-397
[4]   IN-VIVO MONITORING OF LACTATE AND GLUCOSE WITH MICRODIALYSIS AND ENZYME REACTORS IN INTENSIVE-CARE MEDICINE [J].
DEBOER, J ;
KORF, J ;
PLIJTERGROENDIJK, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1994, 17 (03) :163-170
[5]  
KHINKERS H, 1995, SENSOR ACTUAT B-CHEM, V26, P398
[6]   SILICON-WAFER INTEGRATED ENZYME REACTORS [J].
LAURELL, T ;
DROTT, J ;
ROSENGREN, L .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (3-4) :289-299
[7]   Enhanced enzyme activity in silicon integrated enzyme reactors utilizing porous silicon as the coupling matrix [J].
Laurell, T ;
Drott, J ;
Rosengren, L ;
Lindstrom, K .
SENSORS AND ACTUATORS B-CHEMICAL, 1996, 31 (03) :161-166
[8]  
MANZ A, 1993, SENSOR ACTUAT B-CHEM, V1, P244
[9]   FLOW-INJECTION ANALYSIS FOR GLUCOSE USING AN AMPEROMETRIC ENZYME ELECTRODE BASED ON LIPID-MODIFIED GLUCOSE-OXIDASE AS THE DETECTOR [J].
MIZUTANI, F ;
YABUKI, S .
BIOSENSORS & BIOELECTRONICS, 1994, 9 (06) :411-414
[10]   A microdialysis fibre based sampler for flow injection analysis: Determination of L-lactate in biofluids by an electrochemically synthesised bilayer membrane based biosensor [J].
Palmisano, F ;
Centonze, D ;
Quinto, M ;
Zambonin, PG .
BIOSENSORS & BIOELECTRONICS, 1996, 11 (04) :419-425