Wavelet Analysis of the Temporal Dynamics of the Laser Speckle Contrast in Human Skin

被引:27
作者
Mizeva, Irina [1 ,2 ]
Dremin, Viktor [2 ,3 ]
Potapova, Elena [2 ]
Zherebtsov, Evgeny [2 ,3 ]
Kozlov, Igor [2 ]
Dunaev, Andrey [2 ]
机构
[1] Inst Continuous Media Mech, Perm 614013, Russia
[2] Orel State Univ, Res & Dev Ctr Biomed Photon, Oryol 302026, Russia
[3] Univ Oulu, Optoelect & Measurement Tech Unit, Oulu, Finland
基金
俄罗斯科学基金会;
关键词
Speckle; Blood; Imaging; Skin; Oscillators; Spatial resolution; Wavelet transforms; Laser speckle contrast imaging; laser Doppler flowmetry; blood microcirculation; oscillations; wavelets; DOPPLER FLOWMETRY; BLOOD-FLOW; PERFUSION SIGNAL; OSCILLATIONS; MICROCIRCULATION; FLUCTUATIONS;
D O I
10.1109/TBME.2019.2950323
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Spectral analysis of laser Doppler flowmetry (LDF) signals has been widely used in studies of physiological vascular function regulation. An alternative to LDF is the laser speckle contrast imaging method (LSCI), which is based on the same physical principle. In contrast to LDF, LSCI provides non-scanning full-field imaging of a relatively wide skin area and offers high spatial and temporal resolutions, which allows visualization of microvascular structure. This circumstance, together with a large number of works which had shown the effectiveness of temporal LSCI analysis, gave impetus to experimental studies of the relation between LDF and LSCI used to monitor the temporal dynamics of blood flow. Methods: Continuous wavelet transform was applied to construct a time-frequency representation of a signal. Results: Analysis of 10 minute LDF and LSCI output signals recorded simultaneously revealed rather high correlation between oscillating components. It was demonstrated for the first time that the spectral energy of oscillations in the 0.01-2 Hz frequency range of temporal LSCI recordings carries the same information as the conventional LDF recordings and hence it reflects the same physiological vascular tone regulation mechanisms. Conclusion: The approach proposed can be used to investigate speckle pattern dynamics by LSCI in both normal and pathological conditions. Significance: The results of research on the influence of spatial binning and averaging on the spectral characteristics of perfusion monitored by LSCI are of considerable interest for the development of LSCI systems optimized to evaluate temporal dynamics.
引用
收藏
页码:1882 / 1889
页数:8
相关论文
共 50 条
[21]   Assessing spatial resolution versus sensitivity from laser speckle contrast imaging: application to frequency analysis [J].
Bricq, Stephanie ;
Mahe, Guillaume ;
Rousseau, David ;
Humeau-Heurtier, Anne ;
Chapeau-Blondeau, Francois ;
Varela, Julio Rojas ;
Abraham, Pierre .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2012, 50 (10) :1017-1023
[22]   Assessing spatial resolution versus sensitivity from laser speckle contrast imaging: application to frequency analysis [J].
Stéphanie Bricq ;
Guillaume Mahé ;
David Rousseau ;
Anne Humeau-Heurtier ;
François Chapeau-Blondeau ;
Julio Rojas Varela ;
Pierre Abraham .
Medical & Biological Engineering & Computing, 2012, 50 :1017-1023
[23]   Improved detectability of microcirculatory dynamics by laser speckle flowmetry [J].
Postnov, Dmitry D. ;
Sosnovtseva, Olga ;
Tuchin, Valery V. .
JOURNAL OF BIOPHOTONICS, 2015, 8 (10) :790-794
[24]   Linguistic Analysis of Laser Speckle Contrast Images Recorded at Rest and During Biological Zero: Comparison With Laser Doppler Flowmetry Data [J].
Humeau-Heurtier, Anne ;
Abraham, Pierre ;
Mahe, Guillaume .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (12) :2311-2321
[25]   Nephron blood flow dynamics measured by laser speckle contrast imaging [J].
Holstein-Rathlou, Niels-Henrik ;
Sosnovtseva, Olga V. ;
Pavlov, Alexey N. ;
Cupples, William A. ;
Sorensen, Charlotte Mehlin ;
Marsh, Donald J. .
AMERICAN JOURNAL OF PHYSIOLOGY-RENAL PHYSIOLOGY, 2011, 300 (02) :F319-F329
[26]   Cutaneous microvascular functional assessment during exercise: a novel approach using laser speckle contrast imaging [J].
Mahe, G. ;
Abraham, P. ;
Le Faucheur, A. ;
Bruneau, A. ;
Humeau-Heurtier, A. ;
Durand, S. .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 2013, 465 (04) :451-458
[27]   Laser Speckle Contrast Imaging of Skin Changes in Arteriovenous Malformation [J].
Humeau-Heurtier, Anne ;
Martin, Ludovic ;
Bazeries, Paul ;
Abraham, Pierre ;
Henni, Samir .
CIRCULATION-CARDIOVASCULAR IMAGING, 2017, 10 (03)
[28]   Spatiotemporal laser speckle contrast analysis for blood flow imaging with maximized speckle contrast [J].
Qiu, Jianjun ;
Li, Pengcheng ;
Luo, Weihua ;
Wang, Jia ;
Zhang, Hongyan ;
Luo, Qingming .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
[29]   On the equivalence and differences between laser Doppler flowmetry and laser speckle contrast analysis [J].
Fredriksson, Ingemar ;
Larsson, Marcus .
JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (12)
[30]   Aging effect on microcirculation: A multiscale entropy approach on laser speckle contrast images [J].
Khalil, A. ;
Humeau-Heurtier, A. ;
Gascoin, L. ;
Abraham, P. ;
Mahe, G. .
MEDICAL PHYSICS, 2016, 43 (07) :4008-4016