Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques

被引:25
|
作者
Clemente-Gutierrez, Francisco [1 ]
Perez-Vara, Consuelo [1 ]
机构
[1] Hosp Cent Def Gomez Ulla, Secc Radiofis, Serv Oncol Radioterap, Madrid 28047, Spain
来源
关键词
VMAT QA; pretreatment verifications; Mobius3D; COMPASS; CARLO-BASED PHOTON; RADIATION-THERAPY; IMRT QA; DETECTOR ARRAY; PASSING RATES; VMAT QA; BEAM; CONVOLUTION/SUPERPOSITION; CHAMBER; PHANTOM;
D O I
10.1120/jacmp.v16i2.5190
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/3 mm criteria). The clinical use of both systems was successfully evaluated.
引用
收藏
页码:198 / 217
页数:20
相关论文
共 50 条
  • [31] Dosimetric evaluation of helical tomotherapy and volumetric-modulated arc therapy for malignant pleural mesothelioma: a planning study for dose escalation
    Zhang, Fu-Li
    Lu, Na
    Jiang, Hua-Yong
    Chen, Dian-Dian
    Wang, Ya-Di
    TRANSLATIONAL CANCER RESEARCH, 2021, 10 (02) : 914 - 922
  • [32] Validation of the relative insensitivity of volumetric-modulated arc therapy (VMAT) plan quality to gantry space resolution
    Murtaza, Ghulam
    Cora, Stefania
    Khan, Ehsan Ullah
    JOURNAL OF RADIATION RESEARCH, 2017, 58 (04) : 579 - 590
  • [33] Dosimetric precision of 3D gel dosimetry compared with radiochromic films in volumetric modulated arc therapy
    Skyt, P. S.
    Hoye, E. M.
    Yates, E. S.
    Schmidt, M. L.
    Petersen, J. B. B.
    Balling, P.
    Muren, L. P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S409 - S410
  • [34] Prospective Clinical Validation of Virtual Patient-Specific Quality Assurance of Volumetric Modulated Arc Therapy Radiation Therapy Plans
    Wall, Phillip D. H.
    Hirata, Emily
    Morin, Olivier
    Valdes, Gilmer
    Witztum, Alon
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 113 (05): : 1091 - 1102
  • [35] A dose-volume intercomparison of volumetric-modulated arc therapy, 3D static conformal, and rotational conformal techniques for portal vein tumor thrombus in hepatocellular carcinoma
    Ogino, Ryo
    Hosono, Masako
    Ishii, Kentaro
    Tatsumi, Daisaku
    Tsutsumi, Shinichi
    Miki, Yoshitaka
    Masuoka, Yutaka
    Shimatani, Yasuhiko
    Miki, Yukio
    JOURNAL OF RADIATION RESEARCH, 2013, 54 (04) : 697 - 705
  • [36] Correlation analysis between 2D and 3D patient-specific quality assurance for volumetric modulated arc therapy
    Sharma, S.
    Sharma, D.
    Subramani, V.
    Gopishankar, N.
    Bhaskar, S.
    Pathy, S.
    Kumar, P.
    Thulkar, S.
    Chander, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (04): : 685 - 691
  • [37] Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
    Katsuta, Yoshiyuki
    Kadoya, Noriyuki
    Fujita, Yukio
    Shimizu, Eiji
    Matsunaga, Kenichi
    Sawada, Kinya
    Matsushita, Haruo
    Majima, Kazuhiro
    Jingu, Keiichi
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 1220 - 1225
  • [38] 3D Dose Reconstruction of Volumetric Modulated Arc Therapy From Delivered EPID Dose Image
    Yoon, J.
    Jung, J.
    Kim, J.
    Yeo, I.
    MEDICAL PHYSICS, 2014, 41 (06) : 221 - 221
  • [39] Influence of computed tomography contrast agent on radiotherapy dose calculation for pancreatic carcinoma: A dosimetric study based on tomotherapy and volumetric-modulated arc therapy techniques
    Zhu, Feng
    Wu, Weizhang
    Zhu, Fuhai
    Wang, Yong
    Wang, Yingjie
    Xia, Tingyi
    MEDICAL DOSIMETRY, 2017, 42 (04) : 317 - 325
  • [40] Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance
    Mans, Anton
    Rozendaal, Roel
    Janssen, Tomas
    Damen, Eugene
    Kaas, Jochem
    van Mourik, Anke
    Mijnheer, Ben
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 21 : 6 - 10