Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques

被引:25
|
作者
Clemente-Gutierrez, Francisco [1 ]
Perez-Vara, Consuelo [1 ]
机构
[1] Hosp Cent Def Gomez Ulla, Secc Radiofis, Serv Oncol Radioterap, Madrid 28047, Spain
来源
关键词
VMAT QA; pretreatment verifications; Mobius3D; COMPASS; CARLO-BASED PHOTON; RADIATION-THERAPY; IMRT QA; DETECTOR ARRAY; PASSING RATES; VMAT QA; BEAM; CONVOLUTION/SUPERPOSITION; CHAMBER; PHANTOM;
D O I
10.1120/jacmp.v16i2.5190
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/3 mm criteria). The clinical use of both systems was successfully evaluated.
引用
收藏
页码:198 / 217
页数:20
相关论文
共 50 条
  • [21] Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification
    Jin, X.
    Yan, H.
    Han, C.
    Zhou, Y.
    Yi, J.
    Xie, C.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1047):
  • [22] DOSIMETRIC PERFORMANCE OF A 2D DIODE ARRAY FOR VERIFICATION OF VOLUMETRIC MODULATED ARC THERAPY
    Nakada, R.
    Tatsumi, D.
    Sakamoto, M.
    Inoue, M.
    Ichida, T.
    Ishii, K.
    Tsutsumi, S.
    Ogino, R.
    Hosono, M.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S212 - S212
  • [23] Patient Specific Quality Assurance for Volumetric Modulated Arc Therapy (RapidArc) Using COMPASS 3D Dosimetry System
    Arun, G.
    Subramanian, V. Shanmuga
    Swamy, S. Thirumalai
    Kathirvel, M.
    Ramalingam, K.
    Kala, S.
    Kumar, S. Ashok
    Karthikeyan, K.
    Anantharaman, A.
    Kumar, N. Jagdish
    Arivarasan, I.
    Babaiah, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [24] Dosimetric Comparison of Hypofractionated Regimen in Breast Cancer Using Two Different Techniques: Intensity-Modulated Radiation Therapy (IMRT) and Volumetric-Modulated Arc Therapy (VMAT)
    Prasun, Pallav
    Kharade, Vipin
    Pal, Vikas
    Gupta, Manish
    Das, Saikat
    Pasricha, Rajesh
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (04)
  • [25] Quality assurance of volumetric-modulated arc therapy head and neck cancer treatment using PRESAGE (R) dosimeter
    Rehman, Jalil Ur
    Isa, Muhammad
    Ahmad, Nisar
    Asghar, H. M. Noor ul Huda Khan
    Gilani, Zaheer A.
    Chow, James C. L.
    Afzal, Muhammad
    Ibbott, Geoffrey S.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (04) : 441 - 446
  • [26] Delta4-based Dosimetric Error Detection in Volumetric-modulated Arc Therapy: Clinical Significance and Implications
    Udee, Nuntawat
    Commukchik, Supada
    Khamfongkhruea, Chirasak
    Kaewlek, Titipong
    Chusin, Thunyarat
    Yabsantia, Sumalee
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (01) : 56 - 63
  • [27] Analysis of direct clinical consequences of MLC positional errors in volumetric-modulated arc therapy using 3D dosimetry system
    Nithiyanantham, Karthikeyan
    Mani, Ganesh K.
    Subramani, Vikraman
    Mueller, Lutz
    Palaniappan, Karrthick K.
    Kataria, Tejinder
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 296 - 305
  • [28] Investigating the Dosimetric Leaf Gap Correction Factor of Mobius3D Dose Calculation for Volumetric-modulated Arc Radiotherapy Plans
    Sawapabmongkon, Thitipong
    Changkaew, Pimolpun
    Puttanawarut, Chanon
    Tangboonduangjit, Puangpen
    Khachonkham, Suphalak
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (02) : 261 - 269
  • [29] VOLUMETRIC-MODULATED ARC THERAPY: EFFECTIVE AND EFFICIENT END-TO-END PATIENT-SPECIFIC QUALITY ASSURANCE
    O'Daniel, Jennifer
    Das, Shiva
    Wu, Q. Jackie
    Yin, Fang-Fang
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05): : 1567 - 1574
  • [30] Impact of treatment planning quality assurance software on volumetric-modulated arc therapy plans for prostate cancer patients
    Sasaki, Motoharu
    Nakaguchi, Yuji
    Kamomae, Takeshi
    Kajino, Akimi
    Ikushima, Hitoshi
    MEDICAL DOSIMETRY, 2021, 46 (04) : E1 - E6