Dosimetric validation and clinical implementation of two 3D dose verification systems for quality assurance in volumetric-modulated arc therapy techniques

被引:25
作者
Clemente-Gutierrez, Francisco [1 ]
Perez-Vara, Consuelo [1 ]
机构
[1] Hosp Cent Def Gomez Ulla, Secc Radiofis, Serv Oncol Radioterap, Madrid 28047, Spain
关键词
VMAT QA; pretreatment verifications; Mobius3D; COMPASS; CARLO-BASED PHOTON; RADIATION-THERAPY; IMRT QA; DETECTOR ARRAY; PASSING RATES; VMAT QA; BEAM; CONVOLUTION/SUPERPOSITION; CHAMBER; PHANTOM;
D O I
10.1120/jacmp.v16i2.5190
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/3 mm criteria). The clinical use of both systems was successfully evaluated.
引用
收藏
页码:198 / 217
页数:20
相关论文
共 55 条
[1]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[2]  
Alber M, 2008, ESTRO PHYS BOOKLET
[3]   An analysis of tolerance levels in IMRT quality assurance procedures [J].
Basran, Parminder S. ;
Woo, Milton K. .
MEDICAL PHYSICS, 2008, 35 (06) :2300-2307
[4]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[5]   Evaluation of a 2D detector array for patient-specific VMAT QA with different setups [J].
Boggula, Ramesh ;
Birkner, Mattias ;
Lohr, Frank ;
Steil, Volker ;
Wenz, Frederik ;
Wertz, Hansjoerg .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22) :7163-7177
[6]   PATIENT-SPECIFIC 3D PRETREATMENT AND POTENTIAL 3D ONLINE DOSE VERIFICATION OF MONTE CARLO-CALCULATED IMRT PROSTATE TREATMENT PLANS [J].
Boggula, Ramesh ;
Jahnke, Lennart ;
Wertz, Hansjoerg ;
Lohr, Frank ;
Wenz, Frederik .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (04) :1168-1175
[7]   Experimental validation of a commercial 3D dose verification system for intensity-modulated arc therapies [J].
Boggula, Ramesh ;
Lorenz, Friedlieb ;
Mueller, Lutz ;
Birkner, Mattias ;
Wertz, Hansjoerg ;
Stieler, Florian ;
Steil, Volker ;
Lohr, Frank ;
Wenz, Frederik .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (19) :5619-5633
[8]  
Both S, 2007, J APPL CLIN MED PHYS, V8, P1
[9]   Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planning [J].
Boyer, A ;
Xing, L ;
Ma, CM ;
Curran, B ;
Hill, R ;
Kania, A ;
Bleier, A .
MEDICAL PHYSICS, 1999, 26 (02) :187-195
[10]   Comment on "IMRT commissioning: Some causes for concern" [J].
Cadman, Patrick F. .
MEDICAL PHYSICS, 2011, 38 (07) :4464-4465