Accurate white blood cell differential by Alinity hq: A comparison with flow cytometry and manual differential

被引:3
作者
Gambell, Peter [1 ]
Rowley, Grant [1 ]
Pham, Tracy Anh Thu [1 ]
Dang, Theresa Loan [2 ]
Mulumba, Huguette [2 ]
Smith, Larry [2 ]
Lakos, Gabriella [2 ]
机构
[1] Peter MacCallum Canc Ctr, Melbourne, Vic, Australia
[2] Abbott, 4551Great Amer Pkwy, Santa Clara, CA 95054 USA
关键词
differentiation; flow cytometry; hematology analyzer; leukocytes; COUNTS; ENUMERATION; CYTODIFF; SYSTEM;
D O I
10.1111/ijlh.13764
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction White blood cell (WBC) differential by flow cytometry can report a six-part WBC differential and enumerate blasts. Some modern hematology analyzers are also able to provide a six-part WBC differential (including immature granulocytes). Our goal was to compare the WBC differential obtained by the Abbott Alinity hq hematology analyzer to an 8-color single-tube flow cytometry method and to manual WBC differential. Methods Samples from 144 patients were tested with Alinity hq, flow cytometry, and microscopic WBC differential. The WBC count ranged from 1.22 to 359 x 10(9)/L, and 34 samples were flagged by the analyzer for abnormal WBC morphology. Results Strong concordance was demonstrated between Alinity hq and flow cytometry for all six components of the differential, with correlation coefficients ranging from 0.86 (basophils) to 1.00 (lymphocytes). Small, clinically insignificant positive difference was observed between Alinity hq and flow cytometry for mature and total neutrophils (2.51% and 1.85%) and eosinophils (0.14%), and small negative difference for immature granulocytes (-0.65%), lymphocytes (-0.61%), and basophils (-0.21%). No bias was detected between the Alinity hq and flow cytometry monocyte counts. Alinity hq and flow cytometry results agreed with the manual differential, apart from small, clinically insignificant differences. Alinity hq nucleated red blood cell concentrations were equivalent with the manual results (r = 0.95, slope = 1.16). The percentage of blasts by flow cytometry demonstrated good correlation and agreement with the manual count (r = 0.99, slope = 1.35). Conclusion Alinity hq has produced accurate six-part WBC differential in this three-way comparison, equivalent to flow cytometry and morphological classification.
引用
收藏
页码:288 / 295
页数:8
相关论文
共 27 条
[11]  
Hubl W, 1996, CYTOMETRY, V26, P69, DOI 10.1002/(SICI)1097-0320(19960315)26:1<69::AID-CYTO11>3.0.CO
[12]  
2-Q
[13]   Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts [J].
Kahng, Jimin ;
Kim, Yonggoo ;
Kim, Myungshin ;
Oh, Eun-Jee ;
Park, Yeon-Joon ;
Han, Kyungja .
ANNALS OF LABORATORY MEDICINE, 2015, 35 (01) :28-34
[14]   Japanese Society for Laboratory Hematology flow cytometric reference method of determining the differential leukocyte count: external quality assurance using fresh blood samples [J].
Kawai, Y. ;
Nagai, Y. ;
Ogawa, E. ;
Kondo, H. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (02) :202-222
[15]   White blood cell differential counts in severely leukopenic samples: a comparative analysis of different solutions available in modern laboratory hematology [J].
Kim, Ah Hyun ;
Lee, Wonbae ;
Kim, Myungshin ;
Kim, Yonggoo ;
Han, Kyungja .
BLOOD RESEARCH, 2014, 49 (02) :120-126
[16]   Performance evaluation of the CellaVision DM96 system WBC differentials by automated digital image analysis supported by an artificial neural network [J].
Kratz, A ;
Bengtsson, HI ;
Casey, JE ;
Keefe, JM ;
Beatrice, GH ;
Grzybek, DY ;
Lewandrowski, KB ;
Van Cott, EM .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2005, 124 (05) :770-781
[17]   Digital morphology analyzers in hematology: ICSH review and recommendations [J].
Kratz, Alexander ;
Lee, Szu-hee ;
Zini, Gina ;
Riedl, Jurgen A. ;
Hur, Mina ;
Machin, Sam .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 (04) :437-447
[18]   Reference Values for WBC Differential by Hematoflow Analysis [J].
Magierowicz, Marion ;
Lechevalier, Nicolas ;
Freynet, Nicolas ;
Pastoret, Cedric ;
Badaoui, Bouchra ;
Ly-Sunnaram, Beatrice ;
Fest, Thierry ;
Lacombe, Francis ;
Wagner-Ballon, Orianne ;
Roussel, Mikael .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2019, 151 (03) :324-327
[19]  
Novis DA, 2006, ARCH PATHOL LAB MED, V130, P596
[20]   Comparison of the Cytodiff flow cytometric leucocyte differential count system with the Sysmex XE-2100 and Beckman Coulter UniCel DxH 800 [J].
Park, Borae G. ;
Park, Chan-Jeoung ;
Kim, Sue ;
Yoon, Chan-Hee ;
Kim, Duck-Hee ;
Jang, Seongsoo ;
Chi, Hyun-Sook .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 (06) :584-593