Characteristics of gated treatment using an optical surface imaging and gating system on an Elekta linac

被引:46
作者
Freislederer, Philipp [1 ]
Reiner, Michael [1 ]
Hoischen, Winfried [1 ]
Quanz, Anton [1 ]
Heinz, Christian [1 ]
Walter, Franziska [1 ]
Belka, Claus [1 ]
Soehn, Matthias [1 ]
机构
[1] LMU Univ Hosp, Dept Radiat Oncol, D-81377 Munich, Germany
来源
RADIATION ONCOLOGY | 2015年 / 10卷
关键词
Respiratory gating; Catalyst; Latency; Dosimetry; RADIOTHERAPY TREATMENT; MOTION; VARIANCE; DELIVERY;
D O I
10.1186/s13014-015-0376-x
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Knowing the technical characteristics of gated radiotherapy equipment is crucial for ensuring precise and accurate treatment when using techniques such as Deep-Inspiration Breath-Hold and gating under free breathing. With one of the first installations of the novel surface imaging system Catalyst (TM) (C-RAD AB, Sweden) in connection with an Elekta Synergy linear accelerator (Elekta AB, Sweden) via the Elekta Response Interface, characteristics like dose delivery accuracy and time delay were investigated prior to clinical implementation of gated treatments in our institution. Methods: In this study a moving phantom was used to simulate respiratory motion which was registered by the Catalyst (TM) system. The gating level was set manually. Within this gating window a trigger signal is automatically sent to the linac initiating treatment delivery. Dose measurements of gated linac treatment beams with different gating levels were recorded with a static 2D-Diode Array (MapCheck2, Sun Nuclear Co., USA) and compared to ungated reference measurements for different field sizes. In addition, the time delay of gated treatment beams was measured using radiographic film. Results: The difference in dose delivery between gated and ungated treatment decreases with the size of the chosen gating level. For clinically relevant gating levels of about 30%, the differences in dose delivery accuracy remain below 1%. In comparison with other system configurations in literature, the beam-on time delay shows a large deviation of 851 ms +/- 100 ms. Conclusions: When performing gated treatment, especially for free-breathing gating, factors as time delay and dose delivery have to be evaluated regularly in terms of a quality assurance process. Once these parameters are known they can be accounted and compensated for, e.g. by adjusting the pre-selected gating level or the internal target volume margins and by using prediction algorithms for breathing curves. The usage of prediction algorithms becomes inevitable with the high beam-on time delay which is reported here.
引用
收藏
页数:6
相关论文
共 19 条
  • [1] Evaluation of integrated respiratory gating systems on a Novalis Tx system
    Chang, Zheng
    Liu, TongHai
    Cai, Jing
    Chen, Qing
    Wang, Zhiheng
    Yin, Fang-Fang
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (03): : 71 - 79
  • [2] Delivery efficiency of an Elekta linac under gated operation
    Cui, Guoqiang
    Housley, David J.
    Chen, Fan
    Mehta, Vivek K.
    Shepard, David M.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05): : 2 - 11
  • [3] Gating characteristics of an Elekta radiotherapy treatment unit measured with three types of detector
    Evans, Philip M.
    Symonds-Tayler, J. Richard N.
    Colgan, Ruth
    Hugo, Geoffrey D.
    Letts, Nicky
    Sandin, Carlos
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08) : N201 - N210
  • [4] Reduced lung dose during radiotherapy for thoracic esophageal carcinoma: VMAT combined with active breathing control for moderate DIBH
    Gong, Guanzhong
    Wang, Ruozheng
    Guo, Yujie
    Zhai, Deyin
    Liu, Tonghai
    Lu, Jie
    Chen, Jinhu
    Liu, Chengxin
    Yin, Yong
    [J]. RADIATION ONCOLOGY, 2013, 8
  • [5] Stereotactic body radiotherapy using gated radiotherapy with real-time tumor-tracking for stage I non-small cell lung cancer
    Inoue, Tetsuya
    Katoh, Norio
    Onimaru, Rikiya
    Shimizu, Shinichi
    Tsuchiya, Kazuhiko
    Suzuki, Ryusuke
    Sakakibara-Konishi, Jun
    Shinagawa, Naofumi
    Oizumi, Satoshi
    Shirato, Hiroki
    [J]. RADIATION ONCOLOGY, 2013, 8
  • [6] Time delay measurement for linac based treatment delivery in synchronized respiratory gating radiotherapy
    Jin, JY
    Yin, FF
    [J]. MEDICAL PHYSICS, 2005, 32 (05) : 1293 - 1296
  • [7] The management of respiratory motion in radiation oncology report of AAPM Task Group 76
    Keall, Paul J.
    Mageras, Gig S.
    Balter, James M.
    Emery, Richard S.
    Forster, Kenneth M.
    Jiang, Steve B.
    Kapatoes, Jeffrey M.
    Low, Daniel A.
    Murphy, Martin J.
    Murray, Brad R.
    Ramsey, Chester R.
    Van Herk, Marcel B.
    Vedam, S. Sastry
    Wong, John W.
    Yorke, Ellen
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3874 - 3900
  • [8] Task Group 142 report: Quality assurance of medical accelerators
    Klein, Eric E.
    Hanley, Joseph
    Bayouth, John
    Yin, Fang-Fang
    Simon, William
    Dresser, Sean
    Serago, Christopher
    Aguirre, Francisco
    Ma, Lijun
    Arjomandy, Bijan
    Liu, Chihray
    Sandin, Carlos
    Holmes, Todd
    [J]. MEDICAL PHYSICS, 2009, 36 (09) : 4197 - 4212
  • [9] Dosimetric characteristics of a new linear accelerator under gated operation
    Kriminski, Sergey
    Li, Alex N.
    Solberg, Timothy D.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (01): : 65 - 76
  • [10] Respiration gated radiotherapy treatment: A technical study
    Kubo, HD
    Hill, BC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) : 83 - 91