Comparison of established and novel purity tests for the quality control of heparin by means of a set of 177 heparin samples

被引:34
作者
Alban, Susanne [2 ]
Luehn, Susanne [2 ]
Schiemann, Simone [2 ]
Beyer, Tanja [1 ]
Norwig, Jochen [3 ]
Schilling, Claudia [4 ]
Raedler, Oliver [4 ]
Wolf, Bernhard [4 ]
Matz, Magnus [5 ]
Baumann, Knut [5 ]
Holzgrabe, Ulrike [1 ]
机构
[1] Univ Wurzburg, Inst Pharm & Food Chem, D-97074 Wurzburg, Germany
[2] Univ Kiel, Inst Pharmaceut, D-24118 Kiel, Germany
[3] Fed Inst Drugs & Med Device, D-53175 Bonn, Germany
[4] Ratiopharm Grp, D-89079 Ulm, Germany
[5] Univ Technol Braunschweig, Inst Pharmaceut Chem, D-38106 Braunschweig, Germany
关键词
Heparin; OSCS; H-1; NMR; Chemometric classification; Two-step-fluorescence assay; Two-step-aXa-assay; Coagulation assays; Raman spectroscopy; ATR-IR spectroscopy; Purity tests; OVERSULFATED CHONDROITIN SULFATE; LOW-MOLECULAR-WEIGHT; DERMATAN SULFATE; UNFRACTIONATED HEPARIN; SOLE CONTAMINANT; IMPURITIES;
D O I
10.1007/s00216-010-4169-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The widespread occurrence of heparin contaminated with oversulfated chrondroitin sulfate (OSCS) in 2008 initiated a comprehensive revision process of the Pharmacopoeial heparin monographs and stimulated research in analytical techniques for the quality control of heparin. Here, a set of 177 heparin samples from the market in 2008 as well as pure heparin sodium spiked with defined amounts of OSCS and DS were used to evaluate established and novel methods for the quality control of heparin. Besides H-1 nuclear magnetic resonance spectroscopy (NMR), the assessment included two further spectroscopic methods, i.e., attenuated total reflection-infrared spectroscopy (ATR-IR) and Raman spectroscopy, three coagulation assays, i.e., activated partial thromboplastin time (aPTT) performed with both sheep and human plasma and the prothrombin time (PT), and finally two novel purity assays, each consisting of an incubation step with heparinase I followed by either a fluorescence measurement (Inc-PolyH-assay) or by a chromogenic aXa-assay (Inc-aXa-assay). NMR was shown to allow not only sensitive detection, but also quantification of OSCS by using the peak-height method and a response factor determined by calibration. Chemometric evaluation of the NMR, ATR-IR, and Raman spectra by statistical classification techniques turned out to be best with NMR spectra concerning the detection of OSCS. The validity of the aPTT, the current EP assay, could be considerably improved by replacing the sheep plasma by human plasma. In this way, most of the contaminated heparin samples did not meet the novel potency limit of 180 IU/mg. However, also more than 50% of the uncontaminated samples had < 180 IU/MG. In contrast to the aPTT, the PT specifically detects OSCS and other heparin mimetics (LOD 3%). About ten times more sensitive are both the Inc-PolyH-assay and the Inc-aXa-assay, two rapid and simple quantification assays for heparin mimetics. The determined OSCS contents of the heparin samples excellently correlated with those calculated from the NMR spectra. In conclusion, NMR proved to be the current spectroscopic method of choice. The two two-step-assays represent options to supplement NMR, especially as tests for the initial screening, since they detect any heparin mimetic without requiring special expertise for interpretation of the results.
引用
收藏
页码:605 / 620
页数:16
相关论文
共 52 条
  • [1] Alban S, 2009, J THROMB HAEMOST S2, V7, pPP
  • [2] Prothrombin Time for Detection of Contaminated Heparins
    Alban, Susanne
    Luehn, Susanne
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2008, 359 (25) : 2732 - 2734
  • [3] [Anonymous], 2008, 2224 COUNC EUR EUR P
  • [4] [Anonymous], 2010, MON EUR PHARM
  • [5] [Anonymous], 1996, OUT OF BAG ESTIMATIO
  • [6] [Anonymous], HEP SOD MON STAG 2
  • [7] Identification of a simple and sensitive microplate method for the detection of oversulfated chondroitin sulfate in heparin products
    Bairstow, Shawn
    McKee, Jeff
    Nordhaus, Mark
    Johnson, Richard
    [J]. ANALYTICAL BIOCHEMISTRY, 2009, 388 (02) : 317 - 321
  • [8] Quality assessment of unfractionated heparin using 1H nuclear magnetic resonance spectroscopy
    Beyer, T.
    Diehl, B.
    Randel, G.
    Humpfer, E.
    Schaefer, H.
    Spraul, M.
    Schollmayer, C.
    Holzgrabe, U.
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2008, 48 (01) : 13 - 19
  • [9] Composition of OSCS-contaminated heparin occurring in 2008 in batches on the German market
    Beyer, Tanja
    Matz, Magnus
    Brinz, Daniela
    Raedler, Oliver
    Wolf, Bernhard
    Norwig, Jochen
    Baumann, Knut
    Alban, Susanne
    Holzgrabe, Ulrike
    [J]. EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2010, 40 (04) : 297 - 304
  • [10] Improved impurity fingerprinting of heparin by high resolution 1H NMR spectroscopy
    Bigler, Peter
    Brenneisen, Rudolf
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2009, 49 (04) : 1060 - 1064