A novel multicell silicon drift detector module for X-ray spectroscopy and imaging applications

被引:23
作者
Hansen, K [1 ]
Tröger, L [1 ]
机构
[1] DESY, D-22607 Hamburg, Germany
关键词
radiation shield; silicon drift detector array; thermal behavior; thin-film technique; three-dimensional (3-D) module; X-ray imaging; X-ray spectroscopy;
D O I
10.1109/23.901182
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel 61-cell silicon drift detector module is proposed for high-counting rates and high-resolution X-ray spectroscopy and imaging applications at energies up to about 30 keV. Its hexagonal geometry with sloped sidewalls allows a buckyball arrangement. Up to 1860 cells of an active area of 5 mm(2) each can be reached at an average distance to the specimen of only 4.3 cm. An optimized electrical connection concept offers a vertical integration of It sensor array and signal processing electronics. Worst-case crosstalk of 0.3% (-50 dB) and signal loss of 7% (-24 dB) were derived from electrical field analysis. A sandwich of metal foils between X-ray sensor and electronics reduces the dose of radiation at the electronics by more than eight orders of magnitude. Module-induced background fluorescence reaches its maximum at an incident energy of 13 keV, but remains below 10(-4) of the incoming photon Bur at all energies. Three-dimensional (3-D) simulations on thermal module behavior show that the use of graphite as housing material allows an operation with spatial inhomogenities of <1 <degrees>C on the sensor's active area. Empirically, we found that the difference between the temperatures of the sensor and the heat sink depends linearly on the power dissipated by the electronics and sensor chip. The corresponding thermal resistances are 4.3 and 0.8 K/W.
引用
收藏
页码:2748 / 2757
页数:10
相关论文
共 16 条
[1]   Atomic holography with x rays [J].
Adams, B ;
Novikov, DV ;
Hiort, T ;
Materlik, G ;
Kossel, E .
PHYSICAL REVIEW B, 1998, 57 (13) :7526-7534
[2]  
[Anonymous], 1991, TRANSMISSION LINE DE
[3]   X-RAY WAVELENGTHS [J].
BEARDEN, JA .
REVIEWS OF MODERN PHYSICS, 1967, 39 (01) :78-&
[4]   A SUITE OF PROGRAMS FOR CALCULATING X-RAY ABSORPTION, REFLECTION, AND DIFFRACTION PERFORMANCE FOR A VARIETY OF MATERIALS AT ARBITRARY WAVELENGTHS [J].
BRENNAN, S ;
COWAN, PL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :850-853
[5]   Performance of silicon drift detectors in a magnetic field [J].
Castoldi, A ;
Gatti, E ;
Manzari, V ;
Rehak, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 399 (2-3) :227-243
[6]   Continuous charge restoration in semiconductor detectors by means of the gate-to-drain current of the integrated front-end JFET [J].
Fiorini, C ;
Lechner, P .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (03) :761-764
[7]   Thermal behaviour of front-end multi-chip modules [J].
Hansen, K ;
Durica, M ;
Klar, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 408 (2-3) :397-407
[8]   FLUORESCENCE DETECTION OF EXAFS - SENSITIVITY ENHANCEMENT FOR DILUTE SPECIES AND THIN-FILMS [J].
JAKLEVIC, J ;
KIRBY, JA ;
KLEIN, MP ;
ROBERTSON, AS ;
BROWN, GS ;
EISENBERGER, P .
SOLID STATE COMMUNICATIONS, 1977, 23 (09) :679-682
[9]  
KAPPEN P, IN PRESS NUCL INSTRU
[10]  
*MAFIA COLL, US GUID MAFIA VERS 3