Comparison of Peripapillary Retinal Nerve Fiber Layer Thickness Measured by Spectral vs. Time Domain Optical Coherence Tomography

被引:12
作者
Hong, Samin [1 ]
Seong, Gong Je [1 ]
Kim, Sung Soo [1 ]
Kang, Sung Yong [1 ]
Kim, Chan Yun [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Ophthalmol, Inst Vis Res, Seoul 120752, South Korea
关键词
Glaucoma; Optical coherence tomography; Retinal nerve fiber layer; Spectral-domain; Time-domain; SIGNAL STRENGTH; STRATUS OCT; GLAUCOMA; AGREEMENT; DEFINITION; SPEED;
D O I
10.3109/02713683.2010.533807
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Patients and Methods: One hundred and eighty-eight subjects (normal tension glaucoma, n == 80; primary open angle glaucoma, n == 49; healthy controls, n == 59) were included in this study. Peripapillary RNFL thickness was measured by an Optic Disc Cube 200 xx 200 scan of the spectral domain Cirrus HD OCT and by a Fast RNFL scan of the time domain Stratus OCT. They were compared using linear regression (R<SU2</SU), Bland-Altman plots, weighted Kappa (K-w) coefficients, and areas under the receiver operating characteristic curve (AROC). Results: Except for the inferior quadrant, RNFL thickness measurements by Cirrus HD OCT were thinner than Stratus OCT (all p < 0.001); for the inferior quadrant, this tendency was vice versa (p == 0.043). Average RNFL thicknesses of the two OCTs showed a good correlation (R<SU2</SU, 0.852) but not excellent agreement (mean of difference, 4.0 mu mu m; ++1.96 SD, 10.7 mu mu m; -1.96 SD, -18.6 mu mu m). Regarding the probability codes, only the average RNFL thickness had almost perfect agreement (K-w, 0.983). In addition, Cirrus HD OCT had the largest AROC for the superior quadrant (0.795), but Stratus OCT had the largest AROC for average RNFL thickness (0.764) (p == 0.281). Conclusions: Except for the inferior quadrant, the average and quadrants RNFL thickness measured by the Cirrus HD OCT were thinner than the Stratus OCT. Regarding discriminating power for glaucoma detection, the Cirrus HD OCT was better than the Stratus OCT for the superior and inferior quadrant RNFL thicknesses but not for the average RNFL thickness.
引用
收藏
页码:125 / 134
页数:10
相关论文
共 25 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] Dependence of weighted kappa coefficients on the number of categories
    Brenner, H
    Kliebsch, U
    [J]. EPIDEMIOLOGY, 1996, 7 (02) : 199 - 202
  • [3] Reproducibility of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eyes
    Budenz, Donald L.
    Fredette, Marie-Josee
    Feuer, William J.
    Anderson, Douglas R.
    [J]. OPHTHALMOLOGY, 2008, 115 (04) : 661 - 666
  • [4] Sensitivity and Specificity of Time-Domain versus Spectral-Domain Optical Coherence Tomography in Diagnosing Early to Moderate Glaucoma
    Chang, Robert T.
    Knight, O'Rese J.
    Feuer, William J.
    Budenz, Donald L.
    [J]. OPHTHALMOLOGY, 2009, 116 (12) : 2294 - 2299
  • [5] Spectral domain optical coherence tomography - Ultra-high speed, ultra-high resolution ophthalmic imaging
    Chen, TC
    Cense, B
    Pierce, MC
    Nassif, N
    Park, BH
    Yun, SH
    White, BR
    Bouma, BE
    Tearney, GJ
    de Boer, JF
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (12) : 1715 - 1720
  • [6] Relationship between retinal nerve fiber layer measurement and signal strength in optical coherence tomography
    Cheung, Carol Yim Lui
    Leung, Christopher Kai Shun
    Lin, Dusheung
    Pang, Chi-Pui
    Lam, Dennis Shun Chiu
    [J]. OPHTHALMOLOGY, 2008, 115 (08) : 1347 - 1351
  • [7] Role of optic nerve imaging in glaucoma clinical practice and clinical trials
    Greenfield, David S.
    Weinreb, Robert N.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (04) : 598 - 603
  • [8] THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE
    HANLEY, JA
    MCNEIL, BJ
    [J]. RADIOLOGY, 1982, 143 (01) : 29 - 36
  • [9] OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA
    HEE, MR
    IZATT, JA
    SWANSON, EA
    HUANG, D
    SCHUMAN, JS
    LIN, CP
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) : 325 - 332
  • [10] Early glaucoma detection using the Humphrey Matrix perimeter, GDx VCC, stratus OCT, and retinal nerve fiber layer photography
    Hong, Samin
    Ahn, Hyunseok
    Ha, Seung Joo
    Yeom, Ho Yeop
    Seong, Gong Je
    Hong, Young Jae
    [J]. OPHTHALMOLOGY, 2007, 114 (02) : 210 - 215